Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality

被引:51
作者
Fontenele, Rocharles C. [1 ]
Nascimento, Eduarda H. L. [1 ]
Vasconcelos, Taruska V. [1 ]
Noujeim, Marcel [2 ]
Freitas, Deborah Q. [1 ]
机构
[1] Univ Estadual Campinas, Piracicaba Dent Sch, Dept Oral Diag, Div Oral Radiol, Piracicaba, SP, Brazil
[2] Univ Texas Hlth Sci Ctr San Antonio, Dept Comprehens Dent, Div Oral & Maxillofacial Radiol, San Antonio, TX 78229 USA
关键词
cone beam CT; metal artifacts; titanium; zirconium; IN-VITRO ASSESSMENT; COMPUTED-TOMOGRAPHY;
D O I
10.1259/dmfr.20180021
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To evaluate the magnitude of artifacts related to titanium and zirconium implants at different distances and angulations and their impact on cone beam CT(CBCT) image quality. Methods: CBCT images were obtained before and after the insertion of titanium and zirconium implants in a mandible on different CBCT units: Picasso Trio, ProMax 3D and 3D Accuitomo 80. Artifact was assessed by measuring the standard deviation (SD) of gray values and contrast-to-noise ratio (CNR) of 11 regions of interest (ROIs) at different distances (1.5 cm, 2.5 cm and 3.5 cm) and angulations (65 degrees, 90 degrees, 115 degrees and 140 degrees) from implant region. Results: For titanium images, SD values did not differ from those of images without implant in all ROIs; however, some effect occurred in Picasso images as higher values were observed in ROIs closer to the implant (p < 0.05). Zirconium images showed higher SD values than the others in some ROIs for Picasso and ProMax (p < 0.05). In ProMax, the difference was observed even in the farthest ROIs from the implant. CNR values were not influenced by the ROI in Picasso, but presented lower values in ROIs closer to the zirconium implant for ProMax and Accuitomo. Conclusions: The quantity and magnitude of artifacts in CBCT are influenced by the type of implant and CBCT unit. Although they are more pronounced in regions closer to the implant and located at 90 degrees in relation to the mandibular long axis, they can reach as far as 3.5 cm from the artifact-generator object.
引用
收藏
页数:7
相关论文
共 24 条
[1]   Cone beam CT scans with and without artefact reduction in root fracture detection of endodontically treated teeth [J].
Bechara, B. ;
McMahan, C. Alex ;
Moore, W. S. ;
Noujeim, M. ;
Teixeira, F. B. ;
Geha, H. .
DENTOMAXILLOFACIAL RADIOLOGY, 2013, 42 (05)
[2]   Evaluation of a cone beam CT artefact reduction algorithm [J].
Bechara, B. ;
McMahan, C. A. ;
Geha, H. ;
Noujeim, M. .
DENTOMAXILLOFACIAL RADIOLOGY, 2012, 41 (05) :422-428
[3]   Contrast-to-noise ratio with different large volumes in a cone-beam computerized tomography machine: An in vitro study [J].
Bechara, Boulos ;
McMahan, C. Alex ;
Moore, William S. ;
Noujeim, Marcel ;
Geha, Hassem .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2012, 114 (05) :658-665
[4]   In vitro assessment of artifacts induced by titanium dental implants in cone beam computed tomography [J].
Benic, Goran I. ;
Sancho-Puchades, Manuel ;
Jung, Ronald E. ;
Deyhle, Hans ;
Haemmerle, Christoph H. F. .
CLINICAL ORAL IMPLANTS RESEARCH, 2013, 24 (04) :378-383
[5]   Influence of the artefact reduction algorithm of Picasso Trio CBCT system on the diagnosis of vertical root fractures in teeth with metal posts [J].
Bezerra, I. S. Q. ;
Neves, F. S. ;
Vasconcelos, T. V. ;
Ambrosano, G. M. B. ;
Freitas, D. Q. .
DENTOMAXILLOFACIAL RADIOLOGY, 2015, 44 (06)
[6]   ELEVENTH ANNUAL WARREN K. SINCLAIR KEYNOTE ADDRESS-SCIENCE, RADIATION PROTECTION AND NCRP: BUILDING ON THE PAST, LOOKING TO THE FUTURE [J].
Bushberg, Jerrold T. .
HEALTH PHYSICS, 2015, 108 (02) :115-123
[7]  
De Man B, 1999, IEEE T NUCL SCI, V46, P691, DOI 10.1109/23.775600
[8]   Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device [J].
de-Azevedo-Vaz, Sergio Lins ;
Vasconcelos, Karla de Faria ;
Neves, Frederico Sampaio ;
Sousa Melo, Saulo Leonardo ;
Flores Campos, Paulo Sergio ;
Haiter-Neto, Francisco .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 115 (01) :121-127
[9]   E.A.O. guidelines for the use of diagnostic imaging in implant dentistry 2011. A consensus workshop organized by the European Association for Osseointegration at the Medical University of Warsaw [J].
Harris, David ;
Horner, Keith ;
Grondahl, Kerstin ;
Jacobs, Reinhilde ;
Helmrot, Ebba ;
Benic, Goran I. ;
Bornstein, Michael M. ;
Dawood, Andrew ;
Quirynen, Marc .
CLINICAL ORAL IMPLANTS RESEARCH, 2012, 23 (11) :1243-1253
[10]   Guidelines for clinical use of CBCT: a review [J].
Horner, K. ;
O'Malley, L. ;
Taylor, K. ;
Glenny, A-M .
DENTOMAXILLOFACIAL RADIOLOGY, 2015, 44 (01)