A sliding-distance-coupled finite element formulation for polyethylene wear in total hip arthroplasty

被引:133
作者
Maxian, TA [1 ]
Brown, TD [1 ]
Pedersen, DR [1 ]
Callaghan, JJ [1 ]
机构
[1] UNIV IOWA, DEPT BIOMED ENGN, IOWA CITY, IA 52242 USA
关键词
wear; finite element; contact stress; polyethylene; sliding distance;
D O I
10.1016/0021-9290(95)00125-5
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A three-dimensional, nonlinear contact finite element (FE) model of total hip replacement, linked to a sliding-distance-coupled wear algorithm, was used to study polyethylene wear rates for three different Femoral head sizes. Hip resultant loads from a validated gait analysis model were used in the FE model to determine contact stress distributions on the polyethylene bearing surface, for 16 discrete instants of stance phase. Sliding distances of points on the femoral head surface were obtained from the corresponding flexion/extension kinematics. Wear rates were determined by a custom-written computer program that used a relationship that coupled contact stress, sliding distance, and a pin-on-disk determined wear coefficient. The wear rates computed by this formulation were well within clinically observed ranges for each component size.
引用
收藏
页码:687 / 692
页数:6
相关论文
共 14 条
[1]  
AMSTUTZ HC, 1991, CLIN ORTHOP RELAT R, V276, P7
[2]  
[Anonymous], T 17 ANN M SOC BIOM
[3]   CONTACT AND RUBBING OF FLAT SURFACES [J].
ARCHARD, JF .
JOURNAL OF APPLIED PHYSICS, 1953, 24 (08) :981-988
[4]   LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .3. THE MEASUREMENT OF INTERNAL VOLUME CHANGES IN EXPLANTED CHARNLEY SOCKETS AFTER 2-16 YEARS INVIVO AND THE DETERMINATION OF WEAR FACTORS [J].
ATKINSON, JR ;
DOWSON, D ;
ISAAC, JH ;
WROBLEWSKI, BM .
WEAR, 1985, 104 (03) :225-244
[5]  
BARTEL DL, 1991, BRIS MYER Z, P61
[6]   THE EFFECT OF CONFORMITY AND PLASTIC THICKNESS ON CONTACT STRESSES IN METAL-BACKED PLASTIC IMPLANTS [J].
BARTEL, DL ;
BURSTEIN, AH ;
TODA, MD ;
EDWARDS, DL .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1985, 107 (03) :193-199
[7]  
Brand R A, 1994, J Arthroplasty, V9, P45, DOI 10.1016/0883-5403(94)90136-8
[8]  
DALSTRA M, 1993, T 39 ORTH RES SOC, P442
[9]  
Fisher J, 1992, J ENG MED, V205
[10]   EFFECT OF FEMORAL-HEAD SIZE ON WEAR OF THE POLYETHYLENE ACETABULAR COMPONENT [J].
LIVERMORE, J ;
ILSTRUP, D ;
MORREY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :518-528