New distal embolic protection device the fibernet® -: 3 dimensional filter:: First carotid human study

被引:17
作者
Henry, Michel
Polydorou, Antonios
Henry, Isabelle
Liasis, Nikos
Polydorou, Adamantia
Polydorou, Victoria
Demesticha, Theano
Skandalakis, Panayiotis
Kotsiomitis, Evaggelos
Hugel, Michel
Sedgewick, Jerry
Ruth, George
机构
[1] Cabinet Cardiol, F-54000 Nancy, France
[2] Apollo Clin, Global Res Inst, Hyderabad, Andhra Pradesh, India
[3] Panteleimon Gen Hosp, Athens, Greece
[4] Polyclin Bois Bernard, F-62320 Bois Bernard, France
[5] Univ Athens, Dept Anat, GR-10679 Athens, Greece
[6] Univ Minnesota, Biomed Dept, Minneapolis, MN 55455 USA
关键词
angioplasty; CAS; carotid stenting; interventional;
D O I
10.1002/ccd.21129
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Evaluate the performance and safety of the FiberNet (R) Embolic Protection System during carotid artery intervention. Background: Carotid Angioplasty and Stenting (CAS) can be proposed to treat the majority of carotid stenoses. Brain embolization takes place and routine use of Embolic Protection Devices (EPD) is warranted. Many EPDs have significant limitations, which may be addressed by a new EPD, the FiberNet (R) (Lumen Biomedical, Plymouth, MN). Methods: The system consists of a 3-dimensional expandable filter made of fibers, which expand radially, mounted onto a 0.014 '' wire and retrieval catheter. FiberNet can capture particles as small as 40 mu m without compromising flow. Results: 35 lesions treated in 34 patients. Male 67.6%. Age: 71.4 +/- 8.8 (50-85). Average stenosis 84.5% +/- 7.9 (70-99). 29.4% were symptomatic. Technical success: 34/35 (97%). No stroke or death within 30 days. Neurological events: two permanent amaurosis, one amaurosis fugax. All samples visually contained significant amounts of emboli. The mean surface area of debris caught was 63.8 mm(2) (37.7-107.5). Comparisons were made with other EPDs. The mean surface area of debris caught was 12.2 mm(2) (2.7-34.3). No changes were noted in CT/MRI at 30-day post procedure. Conclusion: The first human use of this new novel EPD in carotid artery stenting is encouraging. The FiberNet was easy to use and confirmed the ability to capture particles less than 100 Rm. The feasibility of the FiberNet has been demonstrated. Additional patients will demonstrate the overall safety and efficacy of this new EPD device. (C) 2007 Wiley-Liss, Inc.
引用
收藏
页码:1026 / 1035
页数:10
相关论文
共 78 条
  • [1] Adami CA, 2002, J ENDOVASC THER, V9, P147, DOI 10.1583/1545-1550(2002)009<0147:UOTPAE>2.0.CO
  • [2] 2
  • [3] Multicenter evaluation of carotid artery stenting with a filter protection system
    Al-Mubarak, N
    Colombo, A
    Gaines, PA
    Iyer, SS
    Corvaja, N
    Cleveland, TJ
    Macdonald, S
    Brennan, C
    Vitek, JJ
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (05) : 841 - 846
  • [4] Effect of the distal-balloon protection system on microembolization during carotid stenting
    Al-Mubarak, N
    Roubin, GS
    Vitek, JJ
    Iyer, SS
    New, G
    Leon, MB
    [J]. CIRCULATION, 2001, 104 (17) : 1999 - 2002
  • [5] Cerebral protection during carotid artery stenting -: Collection and histopathologic analysis of embolized debris
    Angelini, A
    Reimers, B
    Della Barbera, M
    Saccà, S
    Pasquetto, G
    Cernetti, C
    Valente, M
    Pascotto, P
    Thiene, G
    [J]. STROKE, 2002, 33 (02) : 456 - 461
  • [6] Silent cerebral ischemia detected by diffusion-weighted MRI after carotid endarterectomy
    Barth, A
    Remonda, L
    Lövblad, KO
    Schroth, G
    Seiler, RW
    [J]. STROKE, 2000, 31 (08) : 1824 - 1828
  • [7] Carotid angioplasty and stenting versus carotid endarterectomy: Randomized trial in a community hospital
    Brooks, WH
    McClure, RR
    Jones, MR
    Coleman, TC
    Breathitt, L
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 38 (06) : 1589 - 1595
  • [8] Brown MM, 2001, LANCET, V357, P1729
  • [9] Coggia M, 2000, J VASC SURG, V31, P550
  • [10] Cremonesi A, 2000, Ital Heart J, V1, P801