Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique

被引:32
作者
Gouda, M. M. [1 ]
Hamzawy, A. [2 ]
Badawi, M. S. [1 ]
El-Khatib, A. M. [1 ]
Thabet, A. A. [3 ]
Abbas, M. I. [1 ]
机构
[1] Univ Alexandria, Fac Sci, Dept Phys, Alexandria 21511, Egypt
[2] Umm Al Qura Univ, Univ Coll, Dept Phys, Makkah Al Mokrama, Saudi Arabia
[3] Pharos Univ Alexandria, Fac Allied Med Sci, Dept Med Equipment Technol, Alexandria, Egypt
关键词
High-purity germanium well-type detector; Cylindrical sources; Self-attenuation; Coincidence summing effect; Efficiency transfer method; COINCIDENCE SUMMING CORRECTIONS; GAMMA-RAY SPECTROMETRY; WELL-TYPE DETECTORS; SELF-ATTENUATION; NAI TL; COMPUTATION; CALIBRATE;
D O I
10.1007/s12648-015-0737-1
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The full-energy peak efficiency of high-purity germanium well-type detector is extremely important to calculate the absolute activities of natural and artificial radionuclides for samples with low radioactivity. In this work, the efficiency transfer method in an integral form is proposed to calculate the full-energy peak efficiency and to correct the coincidence summing effect for a high-purity germanium well-type detector. This technique is based on the calculation of the ratio of the effective solid angles subtended by the well-type detector with cylindrical sources measured inside detector cavity and an axial point source measured out the detector cavity including the attenuation of the photon by the absorber system. This technique can be easily applied in establishing the efficiency calibration curves of well-type detectors. The calculated values of the efficiency are in good agreement with the experimental calibration data obtained with a mixed gamma-ray standard source containing Co-60 and Y-88.
引用
收藏
页码:201 / 210
页数:10
相关论文
共 25 条
  • [1] Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique
    M. M. Gouda
    A. Hamzawy
    M. S. Badawi
    A. M. El-Khatib
    A. A. Thabet
    M. I. Abbas
    Indian Journal of Physics, 2016, 90 : 201 - 210
  • [2] A numerical approach to calculate the full-energy peak efficiency of HPGe well-type detectors using the effective solid angle ratio
    Badawi, M. S.
    Ruskov, I.
    Gouda, M. M.
    El-Khatib, A. M.
    Alotiby, M. F.
    Mohamed, M. M.
    Thabet, A. A.
    Abbas, M. I.
    JOURNAL OF INSTRUMENTATION, 2014, 9
  • [3] Full-Energy Peak Efficiency Determination for the 4 pi Gamma-ray Detector by a Numerical Simulation Method
    Nafee, S. S.
    Abdel-Moneim, A. M.
    Abbas, M. I.
    Badawi, M. S.
    Mahmoud, S. A.
    ARAB JOURNAL OF NUCLEAR SCIENCES AND APPLICATIONS, 2013, 46 (02): : 92 - 106
  • [4] Calculation of full-energy peak efficiency of NaI (Tl) detectors by new analytical approach for parallelepiped sources
    El-Khatib, Ahmed M.
    Badawi, Mohamed S.
    Gouda, Mona M.
    Jovanovic, Slobodan I.
    Dlabac, Aleksandar D.
    Mihaljevic, Nikola N.
    Nafee, Sherif S.
    El-Mallah, Ekram A.
    JOURNAL OF THEORETICAL AND APPLIED PHYSICS, 2013, 7 (01)
  • [5] Calculation of the full-energy peak efficiency of NaI(Tl) detector using the efficiency transfer method for large radioactive cylindrical sources
    El-Khatib, Ahmed. M.
    Gouda, Mona. M.
    Badawi, Mohamed. S.
    Hamzawy, A.
    Hussien, Nancy. S.
    Abbas, Mahmoud. I.
    CHINESE JOURNAL OF PHYSICS, 2016, 54 (04) : 592 - 606
  • [6] HPGe detector full-energy peak efficiency calculation for inhomogeneously activated samples
    Lojze Gačnik
    Radojko Jaćimović
    Journal of Radioanalytical and Nuclear Chemistry, 2017, 311 : 1315 - 1321
  • [7] Full-energy peak determination from total efficiency and peak-to-total ratio calculations
    Aguiar-Amado, Peki P.
    Amado, Valeria A.
    Aguiar, Julio C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 990
  • [8] HPGe detector full-energy peak efficiency calculation for inhomogeneously activated samples
    Gacnik, Lojze
    Jacimovic, Radojko
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2017, 311 (02) : 1315 - 1321
  • [9] Determination of full-energy peak and total efficiency functions for Ge-detectors by iteration calculation using a mixed radionuclide source
    Ishizu, H.
    Yamada, T.
    APPLIED RADIATION AND ISOTOPES, 2018, 134 : 163 - 166
  • [10] A study of the validity of the efficiency transfer method to calculate the peak efficiency using gamma-ray detectors at extremely large distances
    El-Khatib, Ahmed M.
    Badawi, Mohamed S.
    Elzaher, Mohamed A.
    Thabet, Abouzeid A.
    JOURNAL OF THEORETICAL AND APPLIED PHYSICS, 2014, 8 (02)