Determination of trace levels of Pb(II) in tap water by anodic stripping voltammetry with boron-doped diamond electrode

被引:43
作者
Chooto, Pipat [1 ]
Wararatananurak, Puchong [1 ]
Innuphat, Chalermpol [1 ]
机构
[1] Prince Songkla Univ, Dept Chem, Fac Sci, Songkhla 90110, Thailand
来源
SCIENCEASIA | 2010年 / 36卷 / 02期
关键词
voltammetry analysis; heavy metals; electrochemistry; square wave anodic stripping voltammetry; MERCURY FILM ELECTRODE; ELECTROCHEMICAL DETECTION; THIN-FILM; LEAD; CADMIUM; SAMPLES; NICKEL; COBALT; PB2+;
D O I
10.2306/scienceasia1513-1874.2010.36.150
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
A simple and rapid method to determine traces of lead in the Hatyai city tap water by square wave anodic stripping voltammetry with boron-doped diamond (BDD) electrode has been studied. The parameters in the preconcentration steps studied included electrolyte influences, pH, concentrations, and volumes. Effective preconcentration of trace lead was achieved at pH 1.26. The appropriate electrolyte was found to be 0.2 M potassium nitrate (KNO3) in 0.05 M nitric acid (HNO3) to make all lead ionic. By scanning potential in the negative direction, the stripping peak was obtained at -0.460 V. Instrumental parameters including deposition potential, deposition time, scan rate, modulation amplitude, and equilibration time were optimized. The detection limit was found to be 0.3 mu g/l. The accuracy was verified by analysing the standard reference material. Several interfering ions were studied and it was found that the method was not significantly affected by these coexisting ions and can be applied satisfactorily to lead determination in water samples. The results from determination of Pb2+ using the studied method and ICP-OES were compared and found to be in good agreement. The concentrations of lead in various water samples from eleven regions in Hatyai city were found to be in the range of 0.0-0.8 mu g/l, lower than the drinking water contamination standard (< 10.0 mu g/l) issued by the World Health Organization.
引用
收藏
页码:150 / 156
页数:7
相关论文
共 26 条
[1]  
AI G., 2000, J. Environ. Med., V2, P11
[2]  
AIKAMPHON K, 1978, ADSORPTION DISTRIBUT
[3]   Electrochemical detection of trace concentrations of cadmium and lead with a boron-doped diamond electrode:: Effect of KCl and KNO3 electrolytes, interferences and measurement in river water [J].
Babyak, C ;
Smart, RR .
ELECTROANALYSIS, 2004, 16 (03) :175-182
[4]   The silver electrode in square-wave anodic stripping voltammetry. Determination of Pb2+ without removal of oxygen [J].
Brand, M ;
Eshkenazi, I ;
KirovaEisner, E .
ANALYTICAL CHEMISTRY, 1997, 69 (22) :4660-4664
[5]   Adsorption and electrochemical reduction of Co(II)-dimethylglyoxime on mercury [J].
Cordon, F ;
Ramírez, SA ;
Gordillo, GJ .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2002, 534 (02) :131-141
[6]   Detection of trace levels of Pb2+ in tap water at boron-doped diamond electrodes with anodic stripping voltammetry [J].
Dragoe, D ;
Spataru, N ;
Kawasaki, R ;
Manivannan, A ;
Spataru, T ;
Tryk, DA ;
Fujishima, A .
ELECTROCHIMICA ACTA, 2006, 51 (12) :2437-2441
[7]  
*EPA, 2005, 3TS RED LEAD DRINK W
[8]   Direct and simultaneous voltammetric analysis of heavy metals in tap water samples at Assiut city: an approach to improve the analysis time for nickel and cobalt determination at mercury film electrode [J].
Farghaly, OA .
MICROCHEMICAL JOURNAL, 2003, 75 (02) :119-131
[9]   Anodic stripping voltammetry of lead and cadmium using a mercury film electrode and thiocyanate [J].
Fischer, E ;
van den Berg, CMG .
ANALYTICA CHIMICA ACTA, 1999, 385 (1-3) :273-280
[10]  
Goyer R, 1995, BASIC SCI POISONS, P691