Detection limit estimated from slope of calibration curve: An application to competitive ELISA

被引:63
作者
Hayashi, Y
Matsuda, R
Ito, K
Nishimura, W
Imai, K
Maeda, M
机构
[1] Natl Inst Hlth Sci, Tokyo 1538501, Japan
[2] Showa Univ, Sch Pharmaceut Sci, Tokyo 1428555, Japan
[3] Musashino Univ, Pharmaceut Sci Res Inst, Nishitokyo 2028585, Taiwan
关键词
D O I
10.2116/analsci.21.167
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper theoretically derives a general rule that while the slope of the semi-logarithmic plot (Y vs. log X) of a calibration curve varies depending on analyte concentration, X, the slope takes a specific value at the detection limit (L-D)This rule holds good irrespective of the shape of the calibration curve (linear or non-linear) and in this paper, is applied to competitive ELISA (enzyme linked immunosorbent assay). The following, relationship is deduced: slope of log-dose BIBO at L-D = [relative standard deviation (RSD) of blank responses] divided by 0.13. The L-D obtained from the above-mentioned slope corresponds to the dose at which the RSD of dose estimates is 0.3 (= 30%). A commercial kit for 17alpha-hydroxyprogesterone is taken as an example.
引用
收藏
页码:167 / 169
页数:3
相关论文
共 9 条
  • [1] [Anonymous], 1996, ICH Harmonized Tripartite-Guideline for Good Clinical Practice E6 (R1)
  • [2] BORTH R, 1970, ACTA ENDOCR-COP S, V147, P32
  • [3] DUDLEY RA, 1985, CLIN CHEM, V31, P1264
  • [4] EKINS R, 1970, ACTA ENDOCRINOL-COP, P11
  • [5] BASIC PRINCIPLES AND THEORY
    EKINS, RP
    [J]. BRITISH MEDICAL BULLETIN, 1974, 30 (01) : 3 - 11
  • [6] RADIOLIGAND ASSAY
    FINNEY, DJ
    [J]. BIOMETRICS, 1976, 32 (04) : 721 - 740
  • [7] Precision, limit of detection and range of quantitation in competitive ELISA
    Hayashi, Y
    Matsuda, R
    Maitani, T
    Imai, K
    Nishimura, W
    Ito, K
    Maeda, M
    [J]. ANALYTICAL CHEMISTRY, 2004, 76 (05) : 1295 - 1301
  • [9] RODBARD D, 1974, CLIN CHEM, V20, P1255