Thermospray nebulization as sample introduction for inductively coupled plasma atomic emission spectrometry

被引:0
作者
Zhuang, ZX [1 ]
Yang, CL [1 ]
Li, B [1 ]
Yang, PY [1 ]
Wang, XR [1 ]
机构
[1] Xiamen Univ, Dept Chem, Res Lab SEDC Analyt Sci Mat & Life Chem, Xiamen 361005, Peoples R China
来源
CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE | 1997年 / 18卷 / 12期
关键词
ICP-AES; thermospray nebulizer; sample introduction;
D O I
暂无
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
A sample introduction system, thermospray nebulizer was coupled to an inductively coupled plasma atomic emission spectrometer. The optimization procedure includes the ICP operation conditions, the capillary temperature, the sample uptake rate, aerosol spray chamber temperature and the cooling water temperature. Under optimized conditions, the sensitivity for 23 elements obtained with the thermospray system is 14 similar to 25 times higher than with a pneumatic nebulizer. The thermospray systems have been shown to improve detection limits an order of magnitude lower than that obtained with pneumatic sample introduction.
引用
收藏
页码:1939 / 1943
页数:5
相关论文
共 6 条
[1]   CROSSED-BEAM LIQUID CHROMATOGRAPH MASS SPECTROMETER COMBINATION [J].
BLAKLEY, CR ;
MCADAMS, MJ ;
VESTAL, ML .
JOURNAL OF CHROMATOGRAPHY, 1978, 158 (OCT) :261-276
[2]   FUNDAMENTAL CHARACTERISTICS OF THERMOSPRAY AEROSOLS AND SAMPLE INTRODUCTION FOR ATOMIC SPECTROMETRY [J].
KOROPCHAK, JA ;
WINN, DH .
APPLIED SPECTROSCOPY, 1987, 41 (08) :1311-1318
[3]   THERMOSPRAY SAMPLE INTRODUCTION TO ATOMIC SPECTROMETRY [J].
KOROPCHAK, JA ;
VEBER, M .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1992, 23 (03) :113-141
[4]   INFLUENCE OF WATER ON CONDITIONS IN THE INDUCTIVELY COUPLED ARGON PLASMA [J].
LONG, SE ;
BROWNER, RF .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1988, 43 (12) :1461-1471
[5]  
MEYER GA, 1985, ICP INFORMATION NEWS, V10, P955
[6]  
RENG R, 1990, SPECTROCHIM ACTA B, V45, P189