A computational tool for the probabilistic finite element analysis of an uncemented total hip replacement considering variability in bone-implant version angle

被引:20
作者
Dopico-Gonzalez, Carolina [1 ]
New, Andrew M. [1 ]
Browne, Martin [1 ]
机构
[1] Univ Southampton, Bioengn Res Grp, Sch Engn Sci, Southampton SO17 1BJ, Hants, England
关键词
total hip replacement; finite element analysis; probabilistic analysis; biomechanics; STRAIN DISTRIBUTION; TRABECULAR BONE; MUSCLE FORCES; FEMORAL STEM; PROSTHESES; PARAMETERS; DESIGN; MOTION; RANGE; FEMUR;
D O I
10.1080/10255840902911536
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In the present study, a probabilistic finite element tool was implemented to assess an uncemented total hip replacement including variability in bone-implant version angle. The Monte Carlo method was used with two different performance indicators: the bone maximum nodal von-Mises elastic strain and the bone volume (BV) percentage exceeding specified strain limits. Implant version, bone stiffness and load magnitude were the most sensitive parameters. The results were more consistent using percentage BV under specified limit strains as the performance indicator, even for a low number of simulations. The reliability of the computational tool was demonstrated through a comparison with previous studies, and the consistency of the results for all strain limits investigated.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 35 条
[1]  
[Anonymous], 2002, Bones: Structure and Mechanics
[2]   ELASTIC-MODULUS OF TRABECULAR BONE MATERIAL [J].
ASHMAN, RB ;
RHO, JY .
JOURNAL OF BIOMECHANICS, 1988, 21 (03) :177-181
[3]  
BAH MT, 2003, INT C COMP BIOENG ZA
[4]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[5]   Effect of porous coating and loading conditions on total hip femoral stem stability [J].
Biegler, FB ;
Reuben, JD ;
Harrigan, TP ;
Hou, FJ ;
Akin, JE .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :839-847
[6]   Reliability theory for load bearing biomedical implants [J].
Browne, M ;
Langley, RS ;
Gregson, PJ .
BIOMATERIALS, 1999, 20 (14) :1285-1292
[7]  
Cohen J., 2003, Applied Multiple Regression/Correlation Analysis for the Behavioral Sciences
[8]   INFLUENCE OF THIGH MUSCLES ON THE AXIAL STRAINS IN A PROXIMAL FEMUR DURING EARLY STANCE IN GAIT [J].
CRISTOFOLINI, L ;
VICECONTI, M ;
TONI, A ;
GIUNTI, A .
JOURNAL OF BIOMECHANICS, 1995, 28 (05) :617-624
[9]   Changes in strain distribution of loaded proximal femora caused by different types of cementless femoral stems [J].
Decking, Ralf ;
Puhl, Wolfhart ;
Simon, Ulrich ;
Claes, Lutz E. .
CLINICAL BIOMECHANICS, 2006, 21 (05) :495-501
[10]   Probabilistic analysis of an uncemented total hip replacement [J].
Dopico-Gonzalez, Carolina ;
New, Andrew M. ;
Browne, Martin .
MEDICAL ENGINEERING & PHYSICS, 2009, 31 (04) :470-476