Impeller design for a miniaturized centrifugal blood pump

被引:7
作者
Takano, T
Schulte-Eistrup, S
Yoshikawa, M
Nakata, K
Kawahito, S
Maeda, T
Nonaka, K
Linneweber, J
Glueck, J
Fujisawa, A
Makinouchi, K
Yokokawa, M
Nosé, Y
机构
[1] Baylor Coll Med, Dept Surg, Houston, TX 77030 USA
[2] Kyocera Corp, Kyoto, Japan
关键词
centrifugal blood pump; impeller design; miniaturization; ventricular assist device; hemolysis; hydraulic performance;
D O I
10.1046/j.1525-1594.2000.06630.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The impeller design for a miniature centrifugal blood pump is an important consideration since the small diameter impeller requires higher rotational speed, which may cause more blood trauma compared to the larger diameter impeller. Three different impeller vanes (straight vanes with a height of 4 mm and 8 mm, and 8 mm curved vanes) of which the diameter was 35 mm were subjected to hydraulic performance and hemolysis tests in the same pump housing. Both straight vane impellers attained left ventricular assist condition (5 L/min against 100 mm Hg) at 2,900 rpm while the curved vane required 3,280 rpm. There was no significant hemolysis difference between the tall and short vanes. The curved impeller vanes did not exhibit sufficient hydraulic performance when compared to the straight vanes. The straight vane impellers, even with different heights, were incorporated into the same pump housings, and the vane heights did not drastically change the hydraulic performance or hemolysis.
引用
收藏
页码:821 / 825
页数:5
相关论文
共 11 条
[1]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[2]  
IMAKI K, 1993, CENTRIFUGAL PUMP DES, P172
[3]  
Kawahito K, 1997, ARTIF ORGANS, V21, P323
[4]   Computational fluid dynamic analyses to establish design process of centrifugal blood pumps [J].
Miyazoe, Y ;
Sawairi, T ;
Ito, K ;
Konishi, Y ;
Yamane, T ;
Nishida, M ;
Masuzawa, T ;
Takiura, K ;
Taenaka, Y .
ARTIFICIAL ORGANS, 1998, 22 (05) :381-385
[5]   Long-term in vivo left ventricular assist device study with a titanium centrifugal pump [J].
Ohtsuka, G ;
Nakata, K ;
Yoshikawa, M ;
Mueller, J ;
Takano, T ;
Yamane, S ;
Gronau, N ;
Glueck, J ;
Takami, Y ;
Sueoka, A ;
Letsou, G ;
Schima, H ;
Schmallegger, H ;
Wolner, E ;
Koyanagi, H ;
Fujisawa, A ;
Baldwin, JC ;
Nose, Y .
ASAIO JOURNAL, 1998, 44 (05) :M619-M623
[6]   Long-term in vivo left ventricular assist device study for 284 days with Gyro PI pump [J].
Ohtsuka, G ;
Nakata, K ;
Yoshikawa, M ;
Takano, T ;
Glueck, J ;
Sankai, Y ;
Takami, Y ;
Mueller, J ;
Sueoka, A ;
Letsou, G ;
Schima, H ;
Schmallegger, H ;
Wolner, E ;
Koyanagi, H ;
Fujisawa, A ;
Baldwin, JC ;
Nosé, Y .
ARTIFICIAL ORGANS, 1999, 23 (06) :504-507
[7]   MINIMIZATION OF HEMOLYSIS IN CENTRIFUGAL BLOOD PUMPS - INFLUENCE OF DIFFERENT GEOMETRIES [J].
SCHIMA, H ;
MULLER, MR ;
PAPANTONIS, D ;
SCHLUSCHE, C ;
HUBER, L ;
SCHMIDT, C ;
TRUBEL, W ;
THOMA, H ;
LOSERT, U ;
WOLNER, E .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1993, 16 (07) :521-529
[8]   In vivo evaluation of the miniaturized Gyro centrifugal pump as an implantable ventricular assist device [J].
Takami, Y ;
Otsuka, G ;
Mueller, J ;
Sugita, Y ;
Nakata, K ;
Tayama, E ;
Ohashi, Y ;
Schima, H ;
Schmallegger, H ;
Wolner, E ;
Nose, Y .
ARTIFICIAL ORGANS, 1998, 22 (08) :713-720
[9]   Development of an antithrombogenic and antitraumatic blood pump:: The Gyro C1E3 [J].
Takano, T ;
Nakata, K ;
Yoshikawa, M ;
Maeda, T ;
Glueck, J ;
Fujisawa, A ;
Makinouchi, K ;
Yokokawa, M ;
Murabayashi, S ;
Nosé, Y .
ASAIO JOURNAL, 2000, 46 (01) :123-127
[10]   Development of design methods of a centrifugal blood pump with in vitro tests, flow visualization, and computational fluid dynamics: Results in hemolysis tests [J].
Takiura, K ;
Masuzawa, T ;
Endo, S ;
Wakisaka, Y ;
Tatsumi, E ;
Taenaka, Y ;
Takano, H ;
Yamane, T ;
Nishida, M ;
Asztalos, B ;
Konishi, Y ;
Miyazoe, Y ;
Ito, K .
ARTIFICIAL ORGANS, 1998, 22 (05) :393-398