Inaccuracy of the FloTrac/Vigileo™ System in Patients With Low Cardiac Index

被引:17
作者
Maeda, Takuma [1 ]
Yoshitani, Kenji [1 ]
Inatomi, Yuzuru [1 ]
Ohnishi, Yoshihiko [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr, Dept Anesthesiol, Suita, Osaka 5658565, Japan
关键词
cardiac output; intraoperative monitoring; thermodilution; pulse-wave analysis; low cardiac output syndrome; heart failure; TIME 3-DIMENSIONAL ECHOCARDIOGRAPHY; LEFT-VENTRICULAR VOLUMES; CARDIOVASCULAR MAGNETIC-RESONANCE; ACUTE MYOCARDIAL-INFARCTION; WAVE-FORM ANALYSIS; OUTPUT MEASUREMENT; INTERSTUDY REPRODUCIBILITY; PERIPHERAL RESISTANCE; CIRRHOTIC-PATIENTS; VIGILEO-FLOTRAC;
D O I
10.1053/j.jvca.2014.04.013
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives: The goal of this study was to compare cardiac output derived from the FloTrac/Vigileo (TM) system (COFT) with cardiac output measured by 3-dimensional transesophageal echocardiography (CO3D) in patients with severe heart failure undergoing cardiac resynchronization therapy. The impact of preoperative systemic vascular resistance index on the accuracy of the FloTrac/Vigileo (TM) system also was investigated. Design: Prospective clinical study. Setting: Cardiac surgery operating room of a single cardiovascular center. Participants: Forty-one patients undergoing elective cardiac resynchronization therapy lead implantation. Interventions: CO3D as the reference method and COFT were determined simultaneously after induction of anesthesia. Measurements and Main Results: Linear regression analysis showed a poor correlation between CO3D and COFT(R-2 = 0.16). Bland-Altman plots showed wide limits of agreement between CO3D and COFT. Bias was 0.60 +/- 0.63 L/min with a high percentage error of 58.2%. Subgroup analysis showed that the percentage error between CO3D and COFT was 74.1% in patients with a cardiac index <2.2 L/min/m(2) and 17.2% in patients with a cardiac index >= 2.2 L/min/m(2). Systemic vascular resistance index was significantly higher in patients with a cardiac index <2.2 L/min/m(2) (3,037 +/- 820 v 2,461 +/- 878; p = 0.039). Conclusions: The FloTrac/Vigileo (TM) system is not accurate in patients with low cardiac output, especially those with a cardiac index <2.2 L/min/m(2). A high systemic vascular resistance index in patients with low cardiac index may contribute to this inaccuracy. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:1521 / 1526
页数:6
相关论文
共 28 条
[1]   Accuracy of arterial pressure waveform analysis for cardiac output measurement in comparison with thermodilution methods in patients undergoing living donor liver transplantation [J].
Akiyoshi, Kozaburo ;
Kandabashi, Tadashi ;
Kaji, Junko ;
Yamaura, Ken ;
Yoshimura, Hayashi ;
Irita, Kazuo ;
Hoka, Sumio .
JOURNAL OF ANESTHESIA, 2011, 25 (02) :178-183
[2]   Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery [J].
Biancofiore, G. ;
Critchley, L. A. H. ;
Lee, A. ;
Bindi, L. ;
Bisa, M. ;
Esposito, M. ;
Meacci, L. ;
Mozzo, R. ;
DeSimone, P. ;
Urbani, L. ;
Filipponi, F. .
BRITISH JOURNAL OF ANAESTHESIA, 2009, 102 (01) :47-54
[3]   Evaluation of a New Software Version of the FloTrac/Vigileo (Version 3.02) and a Comparison with Previous Data in Cirrhotic Patients Undergoing Liver Transplant Surgery [J].
Biancofiore, Gianni ;
Critchley, Lester A. H. ;
Lee, Anna ;
Yang, Xiao-xing ;
Bindi, Lucia M. ;
Esposito, Massimo ;
Bisa, Massimo ;
Meacci, Luca ;
Mozzo, Roberto ;
Filipponi, Franco .
ANESTHESIA AND ANALGESIA, 2011, 113 (03) :515-522
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   A unique role for 6-O sulfation modification in zebrafish vascular development [J].
Chen, E ;
Stringer, SE ;
Rusch, MA ;
Selleck, SB ;
Ekker, SC .
DEVELOPMENTAL BIOLOGY, 2005, 284 (02) :364-376
[6]   A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques [J].
Critchley, LAH ;
Critchley, JAJH .
JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) :85-91
[7]   Validation and feasibility of Intraoperative three-dimensional transesophageal echocardiographic cardiac output [J].
Culp, William C. ;
Ball, Timothy R. ;
Burnett, Christopher J. .
ANESTHESIA AND ANALGESIA, 2007, 105 (05) :1219-1223
[8]   Performance of 3-Dimensional Echocardiography in Measuring Left Ventricular Volumes and Ejection Fraction [J].
Dorosz, Jennifer L. ;
Lezotte, Dennis C. ;
Weitzenkamp, David A. ;
Allen, Larry A. ;
Salcedo, Ernesto E. .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2012, 59 (20) :1799-1808
[9]   Usefulness of real-time three-dimensional echocardiography for reliable measurement of cardiac output in patients with ischemic or idiopathic dilated cardiomyopathy [J].
Fleming, SM ;
Cumberledge, B ;
Kiesewetter, C ;
Parry, G ;
Kenny, A .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 95 (02) :308-310
[10]   CORRELATIVE CLASSIFICATION OF CLINICAL AND HEMODYNAMIC FUNCTION AFTER ACUTE MYOCARDIAL-INFARCTION [J].
FORRESTER, JS ;
DIAMOND, GA ;
SWAN, HJC .
AMERICAN JOURNAL OF CARDIOLOGY, 1977, 39 (02) :137-145