Contactless conductivity detection for capillary electrophoresis

被引:494
作者
Zemann, AJ [1 ]
Schnell, E [1 ]
Volgger, D [1 ]
Bonn, GK [1 ]
机构
[1] Univ Innsbruck, Inst Analyt Chem & Radiochem, A-6020 Innsbruck, Austria
基金
巴西圣保罗研究基金会;
关键词
D O I
10.1021/ac9707592
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A contactless capacitively coupled conductivity detector for capillary electrophoresis is introduced. The detector consists of two electrodes which are placed cylindrically around the outer polyimide coating of the fused-silica capillary with a detection gap of 2 mm. The electrodes form a cylindrical capacitor, and the electric conductivity of the solution in the gap between the electrodes is measured. A high audio or low ultrasonic frequency for coupling of the ac voltage is used in order to minimize the influence of reactance of the liquid. For an improved version of the detector, two syringe cannulas are used as the electrodes and the capillary is simply assembled into the tubing. This allows an easy placement of the detector on various positions along the capillary. The limit of detection of inorganic cations and anions is 200 ppb, as determined for sodium and chloride, respectively.
引用
收藏
页码:563 / 567