Implantation of Ultraflex nitinol stents in malignant tracheobronchial stenoses

被引:141
作者
Miyazawa, T
Yamakido, M
Ikeda, S
Furukawa, K
Takiguchi, Y
Tada, H
Shirakusa, T
机构
[1] Hiroshima City Hosp, Dept Pulm Med, Naka Ku, Hiroshima 7308518, Japan
[2] Hiroshima Univ, Dept Internal Med, Hiroshima, Japan
[3] Kyoto Katsura Hosp, Dept Surg, Kyoto, Japan
[4] Tokyo Med Univ, Dept Surg, Tokyo, Japan
[5] Chiba Univ, Dept Internal Med, Chiba, Japan
[6] Osaka City Gen Hosp, Dept Surg, Osaka, Japan
[7] Fukuoka Univ, Dept Surg, Fukuoka 81401, Japan
关键词
malignant tracheobronchial stenosis; nitinol stent; Ultraflex stent;
D O I
10.1378/chest.118.4.959
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Study objective: To assess the uncovered Ultraflex nitinol stent (Boston Scientific; Natick, MA) for its efficacy and safety. Design, setting, and patients: Between October 1997 and October 1998, we carried out a prospective multicenter study at six hospitals in Japan. Fifty-four Ultraflex stents were inserted in 34 patients with inoperable malignant airway stenosis using a flexible and/or a rigid bronchoscope under fluoroscopic and endoscopic visualization. Measurements and results: Clinical, endoscopic examination, and pulmonary function on days 1, 30, and 60 after stent implantation showed improvement. In 19 patients (56%), stent implantation was performed as an emergency procedure because of life-threatening tracheobronchial obstruction. Immediate relief of dyspnea was achieved in 82% of the patients. The dyspnea index improved significantly after implantation (before vs days 1, 30, and 60; p < 0.001). Significant improvements were observed in obstruction of airway diameter (81 +/- 15% before vs 14 +/- 17% on day 1, 12 +/- 12% on day 30, and 22 +/- 28% on day 60; p < 0.001). Vital capacity (VC), FEV1, and peak expiratory flow (PEF) increased significantly after stent implantation: before vs immediately after VC (p < 0.01), FEV1 (p < 0.001), and PEF (p < 0.05). The main complications were tumor ingrowth (24%) and tumor overgrowth (21%). After coagulation with an Nd-YAG laser or argon plasma coagulation, mechanical coring out using the bevel of a rigid bronchoscope was necessary in patients showing tumor ingrowth or overgrowth. Removal and reposition were possible in case of misplacement. There were no problems with migration and retained secretions. The median survival time of patients was 3 months. The 1-year survival rate was 25.4%. Conclusions: In this study of the Ultraflex nitinol stent, we have demonstrated that patients were relieved from dyspnea, which contributed to improved quality of life, with minimal complications. This stent can be used safely, even in the subglottic region. Owing to its excellent flexibility and biocompatibility, the stent is also indicated in certain complicated situations, eg, narrow stenosis, hourglass stenosis, curvilinear stenosis, bilateral mainstem bronchial stenoses, and long stenosis of varying diameters.
引用
收藏
页码:959 / 965
页数:7
相关论文
共 24 条
  • [1] BECKER H, 1995, J BRONCHOL, V2, P252
  • [2] Options and results in endobronchial treatment of lung cancer
    Becker, HD
    [J]. MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 1996, 5 (02): : 165 - 178
  • [3] Becker HD., 1995, J BRONCHOL, V2, P98
  • [4] BOLLIGER CT, 1993, CHEST, V104, P1653, DOI 10.1378/chest.104.6.1653
  • [5] COLT HG, 1995, J BRONCHOL, V4, P263
  • [6] Self-expandable metallic airway stent insertion employing flexible bronchoscopy - Preliminary results
    Dasgupta, A
    Dolmatch, BL
    Abi-Saleh, WJ
    Mathur, PN
    Mehta, AC
    [J]. CHEST, 1998, 114 (01) : 106 - 109
  • [7] DUERIG TW, 1997, MED PLAST BIOMAT MAR, P31
  • [8] A DEDICATED TRACHEOBRONCHIAL STENT
    DUMON, JF
    [J]. CHEST, 1990, 97 (02) : 328 - 332
  • [9] Dumon JF, 1996, J BRONCHOLOGY, V3, P6, DOI 10.1097/00128594-199601000-00003
  • [10] Edell Eric S., 1994, P301