129I interlaboratory comparison:: Phase II results
被引:15
作者:
Roberts, ML
论文数: 0引用数: 0
h-index: 0
机构:
Univ Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USAUniv Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USA
Roberts, ML
[1
]
Caffee, MW
论文数: 0引用数: 0
h-index: 0
机构:
Univ Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USAUniv Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USA
Caffee, MW
[1
]
机构:
[1] Univ Calif Lawrence Livermore Natl Lab, Ctr Accelerator Mass Spectrometry, Livermore, CA 94551 USA
accelerator mass spectrometry;
iodine-129;
nuclear non-proliferation;
D O I:
10.1016/S0168-583X(00)00211-1
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Large discrepancies seen in a Phase I I-129 Accelerator Mass Spectrometry (AMS) interlaboratory comparison have prompted a subsequent Phase II comparison. In Phase II of the I-129 AMS interlaboratory comparison, three separate laboratories prepared AgI from two environmental samples (IAEA 375 soil and maple leaves). Each laboratory used its own chemical preparation method; each of these methods being considerably different. The resulting six samples (two sets of three) were then distributed to the participating I-129 AMS facilities and I-129/I-127 ratios measured. Results and discussion of the Phase II interlaboratory comparison are presented. (C) 2000 Elsevier Science B.V. All rights reserved.