Automated segmentation of an intensity calibration phantom in clinical CT images using a convolutional neural network

被引:10
作者
Uemura, Keisuke [1 ,2 ]
Otake, Yoshito [1 ]
Takao, Masaki [3 ]
Soufi, Mazen [1 ]
Kawasaki, Akihiro [1 ]
Sugano, Nobuhiko [2 ]
Sato, Yoshinobu [1 ]
机构
[1] Nara Inst Sci & Technol, Div Informat Sci, Grad Sch Sci & Technol, Ikoma, Nara, Japan
[2] Osaka Univ, Grad Sch Med, Dept Orthopaed Med Engn, Suita, Osaka, Japan
[3] Osaka Univ, Dept Orthopaed, Grad Sch Med, Suita, Osaka, Japan
基金
日本学术振兴会;
关键词
Artificial intelligence; Bone mineral density; Deep learning; Quantitative computed tomography; Phantom segmentation; U-net; BONE; ASSOCIATION; STRENGTH; DENSITY;
D O I
10.1007/s11548-021-02345-w
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose In quantitative computed tomography (CT), manual selection of the intensity calibration phantom's region of interest is necessary for calculating density (mg/cm(3)) from the radiodensity values (Hounsfield units: HU). However, as this manual process requires effort and time, the purposes of this study were to develop a system that applies a convolutional neural network (CNN) to automatically segment intensity calibration phantom regions in CT images and to test the system in a large cohort to evaluate its robustness. Methods This cross-sectional, retrospective study included 1040 cases (520 each from two institutions) in which an intensity calibration phantom (B-MAS200, Kyoto Kagaku, Kyoto, Japan) was used. A training dataset was created by manually segmenting the phantom regions for 40 cases (20 cases for each institution). The CNN model's segmentation accuracy was assessed with the Dice coefficient, and the average symmetric surface distance was assessed through fourfold cross-validation. Further, absolute difference of HU was compared between manually and automatically segmented regions. The system was tested on the remaining 1000 cases. For each institution, linear regression was applied to calculate the correlation coefficients between HU and phantom density. Results The source code and the model used for phantom segmentation can be accessed at https://github.com/keisuke-uemura/CT-Intensity-Calibration-Phantom-Segmentation. The median Dice coefficient was 0.977, and the median average symmetric surface distance was 0.116 mm. The median absolute difference of the segmented regions between manual and automated segmentation was 0.114 HU. For the test cases, the median correlation coefficients were 0.9998 and 0.999 for the two institutions, with a minimum value of 0.9863. Conclusion The proposed CNN model successfully segmented the calibration phantom regions in CT images with excellent accuracy.
引用
收藏
页码:1855 / 1864
页数:10
相关论文
共 23 条
  • [1] Determination of the Hounsfield value for CT-based design of custom femoral stems
    Aamodt, A
    Kvistad, KA
    Andersen, E
    Lund-Larsen, J
    Eine, J
    Benum, P
    Husby, OS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01): : 143 - 147
  • [2] Quantitative computed tomography
    Adams, Judith E.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2009, 71 (03) : 415 - 424
  • [3] Camacho PM, 2020, ENDOCR PRACT, V26, P1, DOI 10.4158/GL-2020-0524SUPPL
  • [4] Application of a deep learning algorithm for detection and visualization of hip fractures on plain pelvic radiographs
    Cheng, Chi-Tung
    Ho, Tsung-Ying
    Lee, Tao-Yi
    Chang, Chih-Chen
    Chou, Ching-Cheng
    Chen, Chih-Chi
    Chung, I-Fang
    Liao, Chien-Hung
    [J]. EUROPEAN RADIOLOGY, 2019, 29 (10) : 5469 - 5477
  • [5] MEASURES OF THE AMOUNT OF ECOLOGIC ASSOCIATION BETWEEN SPECIES
    DICE, LR
    [J]. ECOLOGY, 1945, 26 (03) : 297 - 302
  • [6] FEIT A, 2019, J CLIN DENSITOM J IN
  • [7] Opportunistic Use of CT Imaging for Osteoporosis Screening and Bone Density Assessment A Qualitative Systematic Review
    Gausden, Elizabeth B.
    Nwachukwu, Benedict U.
    Schreiber, Joseph J.
    Lorich, Dean G.
    Lane, Joseph M.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2017, 99 (18) : 1580 - 1590
  • [8] The Effect of Quantitative Computed Tomography Acquisition Protocols on Bone Mineral Density Estimation
    Giambini, Hugo
    Dragomir-Daescu, Dan
    Huddleston, Paul M.
    Camp, Jon J.
    An, Kai-Nan
    Nassr, Ahmad
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2015, 137 (11):
  • [9] Quantitative Computed Tomography (QCT) of the Distal Forearm in Men Using a Spiral Whole-Body CT Scanner - Description of a Method and Reliability Assessment of the QCT Pro Software
    Hanusch, Birgit C.
    Tuck, Stephen P.
    Mekkayil, Babitha
    Shawgi, Mohamed
    McNally, Richard J. Q.
    Walker, Julie
    Francis, Roger M.
    Datta, Harish K.
    [J]. JOURNAL OF CLINICAL DENSITOMETRY, 2020, 23 (03) : 418 - 425
  • [10] Automated Muscle Segmentation from Clinical CT Using Bayesian U-Net for Personalized Musculoskeletal Modeling
    Hiasa, Yuta
    Otake, Yoshito
    Takao, Masaki
    Ogawa, Takeshi
    Sugano, Nobuhiko
    Sato, Yoshinobu
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (04) : 1030 - 1040