Comparison of measured and computed portal dose for IMRT treatment

被引:13
作者
Cilla, Savino [1 ]
Viola, Pietro
Azario, Luigi
Grimaldi, Luca
Craus, Maurizio
D'Onofrio, Guido
Fidanzio, Andrea
Deodato, Francesco
Macchia, Gabriella
Digesu, Cinzia
Morganti, Alessio G.
Piermattei, Angelo
机构
[1] Univ Cattolica S Cuore, Centro Ric Formazione Alta Tecnol Sci Biomed, UO Fis Sanitaria, Campobasso, Italy
[2] Univ Cattolica S Cuore, Centro Ric Formazione Alta Tecnol Sci Biomed, UO Radioterapia, Campobasso, Italy
[3] Univ Cattolica S Coure, Ist Fis, Rome, Italy
关键词
portal dose; IMRT; radiotherapy;
D O I
10.1120/jacmp.v7i3.2281
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A new 2D array Seven 29 (TM) model (PTW, Freiburg), equipped with 729 vented plane-parallel ion chambers, projected for pretreatment verification of radiotherapy plans, was used as a detector for the transmitted or portal dose measurements below a Rando phantom. The dosimetric qualities of the 2D array make it attractive for measuring transmitted dose maps from step-and-shoot intensity-modulated radiotherapy (IMRT). It is well known that for step-and-shoot IMRT beams that use a small number of monitor units (MUs) per sequence, the early and recent electronic portal imaging devices (EPIDs) present a different response at X-ray start-up that affects the accuracy of the measured transmitted dose. The comparison of portal doses measured to those calculated by a commercial treatment-planning system (TPS) can verify correct dose delivery during treatment. This direct validation was tested by irradiating a simulated head tumor in a Rando anthropomorphic phantom by step-and-shoot IMRT beams. The absolute transmitted doses on a plane orthogonal to the beam central axis below the phantom were measured by the 2D array calibrated in terms of dose to water and compared with the computed portal dose extracted by custom software. In a previous paper, the comparison between the IMRT portal doses, computed by a commercial TPS and measured by a linear array that supplied a 1 mm spatial dose resolution, was carried out. The gamma-index analysis supplied an agreement of more than 95% of the dose point with acceptance criteria, in terms of dose difference,. Delta D-max, and distance agreement, Delta d(max), equal to 4% and 4 mm, respectively. In this paper, we verify the possible use of the PTW 2D array for measurements of the transmitted doses during several fractions of head and neck tumor radiotherapy. There are two advantages in the use of this 2D array as a portal dose device for the IMRT quality assurance program: first is the ability to perform absolute dose comparisons for hundreds of measurement positions to verify the correct dose delivery in several fractions of the therapy; second is the efficiency in time to detect these kinds of dose distributions within the field of view area of the CT scanner.
引用
收藏
页码:65 / 79
页数:15
相关论文
共 25 条
[1]   Dosimetric characterization of a large area pixel-segmented ionization chamber [J].
Amerio, S ;
Boriano, A ;
Bourhaleb, F ;
Cirio, R ;
Donetti, M ;
Fidanzio, A ;
Garelli, E ;
Giordanengo, S ;
Madon, E ;
Marchetto, F ;
Nastasi, U ;
Peroni, C ;
Piermattei, A ;
Freire, CJS ;
Sardo, A ;
Trevisiol, E .
MEDICAL PHYSICS, 2004, 31 (02) :414-420
[2]   First clinical tests using a liquid-filled electronic portal imaging device and a convolution model for the verification of the midplane dose [J].
Boellaard, R ;
van Herk, M ;
Uiterwaal, H ;
Mijnheer, B .
RADIOTHERAPY AND ONCOLOGY, 1998, 47 (03) :303-312
[3]  
Chang J, 2003, J Appl Clin Med Phys, V4, P287, DOI 10.1120/1.1615071
[4]   Correction for dose-response variations in a scanning liquid ion chamber EPID as a function of linac gantry angle [J].
Chin, PW ;
Lewis, DG ;
Spezi, E .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (08) :N93-N103
[5]  
CIRIO R, 2004, RADIOTHER ONCOL S1, V73, P256
[6]   Verification of segmented beam delivery using a commercial electronic portal imaging device [J].
Curtin-Savard, AJ ;
Podgorsak, EB .
MEDICAL PHYSICS, 1999, 26 (05) :737-742
[7]   Quality assurance of the dose delivered by small radiation segments [J].
Hansen, VN ;
Evans, PM ;
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brugmans, MJP ;
Wittkamper, FW ;
Mijnheer, BJ ;
Brown, K .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) :2665-2675
[8]  
IAEA, 2001, 398 IAEA TRS
[9]   A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery [J].
Jursinic, PA ;
Nelms, BE .
MEDICAL PHYSICS, 2003, 30 (05) :870-879
[10]   Evaluation of a 2D diode array for IMRT quality assurance [J].
Létourneau, D ;
Gulam, M ;
Yan, D ;
Oldham, M ;
Wong, JW .
RADIOTHERAPY AND ONCOLOGY, 2004, 70 (02) :199-206