The AGC all-polyethylene tibial component: A ten-year clinical evaluation

被引:53
作者
Faris, PM [1 ]
Ritter, MA [1 ]
Keating, EM [1 ]
Meding, JB [1 ]
Harty, LD [1 ]
机构
[1] St Francis Hosp, Ctr Hip & Knee Surg, Mooresville, IN 46158 USA
关键词
D O I
10.2106/00004623-200303000-00014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: While high success rates have been achieved in association with other all-polyethylene tibial components, an alarming number of failures have occurred at our institution in association with the use of an all-polyethylene version of the AGC tibial component. The purpose of the present study was to describe the survival of the AGC all-polyethylene tibial component. Methods: Five hundred and thirty-six AGC all-polyethylene tibial components were implanted in 405 patients and were followed over a ten-year period. The average age of the patients at the time of surgery was 70.3 years, the average weight was 78 kg, and the most common diagnosis was osteoarthritis (prevalence, 92.9%). A clinical and radiographic analysis was performed, Knee Society knee and function scores were determined, and Kaplan-Meier survivorship analysis was conducted. Failure was defined as aseptic loosening as evidenced by progressive radiolucent lines and/or revision due to aseptic loosening or collapse. Results: A high rate of failure was noted in the early postoperative period, with a survival rate of 90.04% (95% confidence interval, 87.35% to 92.72%) after three years. At ten years, the survival rate was 68.11% (95% confidence interval, 57.57% to 78.65%). Fifty-eight (73.4%) of seventy-nine failures occurred in association with loosening or collapse of the bone beneath the medial tibial plateau. Conclusion: While some all-polyethylene tibial designs have been successful, the low success rate among knees treated with the AGC all-polyethylene tibial component suggests that the results associated with all-polyethylene tibial components are design-sensitive. Level of Evidence: Therapeutic study, Level IV (case series [no, or historical, control group]). See Instructions to Authors for a complete description of levels of evidence.
引用
收藏
页码:489 / 493
页数:5
相关论文
共 12 条
[1]   Low-conforming all-polyethylene tibial component not inferior to metal-backed component in cemented total knee arthroplasty -: Prospective, randomized radiostereometric analysis study of the AGC total knee prosthesis [J].
Adalberth, G ;
Nilsson, KG ;
Byström, S ;
Kolstad, K ;
Milbrink, J .
JOURNAL OF ARTHROPLASTY, 2000, 15 (06) :783-792
[2]   PERFORMANCE OF THE TIBIAL COMPONENT IN TOTAL KNEE REPLACEMENT - CONVENTIONAL AND REVISION DESIGNS [J].
BARTEL, DL ;
BURSTEIN, AH ;
SANTAVICCA, EA ;
INSALL, JN .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (07) :1026-1033
[3]   THE TOTAL CONDYLAR KNEE PROSTHESIS IN GONARTHROSIS - A 5 TO 9-YEAR FOLLOW-UP OF THE 1ST 100 CONSECUTIVE REPLACEMENTS [J].
INSALL, JN ;
HOOD, RW ;
FLAWN, LB ;
SULLIVAN, DJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1983, 65 (05) :619-628
[4]   REVIEW OF THE ALL-POLYETHYLENE TIBIAL COMPONENT IN TOTAL KNEE ARTHROPLASTY - A MINIMUM 7-YEAR FOLLOW-UP PERIOD [J].
LEE, JG ;
KEATING, EM ;
RITTER, MA ;
FARIS, PM .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1990, (260) :87-92
[5]   A COMPARATIVE-EVALUATION OF TIBIAL COMPONENT DESIGNS OF TOTAL KNEE PROSTHESES [J].
LEWIS, JL ;
ASKEW, MJ ;
JAYCOX, DP .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1982, 64 (01) :129-135
[6]  
Pavone V, 2001, CLIN ORTHOP RELAT R, P18
[7]  
RANAWAT CS, 1988, CLIN ORTHOP RELAT R, P6
[8]  
REILLY D, 1982, CLIN ORTHOP RELAT R, P273
[9]  
Ritter MA, 2001, CLIN ORTHOP RELAT R, P51
[10]  
RITTER MA, 1984, CLIN ORTHOP RELAT R, P264