Amperometric biosensor based on immobilized acetylcholinesterase on gold nanoparticles and silk fibroin modified platinum electrode for detection of methyl paraoxon, carbofuran and phoxim

被引:70
作者
Yin, Huanshun [1 ]
Ai, Shiyun [1 ]
Xu, Jing [1 ]
Shi, Weijie [1 ]
Zhu, Lusheng [2 ]
机构
[1] Shandong Agr Univ, Coll Chem & Mat Sci, Tai An 271018, Shandong, Peoples R China
[2] Shandong Agr Univ, Coll Resources & Environm, Tai An 271018, Shandong, Peoples R China
基金
中国国家自然科学基金;
关键词
Acetylcholinesterase; Gold nanoparticle; Biosensor; Methyl paraoxon; Carbofuran; Phoxim; PESTICIDE SENSITIVITY; LIQUID-CHROMATOGRAPHY; ENZYME ELECTRODE; CARBARYL; THIOCHOLINE; INHIBITION; SENSORS;
D O I
10.1016/j.jelechem.2009.09.025
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new sensitive and stable amperometric biosensor for the detection of methyl paraoxon, carbofuran and phoxim had been developed based on immobilization of acetylcholinasterase (AChE) on gold nanoparticles (AuNPs) and silk fibroin (SF) modified platinum electrode. The SF provided a biocompatible microenvironment around the enzyme molecule to stabilize its biological activity and effectively prevented it from leaking out of platinum electrode surface. In the presence of acetylthiocholine chloride (ATCl) as a substrate, AuNPs promoted electron transfer reaction at a lower potential and catalyzed the electrochemical oxidation of thiocholine (TCh), thus increasing detection sensitivity. The influences of pH, enzyme loading, temperature and inhibition time on the response of the fabricated biosensor were investigated. Under optimum conditions, the inhibition percentages of methyl paraoxon, carbofuran and phoxim were proportional to their concentrations in the range of 6 x 10(-11)-5 x 10(-8) M. 2 x 10(-10)-1 X 10(-7) M and 5 x 10(-9)-2 x 10(-7) M, respectively. The detection limits were found to be 2 x 10(-11) M for methyl paraoxon, 1x 10(-10) M for carbofuran and 2 x 10(-9) M for phoxim. Moreover. the fabricated biosensor had good reproducibility and acceptable stability. The biosensor is a promising new tool for pesticide analysis. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:21 / 27
页数:7
相关论文
共 42 条
[1]   Enzyme inhibition-based biosensors for food safety and environmental monitoring [J].
Amine, A ;
Mohammadi, H ;
Bourais, I ;
Palleschi, G .
BIOSENSORS & BIOELECTRONICS, 2006, 21 (08) :1405-1423
[2]   Detection of carbamic and organophosphorous pesticides in water samples using a cholinesterase biosensor based on Prussian Blue-modified screen-printed electrode [J].
Arduini, Fabiana ;
Ricci, Francesco ;
Tuta, Catalin S. ;
Moscone, Danila ;
Amine, Aziz ;
Palleschi, Giuseppe .
ANALYTICA CHIMICA ACTA, 2006, 580 (02) :155-162
[3]  
Aspelin L., 1994, PESTICIDES IND SALES
[4]   Biosensors based on highly sensitive acetylcholinesterases for enhanced carbamate insecticides detection [J].
Bucur, B ;
Fournier, D ;
Danet, A ;
Marty, JL .
ANALYTICA CHIMICA ACTA, 2006, 562 (01) :115-121
[5]   MICROBIOLOGICAL AND BIOTECHNOLOGICAL ASPECTS OF METABOLISM OF CARBAMATES AND ORGANOPHOSPHATES [J].
CHAPALAMADUGU, S ;
CHAUDHRY, GR .
CRITICAL REVIEWS IN BIOTECHNOLOGY, 1992, 12 (5-6) :357-389
[6]   Preparation and characterization of noble metal nanocolloids by silk fibroin in situ reduction [J].
Chen, WX ;
Wu, W ;
Chen, HX ;
Shen, ZQ .
SCIENCE IN CHINA SERIES B-CHEMISTRY, 2003, 46 (04) :330-338
[7]   SOLID-PHASE PRECONCENTRATION-FOURIER TRANSFORM-INFRARED SPECTROMETRIC DETERMINATION OF CARBARYL AND 1-NAPHTHOL [J].
DAGHBOUCHE, Y ;
GARRIGUES, S ;
DELAGUARDIA, M .
ANALYTICA CHIMICA ACTA, 1995, 314 (03) :203-212
[8]   IMMOBILIZATION OF GLUCOSE-OXIDASE WITH BOMBYX-MORI SILK FIBROIN BY ONLY STRETCHING TREATMENT AND ITS APPLICATION TO GLUCOSE SENSOR [J].
DEMURA, M ;
ASAKURA, T .
BIOTECHNOLOGY AND BIOENGINEERING, 1989, 33 (05) :598-603
[9]   Application of chemisorption/desorption process of thiocholine for pesticide detection based on acetylcholinesterase biosensor [J].
Du, Dan ;
Ding, Jiawang ;
Tao, Yuan ;
Chen, Xi .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 134 (02) :908-912
[10]   Electrochemical pesticide sensitivity test using acetylcholinesterase biosensor based on colloidal gold nanoparticle modified sol-gel interface [J].
Du, Dan ;
Chen, Shizhen ;
Cai, Jie ;
Zhang, Aidong .
TALANTA, 2008, 74 (04) :766-772