Performance Measurements of a SSPM-LYSO-SSPM Detector Module For Small Animal Positron Emission Tomography

被引:12
作者
Dokhale, Purushottam [1 ]
Stapels, Christopher [1 ]
Christian, James [1 ]
Yang, Yongfeng [2 ]
Cherry, Simon [2 ]
Moses, William [3 ]
Shah, Kanai [1 ]
机构
[1] Radiat Monitoring Devices Inc, Watertown, MA 02472 USA
[2] Univ Calif Davis, Davis, CA 95616 USA
[3] Lawrence Berkeley Natl Lab, Berkeley, CA 94720 USA
来源
2009 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5 | 2009年
关键词
RESOLUTION; DEPTH;
D O I
10.1109/NSSMIC.2009.5401653
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
We present the performance of a compact PET detector module with a depth-of-interaction (DO!) capability based on a LVSO scintillator array coupled at both ends to CMOS solid state photomultipliers (SSPM). In this paper we present flood imaging, energy resolution, coincidence timing and DOI resolution results for a prototype PET detector module consisting of a 4 x 4 block of LYSO scintillators, each crystal measuring 1.43 x 1.43 x 20 mm(3) coupled to 4 x 4 array of SSPMs, each SSPM element is 1.5x1.5 mm(2). All 16 elements in the flood image recorded with Na-22 gamma-ray source at all depths were clearly visible and well separated from each other. The measured FWHM energy resolution with 511 keV gamma rays for all crystals in the array ranged between 13% - 17% (FWHM). The timing resolution measured for the complete detector module in coincidence with LYSO-PMT detector was 2.3 ns (FWHM) and for the individual crystals ranged between 1.1ns - 1.3ns (FWHM). The DOI resolution measured for all crystals in the array was ranged between 2.3mm - 2.9mm.
引用
收藏
页码:2809 / +
页数:2
相关论文
共 13 条
[1]  
BUZHAN P, 2003, NUC I METH A, V504
[2]  
DOKHALE P, 2007, IEEE
[3]   Performance measurements of a depth-encoding PET detector module based on position-sensitive avalanche photodiode read-out [J].
Dokhale, PA ;
Silverman, RW ;
Shah, KS ;
Grazioso, R ;
Farrell, R ;
Glodo, J ;
McClish, MA ;
Entine, G ;
Tran, VH ;
Cherry, SR .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (18) :4293-4304
[4]  
DOLGOSHEIN B, 2005, LARGE AREA SILICON P
[5]   Energy, Timing and Position Resolution Studies With 16-Pixel Silicon Photomultiplier Matrices for Small Animal PET [J].
Llosa, Gabriela ;
Belcari, Nicola ;
Bisogni, M. Giuseppina ;
Collazuol, Gianmaria ;
Marcatili, Sara ;
Barrillon, Pierre ;
de la Taille, Christophe ;
Bondil-Blin, Sylvie ;
Dinu, Nicoleta ;
Melchiorri, Mirko ;
Tarolli, Alessandro ;
Piemonte, Claudio ;
Del Guerra, Alberto .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2009, 56 (05) :2586-2593
[6]  
MOSES W W, 1991, Journal of Nuclear Medicine, V32, P995
[7]  
OLCOTT PD, 2009, IEEE NSS MIC C ORL U
[8]   A test of silicon photomultipliers as readout for PET [J].
Otte, AN ;
Barral, J ;
Dolgoshein, B ;
Hose, J ;
Klemin, S ;
Lorenz, E ;
Mirzoyan, R ;
Popova, E ;
Teshima, M .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2005, 545 (03) :705-715
[9]  
PESTOTNIK, 2008, IEEE, P3123
[10]  
SCHAART, 2008, IEEE, P3581