Use of the Moses Technology to Improve Holmium Laser Lithotripsy Outcomes: A Preclinical Study

被引:178
作者
Elhilali, Mostafa M. [1 ]
Badaan, Shadie [2 ]
Ibrahim, Ahmed [1 ,3 ]
Andonian, Sero [1 ]
机构
[1] McGill Univ, Ctr Hlth, Royal Victoria Hosp, Dept Urol, 1001 Blvd Decarie,D05-5327, Montreal, PQ H4A 3J1, Canada
[2] Rambam Hlth Care Campus, Dept Urol, Haifa, Israel
[3] Al Azhar Univ, Alhussain Univ Hosp, Dept Urol, Cairo, Egypt
来源
JOURNAL OF ENDOUROLOGY | 2017年 / 31卷 / 06期
关键词
Holmium-YAG laser; laser lithotripsy; prospective study; preclinical evaluation; FRAGMENTATION EFFICIENCY; URETERAL CALCULI;
D O I
10.1089/end.2017.0050
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective: To evaluate in vitro and in vivo effects of Moses technology in Holmium laser and to compare it with the Regular mode in terms of lithotripsy efficiency and laser-tissue interactions. Methods: The Lumenis (R) Pulse (TM) P120H holmium laser system together with Moses D/F/L fibers were used to compare the Regular mode with the Moses modes in stone retropulsion by using a high-speed camera, and stone ablation efficiency. In addition, a porcine ureteroscopy model was used to assess stone fragmentation and dusting as well as laser-tissue interaction with the ureteral wall. Results: After a laser pulse, in vitro stone displacement experiments showed a significant reduction in retropulsion when using the Moses mode. The stone movement was reduced by 50 times at 0.8 J and 10 Hz (p < 0.01). The pronounced reduction of retropulsion in the Moses mode was clearly observed during fragmentation setting (high energy) and dusting (low energy, high Hz). In addition, stone fragmentation tests showed that the Moses modes resulted in a significantly higher ablation volume when compared with the Regular mode (160% higher; p < 0.001). In vivo assessment also supported the reduction in retropulsion when treating stones in the porcine kidney. Histological analysis of the porcine ureter after direct lasing in the Moses mode suggested less damage than in the Regular mode. Conclusions: The Moses technology resulted in more efficient laser lithotripsy, in addition to significantly reduced stone retropulsion, and displayed a margin of safety that may result in a shorter procedural time and safer lithotripsy.
引用
收藏
页码:598 / 604
页数:7
相关论文
共 9 条
  • [1] Efficacy and safety of Ho:YAG Laser Lithotripsy for ureteroscopic removal of proximal and distal ureteral calculi
    Khoder, Wael Y.
    Bader, Markus
    Sroka, Ronald
    Stief, Christian
    Waidelich, Raphaela
    [J]. BMC UROLOGY, 2014, 14
  • [2] In vitro fragmentation efficiency of holmium: yttrium-aluminum-garnet (YAG) laser lithotripsy - a comprehensive study encompassing different frequencies, pulse energies, total power levels and laser fibre diameters
    Kronenberg, Peter
    Traxer, Olivier
    [J]. BJU INTERNATIONAL, 2014, 114 (02) : 261 - 267
  • [3] In vitro evaluation of nitinol urological retrieval coil and ureteral occlusion device: Retropulsion and holmium laser fragmentation efficiency
    Lee, Hak J.
    Box, Geoffrey N.
    Abraham, Jose Benito A.
    Deane, Leslie A.
    Elchico, Erick R.
    Eisner, Brian H.
    McDougall, Elspeth M.
    Clayman, Ralph V.
    [J]. JOURNAL OF UROLOGY, 2008, 180 (03) : 969 - 973
  • [4] CUA Guideline: Management of ureteral calculi
    Ordon, Michael
    Andonian, Sero
    Blew, Brian
    Schuler, Trevor
    Chew, Ben
    Pace, Kenneth T.
    [J]. CUAJ-CANADIAN UROLOGICAL ASSOCIATION JOURNAL, 2015, 9 (11-12): : E837 - E851
  • [5] Optimal Power Settings for Holmium:YAG Lithotripsy
    Sea, Jason
    Jonat, Lee M.
    Chew, Ben H.
    Qiu, Jinze
    Wang, Bingqing
    Hoopman, John
    Milner, Thomas
    Teichman, Joel M. H.
    [J]. JOURNAL OF UROLOGY, 2012, 187 (03) : 914 - 919
  • [6] Technical aspects of lasers in urology
    Teichmann, Heinrich-O.
    Herrmann, Thomas R.
    Bach, Thorsten
    [J]. WORLD JOURNAL OF UROLOGY, 2007, 25 (03) : 221 - 225
  • [7] EAU Guidelines on Interventional Treatment for Urolithiasis
    Tuerk, Christian
    Petrik, Ales
    Sarica, Kemal
    Seitz, Christian
    Skolarikos, Andreas
    Straub, Michael
    Knoll, Thomas
    [J]. EUROPEAN UROLOGY, 2016, 69 (03) : 475 - 482
  • [8] The evolution of lasers in urology
    Zarrabi, Amir
    Gross, Andreas J.
    [J]. THERAPEUTIC ADVANCES IN UROLOGY, 2011, 3 (02) : 81 - 89
  • [9] Water content contribution in calculus phantom ablation during Q-switched Tm:YAG laser lithotripsy
    Zhang, Jian J.
    Rajabhandharaks, Danop
    Xuan, Jason Rongwei
    Wang, Hui
    Chia, Ray W. J.
    Hasenberg, Tom
    Kang, Hyun Wook
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2015, 20 (12)