Another cause of difficulty in ventilating a patient

被引:8
作者
Eckhout, GV [1 ]
Bhatia, S [1 ]
机构
[1] Cleveland Clin Fdn, Dept Gen Anesthesiol, Div Anesthesiol & Crit Care Med, Cleveland, OH 44195 USA
关键词
anesthesia delivery systems; anesthetic circuit : mechanical; obstruction; equipment and supplies : anesthesia circuit; expiratory limb;
D O I
10.1016/S0952-8180(02)00515-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Various mechanical problems with the delivery of anesthesia gases have been reported, and preoperative checks of the anesthesia delivery system are designed to prevent these. We report a case of a mechanical obstruction of the expiratory limb of the anesthesia circuit which went undetected before the induction of anesthesia. A check of the circuit was performed before the addition of circuit extensions, thus missing the obstruction. Systematic investigation of the anesthetic system after difficulty in ventilating the patient revealed the obstruction, without any harmful consequence to the patient. This case highlights the importance of a thorough check of the system before any anesthetic, particularly after any change in the circuit. (C) 2003 by Elsevier Science Inc.
引用
收藏
页码:137 / 139
页数:3
相关论文
共 6 条
[1]   Adverse anesthetic outcomes arising from gas delivery equipment - A closed claims analysis [J].
Caplan, RA ;
Vistica, MF ;
Posner, KL ;
Cheney, FW .
ANESTHESIOLOGY, 1997, 87 (04) :741-748
[2]  
HINDMAN BJ, 1986, ANESTHESIOLOGY, V65, P349
[3]   SUDDEN DEVELOPMENT OF ANESTHESIA CIRCUIT OBSTRUCTION BY AN END-TIDAL CO2 CAP IN THE GAS SAMPLING ELBOW [J].
KRENSAVAGE, TJ ;
RICHARDS, E .
ANESTHESIA AND ANALGESIA, 1995, 81 (01) :207-207
[4]  
MCEWAN AI, 1993, ANESTH ANALG, V76, P440
[5]  
Smith C E, 1991, J Clin Anesth, V3, P229, DOI 10.1016/0952-8180(91)90166-K
[6]   Intraoperative breathing circuit obstruction caused by albuterol nebulization [J].
Walton, JS ;
Fears, R ;
Burt, N ;
Dorman, BH .
ANESTHESIA AND ANALGESIA, 1999, 89 (03) :650-651