The novel active fixation coronary sinus lead: efficacy and safety of transvenous extraction procedure

被引:10
|
作者
Bontempi, Luca [1 ]
Vassanelli, Francesca [1 ]
Ashofair, Najat [1 ]
Inama, Lorenza [1 ]
Mariggio, Davide [2 ]
Cerini, Manuel [1 ]
Curnis, Antonio [1 ]
机构
[1] Univ Brescia, Spedali Civili Hosp, Chair & Unit Cardiol, Brescia, Italy
[2] Medtron Italia, Milan, Italy
来源
EUROPACE | 2016年 / 18卷 / 02期
关键词
Coronary sinus lead; Transvenous lead extraction; Cardiac resynchronization therapy; Left ventricular active fixation lead; CARDIAC RESYNCHRONIZATION THERAPY; EXPERIENCE; STABILITY; TERM;
D O I
10.1093/europace/euv227
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Benefits of cardiac resynchronization therapy (CRT) are well known for heart failure; however, some patients might experience complications related to the coronary sinus (CS) lead (high pacing threshold, phrenic nerve stimulation, and dislodgment) with unfavourable impact on quality of life, costs, and management. Lead stability is one of the most common unmet needs for CRT procedures. Recently, new model Medtronic 20066 Attain Stability(A (R)) (Maastricht, The Netherlands) active fixation LV lead has been released, to overcome this issue. The lead has a small side helix of 0.20 mm (0.008 in.) that allows for secure placement of the lead within the vein at the desired location. We report our first experience with the extraction of this novel active fixation left ventricular lead. In our case, to our knowledge the first reported in humans, the extraction of this new model of active fixation lead was proved to be a safe and effective procedure at 8 months after implantation. Indeed, under angiographic and fluoroscopic check, there was no documented dissection or damage to the CS during and after removal of the lead. The rotation manoeuvre was effective when combined with moderate traction of the lead itself.
引用
收藏
页码:301 / 303
页数:3
相关论文
共 50 条
  • [1] Multicenter Experience with Transvenous Lead Extraction of Active Fixation Coronary Sinus Leads
    Maytin, Melanie
    Carrillo, Roger G.
    Baltodano, Pablo
    Schaerf, Raymond H. M.
    Bongiorni, Maria G.
    Di Cori, Andrea
    Curnis, Antonio
    Cooper, Joshua M.
    Kennergren, Charles
    Epstein, Laurence M.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2012, 35 (06): : 641 - 647
  • [2] Transvenous lead extraction: Efficacy and safety of the procedure in female patients
    Segreti, Luca
    Bongiorni, Maria Grazia
    Barletta, Valentina
    Parollo, Matteo
    Di Cori, Andrea
    Fiorentini, Federico
    Santoro, Mario Giannotti
    De Lucia, Raffaele
    Viani, Stefano
    Grifoni, Gino
    Paperini, Luca
    Sodati, Ezio
    Mazzocchetti, Lorenzo
    Canu, Antonio Maria
    HEART RHYTHM O2, 2023, 4 (10): : 625 - 631
  • [3] Clinical predictors of incomplete coronary sinus lead removal during transvenous lead extraction in patients with cardiac resynchronization therapy
    Hayashi, Katsuhide
    Younis, Arwa
    Callahan, Thomas
    Baranowski, Bryan
    Martin, David O.
    Nakhla, Shady
    Wilkoff, Bruce L.
    HEART RHYTHM, 2023, 20 (06) : 872 - 878
  • [4] Chronic performance of an active fixation coronary sinus lead
    Crossley, George H.
    Exner, Derek
    Mead, R. Hardwin
    Sorrentino, Robert A.
    Hokanson, Robert
    Li, Shelby
    Adler, Stuart
    HEART RHYTHM, 2010, 7 (04) : 472 - 478
  • [5] International experience with transvenous lead extractions of an active-fixation coronary sinus pacing lead
    Gabriels, James K.
    Kim, Beom Soo
    Shoda, Morio
    Birgersdotter-Green, Ulrika
    Di Cori, Andrea
    Zucchelli, Giulio
    Curnis, Antonio
    Arabia, Gianmarco
    Epstein, Laurence M.
    HEART RHYTHM, 2024, 21 (05) : 686 - 687
  • [6] Unexpected challenging case of coronary sinus lead extraction
    Bontempi, Luca
    Tempio, Donatella
    De Vito, Raffaella
    Cerini, Manuel
    Salghetti, Francesca
    Dasseni, Niccolo
    Villa, Clara
    Raweh, Abdallah
    Inama, Lorenza
    Vassanelli, Francesca
    Luzi, Mario
    Curnis, Antonio
    WORLD JOURNAL OF CLINICAL CASES, 2017, 5 (02) : 46 - 49
  • [7] Transvenous Lead Extraction Procedure-Indications, Methods, and Complications
    Boarescu, Paul-Mihai
    Rosian, Adela-Nicoleta
    Rosian, Stefan Horia
    BIOMEDICINES, 2022, 10 (11)
  • [8] Efficacy and safety of transvenous lead extraction in the Chinese octogenarian patients
    Zhou, Xu
    Ze, Feng
    Li, Xuebin
    Wang, Bin
    CLINICAL CARDIOLOGY, 2021, 44 (07) : 971 - 977
  • [9] Active fixation mechanism complicates coronary sinus lead extraction and limits subsequent reimplantation targets
    Cronin, E. M.
    Ingelmo, C. P.
    Rickard, J.
    Wazni, O. M.
    Martin, D. O.
    Wilkoff, B. L.
    Baranowski, B.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2013, 36 (01) : 81 - 86
  • [10] Removal of an active fixation coronary sinus pacing lead 5 years postimplant: A case report
    Carretta, Domenico M.
    Troccoli, Rossella
    Carretta, Francesca
    D'Agostino, Carlo