A Pilot Evaluation of a 4-Dimensional Cone-Beam Computed Tomographic Scheme Based on Simultaneous Motion Estimation and Image Reconstruction

被引:17
作者
Dang, Jun [1 ]
Gu, Xuejun [1 ]
Pan, Tinsu [2 ]
Wang, Jing [1 ]
机构
[1] Univ Texas SW Med Ctr Dallas, Dept Radiat Oncol, Dallas, TX 75390 USA
[2] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2015年 / 91卷 / 02期
关键词
C-ARM CT; RESPIRATORY MOTION; DYNAMIC CT; CARDIAC CT; PHASE; MODEL; REGISTRATION; RADIOTHERAPY; ALGORITHMS;
D O I
10.1016/j.ijrobp.2014.10.029
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the performance of a 4-dimensional (4-D) cone-beam computed tomographic (CBCT) reconstruction scheme based on simultaneous motion estimation and image reconstruction (SMEIR) through patient studies. Methods and Materials: The SMEIR algorithm contains 2 alternating steps: (1) motion-compensated CBCT reconstruction using projections from all phases to reconstruct a reference phase 4D-CBCT by explicitly considering the motion models between each different phase and (2) estimation of motion models directly from projections by matching the measured projections to the forward projection of the deformed reference phase 4D-CBCT. Four lung cancer patients were scanned for 4 to 6 minutes to obtain approximately 2000 projections for each patient. To evaluate the performance of the SMEIR algorithm on a conventional 1-minute CBCT scan, the number of projections at each phase was reduced by a factor of 5, 8, or 10 for each patient. Then, 4D-CBCTs were reconstructed from the down-sampled projections using Feldkamp-Davis-Kress, total variation (TV) minimization, prior image constrained compressive sensing (PICCS), and SMEIR. Using the 4D-CBCT reconstructed from the fully sampled projections as a reference, the relative error (RE) of reconstructed images, root mean square error (RMSE), and maximum error (MaxE) of estimated tumor positions were analyzed to quantify the performance of the SMEIR algorithm. Results: The SMEIR algorithm can achieve results consistent with the reference 4D-CBCT reconstructed with many more projections per phase. With an average of 30 to 40 projections per phase, the MaxE in tumor position detection is less than 1 mm in SMEIR for all 4 patients. Conclusion: The results from a limited number of patients show that SMEIR is a promising tool for high-quality 4D-CBCT reconstruction and tumor motion modeling. (C) 2015 Elsevier Inc.
引用
收藏
页码:410 / 418
页数:9
相关论文
共 36 条
[1]   Phase and amplitude binning for 4D-CT imaging [J].
Abdelnour, A. F. ;
Nehmeh, S. A. ;
Pan, T. ;
Humm, J. L. ;
Vernon, P. ;
Schoder, H. ;
Rosenzweig, K. E. ;
Mageras, G. S. ;
Yorke, E. ;
Larson, S. M. ;
Erdi, Y. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (12) :3515-3529
[2]   An investigation of 4D cone-beam CT algorithms for slowly rotating scanners [J].
Bergner, Frank ;
Berkus, Timo ;
Oelhafen, Markus ;
Kunz, Patrik ;
Pan, Tinsu ;
Grimmer, Rainer ;
Ritschl, Ludwig ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2010, 37 (09) :5044-5053
[3]   Autoadaptive phase-correlated (AAPC) reconstruction for 4D CBCT [J].
Bergner, Frank ;
Berkus, Timo ;
Oelhafen, Markus ;
Kunz, Patrik ;
Pan, Tinsu ;
Kachelriess, Marc .
MEDICAL PHYSICS, 2009, 36 (12) :5695-5706
[4]   Observation of interfractional variations in lung tumor position using respiratory gated and ungated megavoltage cone-beam computed tomography [J].
Chang, Jenghwa ;
Mageras, Gig S. ;
Yorke, Ellen ;
De Arruda, Fernando ;
Sillanpaa, Jussi ;
Rosenzweig, Kenneth E. ;
Hertanto, Agung ;
Pham, Hai ;
Seppi, Edward ;
Pevsner, Alex ;
Ling, C. Clifton ;
Amols, Howard .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05) :1548-1558
[5]   Prior image constrained compressed sensing (PICCS): A method to accurately reconstruct dynamic CT images from highly undersampled projection data sets [J].
Chen, Guang-Hong ;
Tang, Jie ;
Leng, Shuai .
MEDICAL PHYSICS, 2008, 35 (02) :660-663
[6]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[7]   4D cone beam CT via spatiotemporal tensor framelet [J].
Gao, Hao ;
Li, Ruijiang ;
Lin, Yuting ;
Xing, Lei .
MEDICAL PHYSICS, 2012, 39 (11) :6943-6946
[8]   Motion estimation for cardiac emission tomography by optical flow methods [J].
Gilland, D. R. ;
Mair, B. A. ;
Parker, J. G. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (11) :2991-3006
[9]   Implementation and evaluation of various demons deformable image registration algorithms on a GPU [J].
Gu, Xuejun ;
Pan, Hubert ;
Liang, Yun ;
Castillo, Richard ;
Yang, Deshan ;
Choi, Dongju ;
Castillo, Edward ;
Majumdar, Amitava ;
Guerrero, Thomas ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (01) :207-219
[10]  
Han GP, 1999, IEEE NUCL SCI CONF R, P1515, DOI 10.1109/NSSMIC.1999.842846