A Highly Sensitive Hydrogen Peroxide Biosensor Based on Hemoglobin Immobilized on Cadmium Sulfide Quantum Dots/Chitosan Composite Modified Glassy Carbon Electrode

被引:13
|
作者
Butwong, Nutthaya [1 ]
Zhou, Lin [2 ,3 ]
Moore, Eric [2 ,3 ,4 ,5 ]
Srijaranai, Supalax [4 ,5 ]
Luong, John H. T. [2 ,3 ]
Glennon, Jeremy D. [2 ,3 ]
机构
[1] Rajamangala Univ Technol, Fac Sci & Liberal Arts, Dept Appl Chem, Nakhon 30000, Ratchasima, Thailand
[2] Natl Univ Ireland Univ Coll Cork, Innovat Chromatog Grp, ISSC, Dept Chem, Cork, Ireland
[3] Natl Univ Ireland Univ Coll Cork, ABCRF, Cork, Ireland
[4] Khon Kaen Univ, Fac Sci, Dept Chem, Khon Kaen 40002, Thailand
[5] Khon Kaen Univ, Ctr Innovat Chem, Fac Sci, Khon Kaen 40002, Thailand
基金
爱尔兰科学基金会;
关键词
Hemoglobin; Cadmium sulphide; Quantum dots; Chitosan; Hydrogen peroxide; LAYER-BY-LAYER; DIRECT ELECTROCHEMISTRY; FLUORESCENCE PROBE; AQUEOUS-SOLUTION; IONIC LIQUID; CHITOSAN; DOTS; NANOPARTICLES; NANOTUBES; FILMS;
D O I
10.1002/elan.201400353
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An L-cysteine capped cadmium sulfide-chitosan nanocomposite has been synthesized, characterized and used for surface modification of a glassy carbon electrode. With direct electron transfer, hemoglobin (Hb) adsorbed strongly on the nanocomposite and displayed excellent bioelectrocatalysis for H2O2. The biosensor was capable of reducing H2O2 at -0.35 V, with a detection limit of 3.13 nM, linearity in the range of 15 nM to 10 mu M and a response time of less than 2 s. The Michaelis-Menten constant (K-M) was 0.57 nM, attesting high bioelectrocatalysis of immobilized Hb for H2O2. Reproducibility of the fabrication method was very satisfactory with a relative standard deviation of 5.3%. The biosensor lost only 6.5% of its original response after 7 days when stored in a pH 7.4 PBS at 4 degrees C.
引用
收藏
页码:2465 / 2473
页数:9
相关论文
共 50 条
  • [41] A sensitive nonenzymatic hydrogen peroxide sensor using cadmium oxide nanoparticles/multiwall carbon nanotube modified glassy carbon electrode
    Butwong, Nutthaya
    Zhou, Lin
    Ng-Eontae, Wittaya
    Burakham, Rodjana
    Moore, Eric
    Srijaranai, Supalax
    Luong, John H. T.
    Glennon, Jeremy D.
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2014, 717 : 41 - 46
  • [42] Direct electrochemistry and biosensing of hydrogen peroxide of horseradish peroxidase immobilized at multiwalled carbon nanotube/alumina-coated silica nanocomposite modified glassy carbon electrode
    Huang, Jian-Lung
    Tsai, Yu-Chen
    SENSORS AND ACTUATORS B-CHEMICAL, 2009, 140 (01): : 267 - 272
  • [43] Horseradish Peroxidase-Immobilized Graphene Oxide-Chitosan Gold Nanocomposites as Highly Sensitive Electrochemical Biosensor for Detection of Hydrogen Peroxide
    Devaraj, Manoj
    Rajendran, Saravanan
    Jebaranjitham, J. Nimita
    Ranjithkumar, Deivasigamani
    Sathiyaraj, Manickam
    Manokaran, Janakiraman
    Sundaravadivel, Elumalai
    Santhanalakshmi, Jayadevan
    Cornejo Ponce, Lorena
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2020, 167 (14)
  • [44] Sensitive Hydrazine Electrochemical Biosensor Based on a Porous Chitosan-Carbon Nanofiber Nanocomposite Modified Electrode
    Ding, Wen
    Wu, Min
    Liang, Meiling
    Ni, Henmei
    Li, Ying
    ANALYTICAL LETTERS, 2015, 48 (10) : 1551 - 1569
  • [45] Hydrogen peroxide biosensor based on direct electrochemistry of soybean Peroxidase immobilized on single-walled carbon nanohorn modified electrode
    Shi, Lihong
    Liu, Xiaoqing
    Niu, Wenxin
    Li, Haijuan
    Han, Shuang
    Chen, Jiuan
    Xu, Guobao
    BIOSENSORS & BIOELECTRONICS, 2009, 24 (05) : 1159 - 1163
  • [46] Glassy carbon electrode modified with a graphene oxide/poly(o-phenylenediamine) composite for the chemical detection of hydrogen peroxide
    Van Hoa Nguyen
    Trung Hieu Tran
    Shim, Jae-Jin
    MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2014, 44 : 144 - 150
  • [47] Direct electrochemistry of sarcosine oxidase on graphene, chitosan and silver nanoparticles modified glassy carbon electrode and its biosensing for hydrogen peroxide
    Zhou, Yunlei
    Yin, Huanshun
    Meng, Xiaomeng
    Xu, Zhenning
    Fu, Yanrong
    Ai, Shiyun
    ELECTROCHIMICA ACTA, 2012, 71 : 294 - 301
  • [48] Hydrogen peroxide biosensor based on hemoglobin-modified gold nanoparticles-screen printed carbon electrode
    Elewi, Ali Saad
    Al-Shammaree, Shatha Abdul Wadood
    Kareem, Abdul M. A.
    Sammarraie, A. L.
    SENSING AND BIO-SENSING RESEARCH, 2020, 28
  • [49] Hydrogen peroxide biosensor based on myoglobin/colloidal gold nanoparticles immobilized on glassy carbon electrode by a Nafion film
    Yang, WW
    Li, YC
    Bai, Y
    Sun, CQ
    SENSORS AND ACTUATORS B-CHEMICAL, 2006, 115 (01) : 42 - 48
  • [50] A sensitive and reliable dopamine biosensor was developed based on the Au@carbon dots-chitosan composite film
    Huang, Qitong
    Zhang, Hanqiang
    Hu, Shirong
    Li, Feiming
    Weng, Wen
    Chen, Jianhua
    Wang, Qingxiang
    He, Yasan
    Zhang, Wuxiang
    Bao, Xiuxiu