Bilateral simultaneous metal inlay dissociation from the polyethylene liner of a metal-on-metal hip replacement

被引:6
|
作者
Girard, J. [1 ]
Soenen, M. [1 ]
Monnin, C. [2 ]
Migaud, H. [1 ]
机构
[1] Univ Lille 2, Dept Orthopaed C, Roger Salengro Hosp, Lille Univ Teaching Hosp Ctr, F-59037 Lille, France
[2] Private Hosp Bois, F-59000 Lille, France
关键词
Total hip arthroplasty; Composite polyethylene liner failure; Prosthetic components dissociation; THR mechanical failure; FOLLOW-UP; ARTHROPLASTY; FRACTURE;
D O I
10.1016/j.otsr.2009.07.003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Hard-on-hard bearings for total hip replacement may require a modular acetabular inlay for which polyethylene is often used in a sandwich-type configuration. However, differences in the elastic modulus of the materials make fixation of this metal insert uncertain. The aim of this study is to report a case of bilateral separation of the metal insert from the polyethylene sandwich in a metal-on-metal bearings prosthesis. Materials and methods: A bilateral total hip arthroplasty was performed in two operations, four months apart, in a 53-year-old woman following a corticosteroid-induced osteonecrosis. The total hip replacement system included a cementless stem, and a press-fit hemispheric cup containing a polyethylene sandwich with a metallic insert (Sikomet (TM)). Results: Three years later, the patient consulted because of abnormal noise in her right hip which appeared normal on conventional X-ray. Three months later she consulted again for persistent noise. Separation of the metal insert from the polyethylene sandwich was diagnosed and an acetabular revision was performed selecting a metal-on-polyethylene articulation system. The postoperative course, for this revision, was uneventful, but the patient returned with the similar symptoms in her left hip four months later, resulting in the same type of revision. During the revisions, osteolysis secondary to metallosis was diagnosed, requiring synovectomy and acetabular reconstruction with morcelized allograft impaction. The left side postoperative course included three dislocations in nine months which were conservatively treated and have not since recurred. Discussion and conclusion: This is the first reported case recording an almost simultaneous bilateral dissociation of a hard-on-hard inlay from its polyethylene sandwich. This bilateral case suggests that the fixation of the metal insert inside the polyethylene was probably defective. This case is also a reminder that mechanical complications (separation, implant fracture) should be searched for in presence of any abnormal noise occurring after hard-on-hard bearings prosthetic implantation. This confirms the necessity of periodical follow-up of hip arthroplasties and the importance of knowing their radiological features. The low carbon content of the Sikomet (TM) bearing may have been the cause of this failure by increasing frictional torque on the bearing surface, causing metallosis which has already been described in the literature in this type of hip replacement system. (c) 2009 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:443 / 446
页数:4
相关论文
共 50 条
  • [21] Metal-on-Metal Hip Replacement: A New Concept for an Old Problem?
    Segal, Ortal
    Maoz-Segal, Ramit
    ISRAEL MEDICAL ASSOCIATION JOURNAL, 2013, 15 (11): : 722 - 724
  • [22] A comparative assessment of small-head metal-on-metal and ceramic-on-polyethylene total hip replacement
    Luebbeke, A.
    Gonzalez, A.
    Garavaglia, G.
    Roussos, C.
    Bonvin, A.
    Stern, R.
    Peter, R.
    Hoffmeyer, P.
    BONE & JOINT JOURNAL, 2014, 96B (07): : 868 - 875
  • [23] Corrosion on the Acetabular Liner Taper from Retrieved Modular Metal-on-Metal Total Hip Replacements
    Gascoyne, Trevor C.
    Dyrkacz, Richard M.
    Turgeon, Thomas R.
    Burnell, Colin D.
    Wyss, Urs P.
    Brandt, Jan-M.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (10): : 2049 - 2052
  • [24] In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty
    Komistek, RD
    Dennis, DA
    Ochoa, JA
    Haas, BD
    Hammill, C
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (10): : 1836 - 1841
  • [25] Metal-on-Metal Hip Resurfacing
    Quesada, Mario J.
    Marker, David R.
    Mont, Michael A.
    JOURNAL OF ARTHROPLASTY, 2008, 23 (07): : 69 - 73
  • [26] Metal-on-metal hip resurfacing
    Roberts, J
    Meek, RMD
    Roberts, P
    Grigoris, P
    SCOTTISH MEDICAL JOURNAL, 2005, 50 (01) : 10 - 12
  • [27] Perfusion MRI in hips with metal-on-metal and metal-on-polyethylene total hip arthroplasty
    Anwander, H.
    Cron, G. O.
    Rakhra, K.
    Beaule, P. E.
    BONE & JOINT RESEARCH, 2016, 5 (03): : 73 - 79
  • [28] Liner dissociation due to locking mechanism failure in Metasul metal-on-metal total hip arthroplasty using an APR cup
    Maezawa, K.
    Nozawa, M.
    Yuasa, T.
    Maruyama, Y.
    Kaneko, K.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2015, 66 (03) : 151 - 152
  • [29] MRI of asymptomatic patients with metal-on-metal and polyethylene-on-metal total hip arthroplasties
    Mistry, A.
    Cahir, J.
    Donell, S. T.
    Nolan, J.
    Toms, A. P.
    CLINICAL RADIOLOGY, 2011, 66 (06) : 540 - 545
  • [30] Metal-on-Metal vs Metal-on-Improved Polyethylene Bearings in Total Hip Arthroplasty
    Molli, Ryan G.
    Lombardi, Adolph V., Jr.
    Berend, Keith R.
    Adams, Joanne B.
    Sneller, Michael A.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (06): : 8 - 13