Development of a fast Monte Carlo dose calculation system for online adaptive radiation therapy quality assurance

被引:20
|
作者
Wang, Yuhe [1 ]
Mazur, Thomas R. [1 ]
Park, Justin C. [1 ]
Yang, Deshan [1 ]
Mutic, Sasa [1 ]
Li, H. Harold [1 ]
机构
[1] Washington Univ, Sch Med, Dept Radiat Oncol, St Louis, MO 63110 USA
关键词
GPU; Monte Carlo; variance reduction; MRI guided radiation therapy; adaptive radiation therapy; RADIOTHERAPY; PHOTON; GEANT4; IMPLEMENTATION; SIMULATION; TRANSPORT; EGSNRC; CODE;
D O I
10.1088/1361-6560/aa6e38
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Online adaptive radiation therapy (ART) based on real-time magnetic resonance imaging represents a paradigm-changing treatment scheme. However, conventional quality assurance (QA) methods based on phantom measurements are not feasible with the patient on the treatment couch. The purpose of this work is to develop a fast Monte Carlo system for validating online re-optimized tri-Co-60 IMRT adaptive plans with both high accuracy and speed. The Monte Carlo system is based on dose planning method (DPM) code with further simplification of electron transport and consideration of external magnetic fields. A vendor-provided head model was incorporated into the code. Both GPU acceleration and variance reduction were implemented. Additionally, to facilitate real-time decision support, a C++ GUI was developed for visualizing 3D dose distributions and performing various analyses in an online adaptive setting. A thoroughly validated Monte Carlo code (gPENELOPE) was used to benchmark the new system, named GPU-accelerated DPM with variance reduction (gDPMvr). The comparison using 15 clinical IMRT plans demonstrated that gDPMvr typically runs 43 times faster with only 0.5% loss in accuracy. Moreover, gDPMvr reached 1% local dose uncertainty within 2.3 min on average, and thus is well-suited for ART QA.
引用
收藏
页码:4970 / 4990
页数:21
相关论文
共 50 条
  • [31] Accurate Monte Carlo simulations for nozzle design, commissioning and quality assurance for a proton radiation therapy facility
    Paganetti, H
    Jiang, H
    Lee, SY
    Kooy, HM
    MEDICAL PHYSICS, 2004, 31 (07) : 2107 - 2118
  • [32] Calculation of radiation dose to the lens of the eye using Monte Carlo simulation
    Meriç, N
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (04) : 557 - 560
  • [33] A phase space model of a Versa HD linear accelerator for application to Monte Carlo dose calculation in a real-time adaptive workflow
    Bedford, James L.
    Nilawar, Rahul
    Nill, Simeon
    Oelfke, Uwe
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (09):
  • [34] Treatment plan prescreening for patient-specific quality assurance measurements using independent Monte Carlo dose calculations
    Xu, Yuan
    Zhang, Ke
    Liu, Zhiqiang
    Liang, Bin
    Ma, Xiangyu
    Ren, Wenting
    Men, Kuo
    Dai, Jianrong
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [35] Monte Carlo calculation of rectal dose when using an intrarectal balloon during prostate radiation therapy
    Song, Jun S.
    Court, Laurence E.
    Cormack, Robert A.
    MEDICAL DOSIMETRY, 2007, 32 (03) : 151 - 156
  • [36] Monte Carlo dose enhancement studies in microbeam radiation therapy
    Martinez-Rovira, I.
    Prezado, Y.
    MEDICAL PHYSICS, 2011, 38 (07) : 4430 - 4439
  • [37] Development and benchmarking of the first fast Monte Carlo engine for helium ion beam dose calculation: MonteRay
    Lysakovski, Peter
    Besuglow, Judith
    Kopp, Benedikt
    Mein, Stewart
    Tessonnier, Thomas
    Ferrari, Alfredo
    Haberer, Thomas
    Debus, Juergen
    Mairani, Andrea
    MEDICAL PHYSICS, 2023, 50 (04) : 2510 - 2524
  • [38] Monte Carlo modeling of radiation dose from radiation therapy with superficial x-rays
    Barghash, Reham
    Martin, Tiffany W.
    Prebble, Amber R.
    Leary, Del
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025,
  • [39] GPU-based fast Monte Carlo simulation for radiotherapy dose calculation
    Jia, Xun
    Gu, Xuejun
    Graves, Yan Jiang
    Folkerts, Michael
    Jiang, Steve B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22) : 7017 - 7031
  • [40] Development of a GPU-based Monte Carlo dose calculation code for coupled electron-photon transport
    Jia, Xun
    Gu, Xuejun
    Sempau, Josep
    Choi, Dongju
    Majumdar, Amitava
    Jiang, Steve B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (11) : 3077 - 3086