Large-area single-photon-counting CdTe detector for synchrotron radiation computed tomography: adedicated pre-processing procedure

被引:32
作者
Brombal, Luca [1 ,2 ]
Donato, Sandro [1 ,2 ]
Brun, Francesco [2 ]
Delogu, Pasquale [3 ,4 ]
Fanti, Viviana [5 ,6 ]
Oliva, Piernicola [6 ,7 ]
Rigon, Luigi [1 ,2 ]
Di Trapani, Vittorio [3 ,4 ]
Longo, Renata [1 ,2 ]
Golosio, Bruno [5 ,6 ]
机构
[1] Univ Trieste, Dept Phys, I-34127 Trieste, Italy
[2] INFN, Div Trieste, I-34127 Trieste, Italy
[3] Univ Siena, Dept Phys Sci Earth & Environm, I-53100 Siena, Italy
[4] INFN, Div Pisa, I-34127 Pisa, Italy
[5] Univ Cagliari, Dept Phys, I-09042 Monserrato, CA, Italy
[6] INFN, Div Cagliari, I-09042 Monserrato, CA, Italy
[7] Univ Sassari, Dept Chem & Pharm, I-07100 Sassari, Italy
关键词
single-photon counting; CdTe; pre-processing; dynamic correction; breast-CT; PHASE RETRIEVAL; IMAGE;
D O I
10.1107/S1600577518006197
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Large-area CdTe single-photon-counting detectors are becoming more and more attractive in view of low-dose imaging applications due to their high efficiency, low intrinsic noise and absence of a scintillating screen which affects spatial resolution. At present, however, since the dimensions of a single sensor are small (typically a few cm(2)), multi-module architectures are needed to obtain a large field of view. This requires coping with inter-module gaps and with close-to-edge pixels, which generally show a non-optimal behavior. Moreover, high-Z detectors often show gain variations in time due to charge trapping: this effect is detrimental especially in computed tomography (CT) applications where a single tomographic image requires hundreds of projections continuously acquired in several seconds. This work has been carried out at the SYRMEP beamline of the Elettra synchrotron radiation facility (Trieste, Italy), in the framework of the SYRMA-3D project, which aims to perform the world's first breast-CT clinical study with synchrotron radiation. An ad hoc data pre-processing procedure has been developed for the PIXIRAD-8 CdTe single-photon-counting detector, comprising an array of eight 30.7mm x 24.8mm modules tiling a 246mm x 25mm sensitive area, which covers the full synchrotron radiation beam. The procedure consists of five building blocks, namely dynamic flat-fielding, gap seaming, dynamic ring removal, projection despeckling and around-gap equalization. Each block is discussed and compared, when existing, with conventional approaches. The effectiveness of the pre-processing is demonstrated for phase-contrast CT images of a human breast specimen. The dynamic nature of the proposed procedure, which provides corrections dependent upon the projection index, allows the effective removal of time-dependent artifacts, preserving the main image features including phase effects.
引用
收藏
页码:1068 / 1077
页数:10
相关论文
共 30 条
[1]   Medical applications of synchrotron radiation at the SYRMEP beamline of ELETTRA [J].
Abrami, A ;
Arfelli, F ;
Barroso, RC ;
Bergamaschi, A ;
Billé, F ;
Bregant, P ;
Brizzi, F ;
Casarin, K ;
Castelli, E ;
Chenda, V ;
Palma, LD ;
Dreossi, D ;
Fava, A ;
Longo, R ;
Mancini, L ;
Menk, RH ;
Montanari, F ;
Olivo, A ;
Pani, S ;
Pillon, A ;
Quai, E ;
Kaiser, SR ;
Rigon, L ;
Rokvic, T ;
Tonutti, M ;
Tromba, G ;
Vaseotto, A ;
Venanzi, C ;
Zanconati, F ;
Zanetti, A ;
Zanini, F .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 548 (1-2) :221-227
[2]   Evaluation of Polarization Effects of e- Collection Schottky CdTe Medipix3RX Hybrid Pixel Detector [J].
Astromskas, Vytautas ;
Gimenez, Eva N. ;
Lohstroh, Annika ;
Tartoni, Nicola .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2016, 63 (01) :252-258
[3]   Review of hybrid pixel detector readout ASICs for spectroscopic X-ray imaging [J].
Ballabriga, R. ;
Alozy, J. ;
Campbell, M. ;
Frojdh, E. ;
Heijne, E. H. M. ;
Koenig, T. ;
Llopart, X. ;
Marchal, J. ;
Pennicard, D. ;
Poikela, T. ;
Tlustos, L. ;
Valerio, P. ;
Wong, W. ;
Zuber, M. .
JOURNAL OF INSTRUMENTATION, 2016, 11
[4]   Chromatic X-ray imaging with a fine pitch CdTe sensor coupled to a large area photon counting pixel ASIC [J].
Bellazzini, R. ;
Spandre, G. ;
Brez, A. ;
Minuti, M. ;
Pinchera, M. ;
Mozzo, P. .
JOURNAL OF INSTRUMENTATION, 2013, 8
[5]   Interface-specific x-ray phase retrieval tomography of complex biological organs [J].
Beltran, M. A. ;
Paganin, D. M. ;
Siu, K. K. W. ;
Fouras, A. ;
Hooper, S. B. ;
Reser, D. H. ;
Kitchen, M. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (23) :7353-7369
[6]   Compensation of ring artefacts in synchrotron tomographic images [J].
Boin, Mirko ;
Haibel, Astrid .
OPTICS EXPRESS, 2006, 14 (25) :12071-12075
[7]  
Brombal L., 2018, P SPIE
[8]   Inpainting approaches to fill in detector gaps in phase contrast computed tomography [J].
Brun, F. ;
Delogu, P. ;
Longo, R. ;
Dreossi, D. ;
Rigon, L. .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2018, 29 (01)
[9]   Phase retrieval in X-ray phase-contrast imaging suitable for tomography [J].
Burvall, Anna ;
Lundstrom, Ulf ;
Takman, Per A. C. ;
Larsson, Daniel H. ;
Hertz, Hans M. .
OPTICS EXPRESS, 2011, 19 (11) :10359-10376
[10]   A proposal for a quality control protocol in breast CT with synchrotron radiation [J].
Contillo, Adriano ;
Veronese, Anna ;
Brombal, Luca ;
Donato, Sandro ;
Rigon, Luigi ;
Taibi, Angelo ;
Tromba, Giuliana ;
Longo, Renata ;
Arfelli, Fulvia .
RADIOLOGY AND ONCOLOGY, 2018, 52 (03) :329-336