Noninvasive monitoring of cardiac output during exercise by inductance cardiography

被引:10
作者
Kaplan, V [1 ]
Bucklar, GB [1 ]
Bloch, KE [1 ]
机构
[1] Univ Zurich Hosp, Dept Internal Med, Div Pulm, CH-8091 Zurich, Switzerland
关键词
stroke volume; physiologic monitoring; inductive plethysmography; thoracocardiography; exercise testing; bicycle exercise; hemodynamic monitoring;
D O I
10.1249/01.MSS.0000064997.58069.A6
中图分类号
G8 [体育];
学科分类号
04 ; 0403 ;
摘要
Introduction/Purpose: Inductance cardiography is a noninvasive technique that monitors changes in cardiac output from an inductance plethysmographic transducer encircling the chest at the level of the heart. The method has been previously validated in supine patients at rest by comparisons to thermodilution. Our purpose was to investigate whether the technique can be employed during bicycle exercise. Methods: We simultaneously measured cardiac output by inductance cardiography and by two gas exchange methods based on the Fick principle during upright cycle ergometry in healthy volunteers. Results: In I I subjects, comparisons of changes in cardiac output by inductance cardiography agreed well with values measured by carbon dioxide rebreathing during a steady-state exercise protocol. In 12 subjects, cardiac output changes measured by inductance cardiography and an oxygen uptake method agreed closely during a progressive ramp exercise protocol to exhaustion. The bias (mean difference to reference methods) and limits of agreement (+/-2 SD of bias) for estimation of relative changes in cardiac output by inductance cardiography were 1% +/- 21% in 67 comparisons to the carbon dioxide rebreathing technique, and 0% +/- 22% in 98 comparisons to the oxygen uptake method. Conclusion: In healthy subjects, inductance cardiography accurately and unobtrusively estimates changes in cardiac output during bicycle exercise in comparison to gas exchange methods.
引用
收藏
页码:747 / 752
页数:6
相关论文
共 20 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Noninvasive monitoring of cardiac output in human neonates and juvenile piglets by inductance cardiography (thoracocardiography) [J].
Bloch, KE ;
Adams, JA ;
Bassuk, JJ ;
Mangino, MJ ;
Sackner, MA .
JOURNAL OF CRITICAL CARE, 2002, 17 (04) :259-266
[3]   Noninvasive monitoring of cardiac output in critically ill patients with thoracocardiography [J].
Bloch, KE ;
Baumann, PC ;
Stocker, R ;
Russi, EW .
AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1997, 155 (01) :26-31
[4]   Thoracocardiography: Noninvasive monitoring of left ventricular stroke volume [J].
Bloch, KE ;
Jugoon, S ;
de Socarraz, H ;
Manning, K ;
Sackner, MA .
JOURNAL OF CRITICAL CARE, 1998, 13 (03) :146-157
[5]   Inductance cardiography (thoracocardiography): A novel, noninvasive technique for monitoring left ventricular filling [J].
Bloch, KE ;
Jugoon, S ;
Sackner, MA .
JOURNAL OF CRITICAL CARE, 1999, 14 (04) :177-185
[6]  
BLOCH KE, 1998, PRINCIPLES PRACTICE, P915
[7]  
BUCKLAR GB, IN PRESS MED BIOL EN
[8]   CARDIAC OUTPUT IN MIDDLE-AGED PATIENTS DETERMINED WITH CO2 REBREATHING METHOD [J].
CLAUSEN, JP ;
LARSEN, OA ;
TRAPJENS.J .
JOURNAL OF APPLIED PHYSIOLOGY, 1970, 28 (03) :337-+
[9]   DETERMINATION OF MIXED VENOUS CO2 TENSIONS BY REBREATHING [J].
COLLIER, CR .
JOURNAL OF APPLIED PHYSIOLOGY, 1956, 9 (01) :25-29
[10]   COMPARISON OF CARDIAC-OUTPUT MEASUREMENT TECHNIQUES - THERMODILUTION, DOPPLER, CO2-REBREATHING AND THE DIRECT FICK METHOD [J].
ESPERSEN, K ;
JENSEN, EW ;
ROSENBORG, D ;
THOMSEN, JK ;
ELIASEN, K ;
OLSEN, NV ;
KANSTRUP, IL .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1995, 39 (02) :245-251