The Effect of Different Closed Suction Catheter Designs and pMDI Adapters on Aerosol Delivery in Simulated Adult Mechanical Ventilation With and Without Exhaled Humidity

被引:9
|
作者
Williams, Jacqueline P. [1 ]
Ari, Arzu [2 ]
Shanmugam, Ramalingam [3 ]
Fink, James B. [4 ]
机构
[1] Emory Univ, Resp Therapy Dept, Midtown Hosp, Emory Healthcare, Atlanta, GA 30322 USA
[2] Texas State Univ, Dept Resp Care, 601 Univ Dr, San Marcos, TX 78666 USA
[3] Texas State Univ, Sch Hlth Adm, San Marcos, TX 78666 USA
[4] Aerogen Pharma, San Mateo, CA USA
关键词
aerosol; airway clearance; mechanical ventilation; closed suction catheters; metered dose inhalers; pMDI adapters; drug delivery; METERED-DOSE INHALER; BRONCHODILATOR DELIVERY; CRITICAL-CARE; LUNG MODEL; EFFICIENCY; DEVICES; HUMIDIFICATION; DEPOSITION; THERAPY; INVITRO;
D O I
10.4187/respcare.06085
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Closed suction catheters are commonly used with pressurized metered-dose inhalers (pMDIs) in ventilator-dependent patients receiving bronchodilator therapy. Although heated/humidified circuits result in a decrease in drug delivery, no studies have investigated the impact of a closed suction catheter with a pMDI on aerosol delivery during mechanical ventilation. We sought to quantify aerosol delivery with a variety of closed suction catheters and pMDI adapters in a simulated ventilator-dependent adult lung model with and without exhaled humidity. METHODS: A ventilator (tidal volume 450 mL, PEEP 5 cm H2O, breathing frequency 15 breaths/min, peak flow 60 L/min, and bias flow 2 L/min) was attached to 3 closed suction catheter designs (T-piece, double swivel elbow, and a multi-access port) with pMDI adapters (small- and large-unidirectional adapters, a bidirectional adapter, and a built-in port) connected to an endotracheal tube (ETT) with a collecting filter and a passive test lung. To simulate exhaled humidity, a heated humidifier was placed between the collection filter and the distal tip of the ETT, set at 36 degrees C. Experiments without exhaled humidity were performed with no heat and humidification. For each test, 4 puffs of albuterol sulfate (432 mu g) were administered. The drug was eluted from the collecting filter and analyzed via spectrophotometer (276 nm). Each experiment was run 3 times. RESULTS: Delivery efficiency of the bidirectional adapter was greatest compared to other adapters tested in this study (P = .01). Aerosol deposition obtained from the combination of the T-piece closed suction catheter and the bidirectional adapter was more than that from the multi-access port (P = .42) and double swivel closed suction catheter (P = .27). Drug delivery without exhaled humidity exceeded values with exhaled humidity by 20-90% using the small unidirectional, large unidirectional, and bidirectional adapters (P = .01, P = .02 and P = .02, respectively). CONCLUSIONS: Closed suction catheter, pMDI spacer/ adapters, and exhalation conditions have an impact on drug delivery during simulated adult mechanical ventilation. Aerosol delivery without exhaled humidity appears to report unrealistically high values in comparison to the more accurate values with simulated exhalation humidity.
引用
收藏
页码:1154 / 1161
页数:8
相关论文
共 4 条
  • [1] Quantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity
    Ari, Arzu
    Harwood, Robert
    Sheard, Meryl
    Alquaimi, Maher Mubarak
    Alhamad, Bshayer
    Fink, James B.
    RESPIRATORY CARE, 2016, 61 (05) : 600 - 606
  • [2] Position of different nebulizer types for aerosol delivery in an adult model of mechanical ventilation
    Hou, Haijia
    Xu, Dongyang
    Dai, Bing
    Zhao, Hongwen
    Wang, Wei
    Kang, Jian
    Tan, Wei
    FRONTIERS IN MEDICINE, 2022, 9
  • [3] The Effects of Inspiratory Flows, Inspiratory Pause, and Suction Catheter on Aerosol Drug Delivery with Vibrating Mesh Nebulizers During Mechanical Ventilation
    Lin, Hui-Ling
    Fink, James B.
    Li, Jie
    JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2024, 37 (03) : 125 - 131
  • [4] Influence of Nebulizer Type, Position, and Bias Flow on Aerosol Drug Delivery in Simulated Pediatric and Adult Lung Models During Mechanical Ventilation
    Ari, Arzu
    Atalay, Orcin Telli
    Harwood, Robert
    Sheard, Meryl M.
    Aljamhan, Essam A.
    Fink, James B.
    RESPIRATORY CARE, 2010, 55 (07) : 845 - 851