Effects of varied dimensions of surgical guides on implant angulations

被引:39
作者
Choi, M [1 ]
Romberg, E [1 ]
Driscoll, CF [1 ]
机构
[1] Univ Maryland Baltimore Cty, Baltimore Coll Dent Surg, Baltimore, MD 21228 USA
关键词
D O I
10.1016/j.prosdent.2004.08.010
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Fabrication of a proper surgical guide is critical for success of implant restorations. The effects of the dimensional factors of the surgical guide on implant placement have not been Studied. Purpose. The purpose of this study was to determine the effect of varied dimensions (diameter, length, and distance between the underside of the surgical guide and the implant recipient site) of a surgical guide on the accuracy of implant angulation. Material and methods. In this in vitro study, 240 implant recipient sites were randomly prepared using varied dimensions of a surgical guide. The varied dimensions of the surgical guide's channel and distance were: channel diameter (2, 3, 4, or 5 mm), channel length (6 or 9 mm), and distance between the underside of the surgical guide and the simulated implant recipient site (2 or 4 mm). From these varying dimensions and distances, 16 combinations of dimensions and distances were tested. For each combination, 15 simulated implant recipient site (SIRS) specimens were prepared. The deviated angulation (DA) from the midpoint of the top surface of the 1- x 1-inch simulated implant recipient site (each simulated implant recipient acrylic block contained 5 SIRS of 1 x 1 inch), in the right-to-left (DA(RL)) and front-to-back (DA(FB)) directions, were measured in degrees using a protractor. The data was analyzed using factorial analysis of variance and Tukey's HSD test (alpha = .05). Results. The DA(RL), in degrees, at a channel length of 9.0 mm (2.33 +/- 1.27) was significantly smaller than at a channel length of 6.0 mm (3.0 +/- 1.42, P = .0001). The DA(RL), in degrees, at a distance of 4.0 min (2.13 +/- 1.16) was significantly smaller than at a distance of 2.0 mm (3.16 +/- 1.39, P = .0001). Also, a significant interaction for DA(RL) was found between diameter and distance (P < .05). For DA(FB), the varying diameters (P < .05), lengths (P = .0001), and distances (P = .0001) showed significant differences. The DA(FB) at a channel length of 9.0 mm (2.56 degrees +/- 1.51) was significantly smaller than that at 6.0 mm (3.82 degrees +/- 1.87). Significant interactions found for DAFB were: diameter by length (P = .0001), diameter by distance (F 4.547, P = .004), and length by distance (F = 11.512, P = .001). Conclusion. Within the limitations of this study, the results suggest channel length as the primary controlling factor in minimizing deviated angulations.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 15 条
[1]  
Blustein R, 1986, Int J Oral Maxillofac Implants, V1, P47
[2]   TEMPLATE FOR POSITIONING AND ANGULATION OF INTRAOSSEOUS IMPLANTS [J].
BURNS, DR ;
CRABTREE, DG ;
BELL, DH .
JOURNAL OF PROSTHETIC DENTISTRY, 1988, 60 (04) :479-483
[3]  
Clelland N L, 1991, Int J Oral Maxillofac Implants, V6, P391
[4]   Surgical guide fabrication for an angled implant [J].
Dixon, DL ;
Breeding, LC .
JOURNAL OF PROSTHETIC DENTISTRY, 1996, 75 (05) :562-565
[5]   OPTIMUM PLACEMENT OF OSSEOINTEGRATED IMPLANTS [J].
ENGELMAN, MJ ;
SORENSEN, JA ;
MOY, P .
JOURNAL OF PROSTHETIC DENTISTRY, 1988, 59 (04) :467-473
[6]   Fabrication of a radiographic and surgical stent for implants with a vacuum former [J].
Ku, YC ;
Shen, YF .
JOURNAL OF PROSTHETIC DENTISTRY, 2000, 83 (02) :252-253
[7]  
Lazzara R J, 1993, J Esthet Dent, V5, P265, DOI 10.1111/j.1708-8240.1993.tb00791.x
[8]   SURGICAL IMPLANT STENT - A DESIGN MODIFICATION AND SIMPLIFIED FABRICATION TECHNIQUE [J].
NEIDLINGER, J ;
LILIEN, BA ;
KALANT, DC .
JOURNAL OF PROSTHETIC DENTISTRY, 1993, 69 (01) :70-72
[9]  
PAREL S M, 1991, International Journal of Oral and Maxillofacial Implants, V6, P207
[10]   FINITE-ELEMENT ANALYSIS OF 6 ENDOSSEOUS IMPLANTS [J].
RIEGER, MR ;
MAYBERRY, M ;
BROSE, MO .
JOURNAL OF PROSTHETIC DENTISTRY, 1990, 63 (06) :671-676