Pressure Regulating Ear Plug Testing in a Pressure Chamber

被引:11
作者
Jumah, Masen Dirk [1 ]
Schlachta, Miriam [1 ]
Hoelzl, Matthias [1 ]
Werner, Andreas [2 ]
Sedlmaier, Benedikt [1 ]
机构
[1] Charite, Dept Otolaryngol Head & Neck Surg, D-10117 Berlin, Germany
[2] Ctr Space Med Berlin ZWMB, Dept Physiol, Berlin, Germany
来源
AVIATION SPACE AND ENVIRONMENTAL MEDICINE | 2010年 / 81卷 / 06期
关键词
ear barotrauma; eustachian tube; middle ear ventilation; flying discomfort; impedance measurement; EUSTACHIAN-TUBE FUNCTION; EQUALIZING EARPLUGS; OTIC BAROTRAUMA; AIR-TRAVEL; CHILDREN; PREVALENCE; BAROTITIS; ALTITUDE;
D O I
10.3357/ASEM.2717.2010
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
JUMAH MD, SCHLACHTA M, HOELZL M, WERNER A, SEDLMAIER B. Pressure regulating ear plug testing in a pressure chamber. Aviat Space Environ Med 2010; 81:560-5. Introduction: Middle ear barotrauma is a condition frequently associated with flying. It is usually caused by Eustachian tube (ET) dysfunction. Pressure-regulating earplugs (PREP) should improve complaints due to pressure equalization problems. Methods: There were 21 patients with a history of pressure equalization problems while flying who were examined. ET function was measured with exploratory tests and in a pressure chamber. In a double-blind study, PREP were examined using continuous impedance measurement in the pressure chamber. Eardrum deflection and pressure-equalizing maneuvers were also examined. During pressure exposure the subjective state of patients with and without PREP was compared. Results: Evidence of ET dysfunction was found in twice as many patients with the impedance method in a pressure chamber compared to tympanometry or the Valsalva test. Use of PREP reduced the rate of pressure changes in the external auditory canal. Maximum pressure was reached with a delay of about 7 min. The number of pressure-equalizing maneuvers did not differ significantly. Using PREP the patients reported a significantly better subjective state on the VAS scale of 2.19 +/- 1.50 compared to 3.38 +/- 2.33 without PREP. Conclusions: Continuous impedance measurements in the pressure chamber identified ET dysfunction more often than the exploratory ET function tests commonly used in routine diagnostics. This novel finding requires further validation. Although use of PREP showed no signs of improving ET function, it did significantly improve individual subjective state during sudden pressure changes.
引用
收藏
页码:560 / 565
页数:6
相关论文
共 20 条
[1]   INTERSTITIAL-LYMPHATIC MECHANISMS IN THE CONTROL OF EXTRACELLULAR FLUID VOLUME [J].
AUKLAND, K ;
REED, RK .
PHYSIOLOGICAL REVIEWS, 1993, 73 (01) :1-78
[2]  
Berger EH, 2006, AVIAT SPACE ENVIR MD, V77, P766
[3]  
Buchanan B, 1999, ARCH PEDIAT ADOL MED, V153, P466
[4]  
FULCO CS, 1989, AVIAT SPACE ENVIR MD, V60, P1049
[5]   RELIABILITY IN TESTS OF THE EUSTACHIAN-TUBE FUNCTION [J].
GROTH, P ;
IVARSSON, A ;
TJERNSTROM, O .
ACTA OTO-LARYNGOLOGICA, 1982, 93 (3-4) :261-267
[6]   Passenger well-being in airplanes [J].
Hinninghofen, H. ;
Enck, P. .
AUTONOMIC NEUROSCIENCE-BASIC & CLINICAL, 2006, 129 (1-2) :80-85
[7]  
Hüttenbrink KB, 2002, HNO, V50, P863, DOI 10.1007/s00106-002-0670-4
[8]  
Ingelstedt S., 1967, Acta Otolaryngol, V228, P1
[10]  
JERGER J, 1970, ARCHIV OTOLARYNGOL, V92, P311