Proof of concept non-invasive estimation of peripheral venous oxygen saturation

被引:19
作者
Khan, Musabbir [1 ]
Pretty, Chris G. [1 ]
Amies, Alexander C. [1 ]
Balmer, Joel [1 ]
Banna, Houda E. [2 ]
Shaw, Geoffrey M. [2 ]
Chase, J. Geoffrey [1 ]
机构
[1] Univ Canterbury, Dept Mech Engn, Ctr Bioengn, Private Bag 4800, Christchurch 8140, New Zealand
[2] Christchurch Hosp, Dept Intens Care Unit, Private Bag 4710, Christchurch 8140, New Zealand
来源
BIOMEDICAL ENGINEERING ONLINE | 2017年 / 16卷
关键词
Non-invasive; Pulse oximeter; Photoplethysmograph; SvO(2) estimation; Artificial modulation; NEAR-INFRARED SPECTROSCOPY; PHOTOPLETHYSMOGRAPH WAVE-FORM; PULSE OXIMETRY; SEVERE SEPSIS; INTENSIVE-CARE; BLOOD-FLOW; OCCLUSION; SHOCK; TRANSPORT; PERFUSION;
D O I
10.1186/s12938-017-0351-x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Pulse oximeters continuously monitor arterial oxygen saturation. Continuous monitoring of venous oxygen saturation (SvO(2)) would enable real-time assessment of tissue oxygen extraction (O2E) and perfusion changes leading to improved diagnosis of clinical conditions, such as sepsis. Methods: This study presents the proof of concept of a novel pulse oximeter method that utilises the compliance difference between arteries and veins to induce artificial respiration-like modulations to the peripheral vasculature. These modulations make the venous blood pulsatile, which are then detected by a pulse oximeter sensor. The resulting photoplethysmograph (PPG) signals from the pulse oximeter are processed and analysed to develop a calibration model to estimate regional venous oxygen saturation (SpvO(2)), in parallel to arterial oxygen saturation estimation (SpaO(2)). A clinical study with healthy adult volunteers (n = 8) was conducted to assess peripheral SvO(2) using this pulse oximeter method. A range of physiologically realistic SvO(2) values were induced using arm lift and vascular occlusion tests. Gold standard, arterial and venous blood gas measurements were used as reference measurements. Modulation ratios related to arterial and venous systems were determined using a frequency domain analysis of the PPG signals. Results: A strong, linear correlation (r(2) = 0.95) was found between estimated venous modulation ratio (R-Ven) and measured SvO(2), providing a calibration curve relating measured R-Ven to venous oxygen saturation. There is a significant difference in gradient between the SpvO(2) estimation model (SpvO(2) = 111 -40.6*R) and the empirical SpaO(2) estimation model (SpaO(2) = 110 -25*R), which yields the expected arterial-venous differences. Median venous and arterial oxygen saturation accuracies of paired measurements between pulse oximeter estimated and gold standard measurements were 0.29 and 0.65%, respectively, showing good accuracy of the pulse oximeter system. Conclusions: The main outcome of this study is the proof of concept validation of a novel pulse oximeter sensor and calibration model to assess peripheral SvO(2), and thus O2E, using the method used in this study. Further validation, improvement, and application of this model can aid in clinical diagnosis of microcirculation failures due to alterations in oxygen extraction.
引用
收藏
页数:16
相关论文
共 41 条
[1]  
Abbott, 2013, I STAT 1 SYST MAN
[2]  
[Anonymous], 2012, MECH CIRCULATION
[3]   Pulse oximetry: Its invention, theory, and future [J].
Takuo Aoyagi .
Journal of Anesthesia, 2003, 17 (4) :259-266
[4]   MICROVASCULAR FUNCTION AND RHEOLOGICAL CHANGES IN HYPERDYNAMIC SEPSIS [J].
ASTIZ, ME ;
DEGENT, GE ;
LIN, RY ;
RACKOW, EC .
CRITICAL CARE MEDICINE, 1995, 23 (02) :265-271
[5]   Maximizing the Laboratory Setting for Testing Devices and Understanding Statistical Output in Pulse Oximetry [J].
Batchelder, Paul B. ;
Raley, Dena M. .
ANESTHESIA AND ANALGESIA, 2007, 105 :S85-S94
[6]  
Casciani JR, 2008, US Patent No, Patent No. [7,415,298, 7415298]
[7]   Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine [J].
Cecconi, Maurizio ;
De Backer, Daniel ;
Antonelli, Massimo ;
Beale, Richard ;
Bakker, Jan ;
Hofer, Christoph ;
Jaeschke, Roman ;
Mebazaa, Alexandre ;
Pinsky, Michael R. ;
Teboul, Jean Louis ;
Vincent, Jean Louis ;
Rhodes, Andrew .
INTENSIVE CARE MEDICINE, 2014, 40 (12) :1795-1815
[8]  
Daly SM, 2013, J BIOPHOTONICS, V6, P217, DOI 10.1002/jbio.201200071
[9]   Increased oxygen extraction fraction is associated with prior ischemic events in patients with carotid occlusion [J].
Derdeyn, CP ;
Yundt, KD ;
Videen, TO ;
Carpenter, DA ;
Grubb, RL ;
Powers, WJ .
STROKE, 1998, 29 (04) :754-758
[10]   Non-invasive measurement of peripheral venous oxygen saturation using a new venous oximetry method: evaluation during bypass in heart surgery [J].
Echiadis, A. S. ;
Crabtree, V. P. ;
Bence, J. ;
Hadjinikolaou, L. ;
Alexiou, C. ;
Spyt, T. J. ;
Hu, S. .
PHYSIOLOGICAL MEASUREMENT, 2007, 28 (08) :897-911