Comparison of Choriocapillaris Flow Measurements between Two Optical Coherence Tomography Angiography Devices

被引:14
|
作者
Lauermann, Jost L. [1 ]
Heiduschka, Peter [1 ]
Nelis, Pieter [1 ]
Treder, Maximilian [1 ]
Alnawaiseh, Maged [1 ]
Clemens, Christoph R. [1 ]
Eter, Nicole [1 ]
Alten, Florian [1 ]
机构
[1] Univ Munster, Med Ctr, Dept Ophthalmol, Domagkstr 15, DE-48149 Munster, Germany
关键词
Optical coherence tomography; Angiography; Choriocapillaris; Choroidal blood flow; Retinal imaging; AMPLITUDE-DECORRELATION ANGIOGRAPHY; OCT ANGIOGRAPHY; MORPHOMETRIC-ANALYSIS; MOTION CORRECTION; QUANTIFICATION; FLUORESCEIN; THICKNESS; DENSITY; EYES; RPE;
D O I
10.1159/000464355
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate choriocapillaris (CC) perfusion in healthy subjects using 2 different optical coherence tomography angiography (OCT-A) devices. Procedures: Macular OCT-A imaging (36 eyes of 36 subjects) was performed using Optovue AngioVue and Zeiss AngioPlex devices. CC decorrelation signal index was assessed, and CC data were analyzed regarding intra-device variability, inter-device correlation, age, signal strength, and fields of view. Results: The intra-device variability of CC measurements in the 3 x 3 mm(2) field was 5.3 and 2.6% (Angiovue and Angioplex, coefficients of variation; 6 x 6 mm(2) : 8.0 and 2.8%, respectively). Mean CC decorrelation signal index in 3 x 3 mm(2) was 104.3 +/- 6.7 (Angiovue) and 81.3 +/- 9.2 (Angioplex) (6 x 6 mm(2) : 95.6 +/- 8.1, 81.1 +/- 6.5) with high correlation between both devices (3 x 3 mm(2) : p = 0.0053; 6 x 6 mm(2) : p = 0.0139). CC decorrelation signal index in 3 x 3 mm(2) was significantly higher in subjects aged <= 58 years compared to subjects aged >= 59 years (Angiovue: 107.3 +/- 3.6, 101.3 +/- 7.7, p = 0.0156; Angioplex: 84.6 +/- 7.6, 78.0 +/- 9.5, p = 0.0371). Signal strength was 64.6 +/- 8.9 (Angiovue) and 9.5 +/- 0.8 (Angioplex). Conclusion: Both devices showed low intra-device variability and a high inter-device correlation. CC decorrelation signal index was negatively correlated with advancing age. (C) 2017 S. Karger AG, Basel
引用
收藏
页码:238 / 246
页数:9
相关论文
共 50 条
  • [41] Handheld spectrally encoded coherence tomography and reflectometry for motion-corrected ophthalmic optical coherence tomography and optical coherence tomography angiography
    Malone, Joseph D.
    El-Haddad, Mohamed Tarek
    Yerramreddy, Suhaas S.
    Oguz, Ipek
    Tao, Yuankai K.
    NEUROPHOTONICS, 2019, 6 (04)
  • [42] Topographic analysis of macular choriocapillaris flow deficits in diabetic retinopathy using swept–source optical coherence tomography angiography
    Isaac Gendelman
    A. Yasin Alibhai
    Eric M. Moult
    Emily S. Levine
    Phillip X. Braun
    Nihaal Mehta
    Yi Zhao
    Akihiro Ishibazawa
    Osama A. Sorour
    Caroline R. Baumal
    Andre J. Witkin
    Elias Reichel
    James G. Fujimoto
    Jay S. Duker
    Nadia K. Waheed
    International Journal of Retina and Vitreous, 6
  • [43] Visualizing choriocapillaris using swept-source optical coherence tomography angiography with various probe beam sizes
    Zhou, Kanheng
    Song, Shaozhen
    Zhang, Qinqin
    Chu, Zhongdi
    Huang, Zhihong
    Wang, Ruikang K.
    BIOMEDICAL OPTICS EXPRESS, 2019, 10 (06) : 2847 - 2860
  • [44] Optical Coherence Tomography-Angiography in Geographic Atrophy
    Mueller, Philipp L.
    Pfau, Maximilian
    Schmitz-Valckenberg, Steffen
    Fleckenstein, Monika
    Holz, Frank G.
    OPHTHALMOLOGICA, 2021, 244 (01) : 42 - 50
  • [45] CHOROIDAL NEOVASCULAR AREA AND VESSEL DENSITY COMPARISON BETWEEN TWO SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY ANGIOGRAPHY DEVICES
    Ohayon, Avi
    Sacconi, Riccardo
    Semoun, Oudy
    Corbelli, Eleonora
    Souied, Eric H.
    Querques, Giuseppe
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2020, 40 (03): : 521 - 528
  • [46] Optical Coherence Tomography and Optical Coherence Tomography Angiography Findings in Multiple Sclerosis Patients
    Balikci, Ayse
    Yener, Neslihan Parmak
    Seferoglu, Meral
    NEURO-OPHTHALMOLOGY, 2022, 46 (01) : 19 - 33
  • [47] Adaptive optics optical coherence tomography angiography for morphometric analysis of choriocapillaris [Invited]
    Kurokawa, Kazuhiro
    Liu, Zhuolin
    Miller, Donald T.
    BIOMEDICAL OPTICS EXPRESS, 2017, 8 (03): : 1803 - 1822
  • [48] Retinal vessel calibre measurements by optical coherence tomography angiography
    Falavarjani, Khalil Ghasemi
    Al-Sheikh, Mayss
    Darvizeh, Fatemeh
    Sadun, Alfredo A.
    Sadda, Srinivas R.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2017, 101 (07) : 989 - 992
  • [49] Optical Coherence Tomography Angiography of the Choriocapillaris in Subclinical Toxoplasma Chorioretinitis: A Case Report
    Alsatrawi, Ahmed B.
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (07)
  • [50] Comparison between optical coherence tomography angiography and fundus fluorescein angiography for early detection of diabetic retinopathy
    Badr, Amal S. S.
    El-Baha, Samir M.
    Souka, Ahmed A. R.
    Goma, Amir R.
    JOURNAL OF THE EGYPTIAN OPHTHALMOLOGICAL SOCIETY, 2022, 115 (01) : 29 - 34