Application of Artificial Intelligence-based Image Optimization for Computed Tomography Angiography of the Aorta With Low Tube Voltage and Reduced Contrast Medium Volume

被引:12
|
作者
Wang, Yining [1 ]
Yu, Min [2 ]
Wang, Ming [1 ]
Wang, Yun [1 ]
Kong, Lingyan [1 ]
Yi, Yan [1 ]
Wang, Man [1 ]
Li, Yumei [1 ]
Jin, Zhengyu [1 ]
机构
[1] Peking Union Med Coll Hosp, Dept Radiol, Beijing, Peoples R China
[2] Neusoft Med Syst Co, CT Business Unit, Shenyang, Liaoning, Peoples R China
基金
中国国家自然科学基金;
关键词
artificial intelligence; computed tomography angiography; image quality; tube voltage; contrast medium; image noise; noise reduction; ITERATIVE RECONSTRUCTION; CT ANGIOGRAPHY; RADIATION-EXPOSURE; QUALITY;
D O I
10.1097/RTI.0000000000000438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to evaluate the impact of artificial intelligence (AI)-based noise reduction algorithm on aorta computed tomography angiography (CTA) image quality (IQ) at 80 kVp tube voltage and 40 mL contrast medium (CM). Materials and Methods: After obtaining institutional review board approval and 8 written informed consents, 60 patients (35 men, 25 women; age range: 18 to 85 y) referred for aorta CTA examination were assigned to 2 groups at random. Group A underwent an 80 kVp protocol with 40 mL CM (320 mg I/mL). Group A reconstructed with iterative reconstruction was named as group A1 and further AI-based noise reduction was named as group A2. Group B was scanned with standard 120 kVp, 80 mL CM, and iterative reconstruction algorithm. The quantitative assessment of IQ included aorta CT attenuation, noise, signal-to-noise ratio, and contrast-to-noise ratio. A 5-point scale (5-excellent, 1-poor) was used by 2 radiologists independently for qualitative IQ analysis. Results: The image noise significantly decreased while signal-to-noise ratio and contrast-to-noise ratio significantly increased in the order of group A1, B, and A2 (all P<0.05). Compared with group B, the subjective IQ score of group A1 was significantly lower (P<0.05), while that of group A2 had no significant difference (P>0.05). The effective dose and CM volume of group A were reduced by 79.18% and 50%, respectively, than that of group B. Conclusions: The AI-based noise reduction could improve the IQ of aorta CTA with low kV and reduced CM, which achieved the potential of radiation dose and contrast media reduction compared with conventional aorta CTA protocol.
引用
收藏
页码:393 / 399
页数:7
相关论文
共 47 条
  • [1] Artificial Intelligence-Based Prediction of Contrast Medium Doses for Computed Tomography Angiography Using Optimized Clinical Parameter Sets
    Fleitmann, Marja
    Uzunova, Hristina
    Pallenberg, Rene
    Stroth, Andreas M.
    Gerlach, Jan
    Fuerschke, Alexander
    Barkhausen, Joerg
    Bischof, Arpad
    Handels, Heinz
    METHODS OF INFORMATION IN MEDICINE, 2024, 63 (01/02) : 11 - 20
  • [2] Personalized administration of contrast medium with high delivery rate in low tube voltage coronary computed tomography angiography
    Tan, Sock Keow
    Ng, Kwan Hoong
    Yeong, Chai Hong
    Aman, Raja Rizal Azman Raja
    Sani, Fadhli Mohamed
    Aziz, Yang Faridah Abdul
    Sun, Zhonghua
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (04) : 552 - +
  • [3] Artificial Intelligence-Based Image Reconstruction for Computed Tomography: A Survey
    Yan, Quan
    Ye, Yunfan
    Xia, Jing
    Cai, Zhiping
    Wang, Zhilin
    Ni, Qiang
    INTELLIGENT AUTOMATION AND SOFT COMPUTING, 2023, 36 (03): : 2545 - 2558
  • [4] Artificial intelligence-based full aortic CT angiography imaging with ultra-low-dose contrast medium: a preliminary study
    Zhou, Zhen
    Gao, Yifeng
    Zhang, Weiwei
    Bo, Kairui
    Zhang, Nan
    Wang, Hui
    Wang, Rui
    Du, Zhiqiang
    Firmin, David
    Yang, Guang
    Zhang, Heye
    Xu, Lei
    EUROPEAN RADIOLOGY, 2023, 33 (01) : 678 - 689
  • [5] Computed Tomography Angiography of the Aorta-Optimization of Automatic Tube Voltage Selection Settings to Reduce Radiation Dose or Contrast Medium in a Prospective Randomized Trial
    Euler, Andre
    Taslimi, Tilo
    Eberhard, Matthias
    Kobe, Adrian
    Reeve, Kelly
    Zimmermann, Alexander
    Krauss, Andreas
    Gutjahr, Ralf
    Schmidt, Bernhard
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2021, 56 (05) : 283 - 291
  • [6] Low concentration contrast medium for dual-source computed tomography coronary angiography by a combination of iterative reconstruction and low-tube-voltage technique: Feasibility study
    Zheng, Minwen
    Liu, Ying
    Wei, Mengqi
    Wu, Yongjie
    Zhao, Hongliang
    Li, Jian
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) : E92 - E99
  • [7] Computed Tomography Angiography of Coronary Artery Bypass Grafts Low Contrast Media Volume Protocols Adapted to Tube Voltage
    Higashigaito, Kai
    Husarik, Daniela B.
    Barthelmes, Jens
    Plass, Andre R.
    Manka, Robert
    Maisano, Francesco
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2016, 51 (04) : 241 - 248
  • [8] Assessing the Accuracy of an Artificial Intelligence-Based Segmentation Algorithm for the Thoracic Aorta in Computed Tomography Applications
    Artzner, Christoph
    Bongers, Malte N.
    Kaergel, Rainer
    Faby, Sebastian
    Hefferman, Gerald
    Herrmann, Judith
    Nopper, Svenja L.
    Perl, Regine M.
    Walter, Sven S.
    DIAGNOSTICS, 2022, 12 (08)
  • [9] Image quality assessment of artificial intelligence iterative reconstruction for low dose aortic CTA: A feasibility study of 70 kVp and reduced contrast medium volume
    Li, Wanjiang
    You, Yongchun
    Zhong, Sihua
    Shuai, Tao
    Liao, Kai
    Yu, Jianqun
    Zhao, Jin
    Li, Zhenlin
    Lu, Chunyan
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 149
  • [10] Evaluation of image quality and radiation dose in low tube voltage coronary computed tomography angiography
    Mousavi-Gazafroudi, Seyedeh Shokoofeh
    Sajjadieh-Khajouei, Amirreza
    Moradi, Maryam
    Mousavi-Gazafroudi, Seyedeh Shabnam
    Yadegarfar, Ghasem
    Tavakoli, Mohammad Bagher
    ARYA ATHEROSCLEROSIS, 2019, 15 (05) : 205 - 210