Liquid ionization chamber measurements of dose distributions in small 6 MV photon beams

被引:19
作者
Dasu, A
Lofroth, PO
Wickman, G
机构
[1] Umea Univ, Dept Radiat Phys, S-90185 Umea, Sweden
[2] Alexandru Ioan Cuza Univ, Dept Med Phys & Biophys, R-6600 Iasi 7, Romania
关键词
D O I
10.1088/0031-9155/43/1/002
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A new liquid ionization chamber (LIC) design optimized for high spatial resolution was used for measurements of dose distributions in radiation fields intended for stereotactic radiosurgery (SRS). This work was mainly focused on the properties of this detector in radiation fields from linear accelerators for clinical radiotherapy (pulsed radiation with dose rates from approximately 0.5 to 5 Gy min(-1) and beam diameters down to 8 mm). The narrow beams used in stereotactic radiosurgery require detectors with small sizes in order to provide a good spatial resolution. The LIC is investigated to see whether it can be used as a detector for dose measurements in beams currently used for stereotactic radiosurgery. Its properties are compared with those of silicon diodes. The comparisons include output factor (OF), depth dose and profile measurements in 6 MV photon fields of different sizes. For OF measurements, an NACP air ionization chamber was also used in the comparison. The dependence of the response on the detector orientation in the photon beam is also investigated for the diodes and the LIC. The results suggest that LICs can provide better properties than diodes for measuring dose distributions in narrow photon beams.
引用
收藏
页码:21 / 36
页数:16
相关论文
共 30 条
  • [1] [Anonymous], 1983, MED PHYS, V10, P741
  • [2] ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY
    BEDDAR, AS
    MASON, DJ
    OBRIEN, PF
    [J]. MEDICAL PHYSICS, 1994, 21 (07) : 1075 - 1079
  • [3] Burgemeister E. A., 1984, Radiation Protection Dosimetry, V6, P145
  • [4] Use of a micro-ionization chamber and an anthropomorphic head phantom in a quality assurance program for stereotactic radiosurgery
    Duggan, DM
    Coffey, CW
    [J]. MEDICAL PHYSICS, 1996, 23 (04) : 513 - 516
  • [5] DUTREIX A, 1997, PHYSICS CLIN RADIOTH, P104
  • [6] FOWLER JF, 1966, RAD DOSIMETRY, P291
  • [7] LINEARITY WITH DOSE-RATE OF LOW-RESISTIVITY P-TYPE SILICON SEMICONDUCTOR-DETECTORS
    GRUSELL, E
    RIKNER, G
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06) : 785 - 792
  • [8] A comparison of dosimetry techniques in stereotactic radiosurgery
    Heydarian, M
    Hoban, PW
    Beddoe, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) : 93 - 110
  • [9] DOSE-RATE DEPENDENCE OF A PTW DIAMOND DETECTOR IN THE DOSIMETRY OF A 6 MV PHOTON-BEAM
    HOBAN, PW
    HEYDARIAN, M
    BECKHAM, WA
    BEDDOE, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (08) : 1219 - 1229
  • [10] ICRU, 1984, 37 ICRU