Parameters of Implant Stability Measurements Based on Resonance Frequency and Damping Capacity: A Comparative Finite Element Analysis

被引:0
|
作者
Winter, Werner [2 ]
Moehrle, Stefan [2 ]
Holst, Stefan [1 ]
Karl, Matthias [1 ]
机构
[1] Univ Erlangen Nurnberg, Dept Prosthodont, D-91054 Erlangen, Germany
[2] Univ Erlangen Nurnberg, Inst Appl Mech, D-91054 Erlangen, Germany
关键词
damping capacity; finite element analysis; implant stability; Osstell Mentor; Periotest; resonance frequency; DENTAL IMPLANT; IN-VITRO; NUMERICAL APPROACH; BONE LOSS; OSSEOINTEGRATION; PERIOTEST; BEHAVIOR;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: Contradictory results have been reported on the comparability of implant stability measurements performed with the Periotest and the Osstell Mentor devices. The purpose of the present finite element analysis was to simulate the influence of the parameters implant length, bone quality (cortical thickness and damping factor), bone loss, and quality of transducer fixation on resonance frequency analysis (RFA) and damping capacity measurements. Materials and Methods: Three-dimensional finite element models of implants placed in human mandibular bone were designed for the simulation of Periotest (Periotest value; PTV) and RFA (implant stability quotient) measurements. Three values for each of the parameters implant length, damping capacity of cortical and trabecular bone, thickness of cortical bone, bone loss, and quality of transducer fixation were obtained. Measurements were simulated at four stages of osseointegration. Results: For all parameters, an increase in implant stability was found with increasing levels of osseointegration. Implant stability was positively correlated with implant length and thickness of cortical bone, with slightly converging values at increased levels of osseointegration. Varying the damping factor of bone had no significant effect. Implant stability was negatively correlated with bone loss, with slightly converging values at increased levels of osseointegration. Linear changes in implant length and bone loss caused nonlinear effects in implant stability values. Stiffness of transducer fixation had an impact on RFA measurements when values below 10 GPa were applied. Conclusion: Although both measuring devices reacted similarly when different parameters of implant stability were changed, good correlation between Periotest values and implant stability quotients was observed only when measurement values of implants without bone loss were considered. INT J ORAL MAXILLOFAC IMPLANTS 2010;25:532-539
引用
收藏
页码:532 / 539
页数:8
相关论文
共 50 条
  • [1] Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study
    Lee, Jungwon
    Pyo, Se-Wook
    Cho, Hyun-Jae
    An, Jung-Sub
    Lee, Jae-Hyun
    Koo, Ki-Tae
    Lee, Yong-Moo
    JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2020, 50 (01) : 56 - 66
  • [2] Influence of Implant Geometry on Primary Insertion Stability and Simulated Pen-implant Bone Loss: An In Vitro Study Using Resonance Frequency Analysis and Damping Capacity Assessment
    Lachmann, Stefan
    Laval, Johannes Yves
    Axmann, Detlef
    Weber, Heiner
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2011, 26 (02) : 347 - 355
  • [3] Influence of Osseointegration Degree and Pattern on Resonance Frequency in the Assessment of Dental Implant Stability Using Finite Element Analysis
    Deng, Bin
    Tan, Keson B.
    Liu, Gui Rong
    Lu, Yi
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (06) : 1082 - 1088
  • [4] Measurement of dental implant stability by resonance frequency analysis and damping capacity assessment: Comparison of both techniques in a clinical trial
    Zix, Juergen
    Hug, Stefan
    Kessler-Liechti, Gerda
    Mericske-Stern, Regina
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (03) : 525 - 530
  • [5] Finite element based bone remodeling and resonance frequency analysis for osseointegration assessment of dental implants
    Li, Wei
    Lin, Daniel
    Rungsiyakull, Chaiy
    Zhou, Shiwei
    Swain, Michael
    Li, Qing
    FINITE ELEMENTS IN ANALYSIS AND DESIGN, 2011, 47 (08) : 898 - 905
  • [6] Resonance frequency analysis and damping capacity assessment -: Part I:: an in vitro study on measurement reliability and a method of comparison in the determination of primary dental implant stability
    Lachmann, S
    Jäger, B
    Axmann, D
    Gomez-Roman, G
    Groten, M
    Weber, H
    CLINICAL ORAL IMPLANTS RESEARCH, 2006, 17 (01) : 75 - 79
  • [7] Optimization of Pedicle Screw Parameters for Enhancing Implant Stability Based on Finite Element Analysis
    Li, Qiang
    Zhao, Hu
    Yang, Jinshuai
    Song, Shihong
    Liu, Xuyan
    WORLD NEUROSURGERY, 2024, 183 : E345 - E354
  • [8] Torsional resonance frequency analysis: a novel method for assessment of dental implant stability
    Tang, Yu-Long
    Li, Bing
    Jin, Wei
    Li, De-Hua
    CLINICAL ORAL IMPLANTS RESEARCH, 2015, 26 (06) : 615 - 622
  • [9] Longitudinal Implant Stability Measurements Based on Resonance Frequency Analysis After Placement in Healed or Regenerated Bone
    Deli, Giorgio
    Petrone, Vincenzo
    De Risi, Valeria
    Tadic, Drazen
    Zafiropoulos, Gregory-George
    JOURNAL OF ORAL IMPLANTOLOGY, 2014, 40 (04) : 438 - 447
  • [10] In vitro comparative analysis of two resonance frequency measurement devices: Osstell implant stability coefficient and Penguin resonance frequency analysis
    Herrero-Climent, Mariano
    Falcao, Artur
    Lopez-Jarana, Paula
    Diaz-Castro, Carmen M.
    Rios-Carrasco, Blanca
    Vicente Rios-Santos, Jose
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2019, 21 (06) : 1124 - 1131