A simple, fast, low-cost, portable, and eco-friendly method using a spot test on a paper platform, together with diffuse reflectance spectroscopy, was developed and validated for the quantification of aluminum hydrochloride, a potential neurotoxic agent, in antiperspirant samples. The determination of aluminum hydrochloride was performed at a wavelength of 615 nm, by measuring consumption of the purple colorimetric reagent Alizarin S, due to reaction with aluminum. The linear range was from 10.0 to 125.0 mg L-1 and could be described by the equation: A(R) = 0.4479-0.002543 C-AI (R = 0.999). The limits of detection (LOD) and quantification (LOQ) were 3.06 and 10.2 mg L-1 respectively. The method was specific, accurate, and repeatable, with relative standard deviation (RSD) <5.0%. The recovery was between 92.2 and 103.4%. The method was successfully used for the determination of aluminum hydrochloride in commercial antiperspirant samples, revealing concentrations below the maximum permitted by current legislation. (C) 2018 Elsevier B.V. All rights reserved.