LOW TEMPERATURE CALIBRATION FACILITIES AT KRISS

被引:0
作者
Yang, Inseok [1 ]
Kim, Yong-Gyoo [1 ]
Song, Chang Ho [1 ]
Kang, Kee Hoon [1 ]
Gam, Kee Sool [1 ]
机构
[1] Korea Res Inst Stand & Sci, Taejon, South Korea
来源
XIX IMEKO WORLD CONGRESS: FUNDAMENTAL AND APPLIED METROLOGY, PROCEEDINGS | 2009年
关键词
low temperature; cryogenic fixed point; calibration;
D O I
暂无
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We review the low temperature calibration facilities at the Korea Research Institute of Standards and Science. For the calibration of capsule-type standard resistance thermometers, open and sealed cells of cryogenic fixed points and their realization system were manufactured and used. A sealed cell of the triple point of Argon for long-stem standard platinum resistance thermometers was also manufactured. For calibration by comparison, a controllable comparison cryostat and a nitrogen boiling point temperature block were made. We also review our traceability scheme and supporting international key comparisons for the dissemination of low temperature standards.
引用
收藏
页码:1563 / 1566
页数:4
相关论文
共 6 条
[1]   CALIBRATION APPARATUS FOR LONG-STEM AND CAPSULE-TYPE PT RESISTANCE THERMOMETERS BETWEEN THE TRIPLE POINTS OF AR AND IN [J].
ANCSIN, J ;
PHILLIPS, JM .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1984, 55 (08) :1321-1324
[2]   Summary of comparison of realizations of the ITS-90 over the range 83.8058 K to 933.473 K: CCT Key Comparison CCT-K3 [J].
Mangum, BW ;
Strouse, GF ;
Guthrie, WF ;
Pello, R ;
Stock, M ;
Renaot, E ;
Hermier, Y ;
Bonnier, G ;
Marcarino, P ;
Gam, KS ;
Kang, KH ;
Kim, YG ;
Nicholas, JV ;
White, DR ;
Dransfield, TD ;
Duan, Y ;
Qu, Y ;
Connolly, J ;
Rusby, RL ;
Gray, J ;
Sutton, GJM ;
Head, DL ;
Hill, KD ;
Steele, A ;
Nara, K ;
Tegeler, E ;
Noatsch, U ;
Heyer, D ;
Fellmuth, B ;
Thiele-Krivoj, B ;
Duris, S ;
Pokhodun, AI ;
Moiseeva, NP ;
Ivanova, AG ;
de Groot, MJ ;
Dubbeldam, JF .
METROLOGIA, 2002, 39 (02) :179-205
[3]   THE INTERNATIONAL TEMPERATURE SCALE OF 1990 (ITS-90) [J].
PRESTONTHOMAS, H .
METROLOGIA, 1990, 27 (01) :3-10
[4]   CCT-K2: key comparison of capsule-type standard platinum resistance thermometers from 13.8 K to 273.16K [J].
Steele, AG ;
Fellmuth, B ;
Head, DL ;
Hermier, Y ;
Kang, KH ;
Steur, PPM ;
Tew, WL .
METROLOGIA, 2002, 39 (06) :551-571
[5]   Development of the sealed-type triple-point-of-argon cell for long-stem SPRT calibration at KRISS [J].
Yang, I. ;
Song, C. H. ;
Kang, K. H. ;
Kim, Y. -G. ;
Gam, K. S. .
INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2008, 29 (05) :1740-1748
[6]   First-order compensation for time lag in the dynamic calibration of industrial thermometers [J].
Yang, Inseok ;
Song, Chang Ho ;
Kang, Kee Hoon ;
Kim, Yong-Gyoo .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2009, 20 (01)