Highly sensitive lactate biosensor by engineering chitosan/PVI-Os/CNT/LOD network nanocomposite

被引:110
作者
Cui, Xiaoqiang
Li, Chang Ming
Zang, Jianfeng
Yu, Shucong
机构
[1] Nanyang Technol Univ, Sch Chem & Biomed Engn, Singapore 637457, Singapore
[2] Nanyang Technol Univ, Ctr Adv Bionanosyst, Singapore 637457, Singapore
关键词
lactate biosensor; chitosan; redox polymer; carbon nanotubes; network nanocomposite;
D O I
10.1016/j.bios.2007.03.004
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel chitosan/PVI-Os(polyvinylitnidazole-Os)/CNT(carbon nanotube)/LOD (lactate oxidase) network nanocomposite was constructed on gold electrode for detection of lactate. The composite was nanoengineered by selected matched material components and optimized composition ratio to produce a superior lactate sensor. Positively charged chitosan and PVI-Os were used as the matrix and the mediator to immobilize the negatively charged LOD and to enhance the electron transfer, respectively. CNTs were introduced as the essential component in the composite for the network nanostructure. FESEM (field emission scan electron microscopy) and electrochemical characterization demonstrated that CNT behaved as a cross-linker to network PVI and chitosan due to its nanoscaled and negative charged nature. This significantly improved the conductivity, stability and electroactivity for detection of lactate. The standard deviation of the sensor without CNT in the composite was greatly reduced from 19.6 to 4.9% by addition of CNTs. With optimized conditions the sensitivity and detection limit of the lactate sensor was 19.7 mu A mM(-1) cm(-2) and 5 mu M, respectively. The sensitivity was remarkably improved in comparison to the newly reported values of 0.15-3.85 mu A mM(-1) cm(-2). This novel nanoengineering approach for selecting matched components to form a network nanostructure could be extended to other enzyme biosensors, and to have broad potential applications in diagnostics, life science and food analysis. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:3288 / 3292
页数:5
相关论文
共 33 条
[1]  
[Anonymous], 1987, Biosensors Fundamentals and Applications
[2]   THE RELATIVE PROGNOSTIC VALUE OF LACTATE AND HEMODYNAMIC MEASUREMENTS IN EARLY SHOCK [J].
COWAN, BN ;
BURNS, HJG ;
BOYLE, P ;
LEDINGHAM, IM .
ANAESTHESIA, 1984, 39 (08) :750-755
[3]  
F Allen J Bard L.R., 2001, Electrochemical Methods: Fundamentals and Applications
[4]   Layer-by-layer electrodeposition of redox polymers and enzymes on screen-printed carbon electrodes for the preparation of reagentless biosensors [J].
Gao, Q ;
Yang, MR .
CHEMICAL COMMUNICATIONS, 2004, (01) :30-31
[5]  
Gao ZQ, 2002, ANGEW CHEM INT EDIT, V41, P810, DOI 10.1002/1521-3773(20020301)41:5<810::AID-ANIE810>3.0.CO
[6]  
2-I
[7]   CROSS-LINKED REDOX GELS CONTAINING GLUCOSE-OXIDASE FOR AMPEROMETRIC BIOSENSOR APPLICATIONS [J].
GREGG, BA ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1990, 62 (03) :258-263
[8]   Reagentless amperometric detection of L-lactate on an enzyme-modified conducting copolymer poly(5-hydroxy-1,4-naphthoquinone-co-5-hydroxy-3-thioacetic acid-1,4-naphthoquinone) [J].
Haccoun, J ;
Piro, B ;
Tran, LD ;
Dang, LA ;
Pham, MC .
BIOSENSORS & BIOELECTRONICS, 2004, 19 (10) :1325-1329
[9]   Design of novel electron transfer mediators based on indophenol derivatives for lactate sensor [J].
Hirano, K ;
Yamato, H ;
Kunimoto, K ;
Ohwa, M .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (04) :315-322
[10]   99.9% purity multi-walled carbon nanotubes by vacuum high-temperature annealing [J].
Huang, W ;
Wang, Y ;
Luo, GH ;
Wei, F .
CARBON, 2003, 41 (13) :2585-2590