Sequential Activation of Ground Pads Reduces Skin Heating During Radiofrequency Tumor Ablation: In Vivo Porcine Results

被引:6
作者
Schutt, David J. [1 ]
Swindle, M. Michael [2 ]
Helke, Kristi L. [2 ]
Bastarrika, Gorka [3 ]
Schwarz, Florian [4 ]
Haemmerich, Dieter [1 ]
机构
[1] Med Univ S Carolina, Div Pediat Cardiol, Charleston, SC 29403 USA
[2] Med Univ S Carolina, Dept Comparat Med, Charleston, SC 29403 USA
[3] Univ Navarra Clin, Dept Radiol, Pamplona 31008, Spain
[4] Med Univ S Carolina, Dept Radiol, Charleston, SC 29403 USA
基金
美国国家卫生研究院;
关键词
RF ablation; skin burns; tumor ablation; HIGH-POWER GENERATOR; LIVER-TUMORS; THERMAL ABLATION; BURN DEPTH; EX-VIVO; COMPLICATIONS; ELECTRODES; CARCINOMA; MODEL; EFFICACY;
D O I
10.1109/TBME.2009.2033385
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Skin burns below ground pads during monopolar RF ablation are increasingly prevalent, thereby hindering the development of higher power RF generators capable of creating larger tumor ablation zones in combination with multiple or new applicators. Our goal was to evaluate reduction in skin temperatures via additional ground pads in an in vivo porcine model. Three ground pads placed on the animal's abdomen were activated either simultaneously or sequentially, where activation timing was adjusted to equilibrate skin temperature below each pad. Thirteen RF ablations (n = 4 simultaneous at 300 W, n = 5 sequential at 300 W, and n = 4 sequential at 375 W) were performed for 12 min via two internally cooled cluster electrodes placed in the gluteus maximus of domestic swine. Temperature rise at each pad and burn degree as determined via histology were compared. Ablation zone size was determined via T2-weighted MRI. Maximum temperature rise was significantly higher with simultaneous activation than with either of the sequential activation group (21.4 degrees C versus 8.1 degrees C or 9.6 degrees C, p < 0.01). Ablation zone diameters during simultaneous ( 300 W) and sequential activations ( 300 and 375 W) were and 6.9 +/- 0.3, 5.6 +/- 0.3, and 7.5 +/- 0.6 cm, respectively. Sequential activation of multiple ground pads results in significantly lower skin temperatures and less severe burns, as measured by histological examination.
引用
收藏
页码:746 / 753
页数:8
相关论文
共 29 条
[11]   Estimation of current density distribution under electrodes for external defibrillation [J].
Krasteva, Vessela Tz ;
Papazov, Sava P. .
BIOMEDICAL ENGINEERING ONLINE, 2002, 1 (1)
[12]   Treatment of focal liver tumors with percutaneous radio-frequency ablation: Complications encountered in a multicenter study [J].
Livraghi, T ;
Solbiati, L ;
Meloni, MF ;
Gazelle, GS ;
Halpern, EF ;
Goldberg, SN .
RADIOLOGY, 2003, 226 (02) :441-451
[13]   Prevention of dispersive pad skin burns during RFA by a simple method [J].
Machi, J .
SURGICAL LAPAROSCOPY ENDOSCOPY & PERCUTANEOUS TECHNIQUES, 2003, 13 (06) :372-373
[14]   Adrenal neoplasms: CT-guided radiofrequency ablation - Preliminary results [J].
Mayo-Smith, WW ;
Dupuy, DE .
RADIOLOGY, 2004, 231 (01) :225-230
[15]   Complications of radiofrequency coagulation of liver tumours [J].
Mulier, S ;
Mulier, P ;
Ni, Y ;
Miao, Y ;
Dupas, B ;
Marchal, G ;
De Wever, I ;
Michel, L .
BRITISH JOURNAL OF SURGERY, 2002, 89 (10) :1206-1222
[16]  
Neeman Ziv, 2002, Tech Vasc Interv Radiol, V5, P156, DOI 10.1053/tvir.2002.36419
[17]   The progression of burn depth in experimental burns:: a histological and methodological study [J].
Papp, A ;
Kiraly, K ;
Härmä, M ;
Lahtinen, T ;
Uusaro, A ;
Alhava, E .
BURNS, 2004, 30 (07) :684-690
[18]   Complications of radiofrequency ablation in hepatocellular carcinoma [J].
Rhim, H .
ABDOMINAL IMAGING, 2005, 30 (04) :409-418
[19]   Osteoid osteoma: Percutaneous treatment with radiofrequency energy [J].
Rosenthal, DI ;
Hornicek, FJ ;
Torriani, M ;
Gebhardt, MC ;
Mankin, HJ .
RADIOLOGY, 2003, 229 (01) :171-175
[20]   Sequential activation of a segmented ground pad reduces skin heating during radiofrequency tumor ablation: Optimization via computational models [J].
Schutt, David J. ;
Haemmerich, Dieter .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (07) :1881-1889