Influence of observers, threshold values, and measurement methods on volumetric analysis of cerebral aneurysms with three-dimensional rotational angiography

被引:0
|
作者
Cho, Won-Sang [1 ]
Shin, Dong Hoon [2 ]
Lee, Joongyub [3 ]
Lee, Seung Jin [1 ]
机构
[1] Kangwon Natl Univ, Sch Med, Kangwon Natl Univ Hosp, Dept Neurosurg, Chuncheon Si, Gangwon Do, South Korea
[2] Kangwon Natl Univ Hosp, Dept Radiol, Chuncheon Si, Gangwon Do, South Korea
[3] Seoul Natl Univ Hosp, Div Clin Epidemiol, Med Res Collaborating Ctr, Biomed Res Inst, Seoul 110744, South Korea
关键词
DIGITAL-SUBTRACTION-ANGIOGRAPHY; BARE PLATINUM COILS; INTRACRANIAL ANEURYSMS; PACKING; APPROXIMATION; TRIAL;
D O I
10.1136/neurintsurg-2014-011444
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Introduction Exact measurements of cerebral aneurysms are important in terms of their treatment. However, there is no definitive way to estimate the exact volume of an aneurysm. Our aim was to compare aneurysm volume measured under different conditions: threshold values, observers, and measurement methods. Methods 40 aneurysms and 7 phantom models were included in the study. Three-dimensional rotational angiography was used, and volumes were compared based on the following factors: two methods (two-dimensional formula calculation and three-dimensional software measurement); three observers; and three threshold values for imaging. In addition, in the phantom models, measured volumes were compared with real volumes. The consistency of the volume measurements under different conditions was assessed using the intraclass correlation coefficient (ICC). Results The consistency of the measured volumes was excellent in both the patient aneurysms (ICC >0.98) and phantom models (ICC >0.95), irrespective of the influencing factors. Median volume differences were small for observers, threshold values, and methods. When the measured and real volumes of the phantom models were compared, the software measurement achieved the highest reproducibility for real volumes compared with the formula calculation (ICC=0.86-0.93 vs 0.71-0.80). Conclusions Measurement of aneurysm volume showed high agreement within each influencing factor, such as methods, observers, and threshold values.
引用
收藏
页码:203 / 207
页数:5
相关论文
共 22 条
  • [21] Meta-analysis of diagnostic significance of sixty-four-row multi-section computed tomography angiography and three-dimensional digital subtraction angiography in patients with cerebral artery aneurysm
    Guo, Wei
    He, Xu-Ying
    Li, Xi-Feng
    Qian, Dong-Xiang
    Yan, Jian-Quan
    Bu, De-Lin
    Duan, Chuan-Zhi
    JOURNAL OF THE NEUROLOGICAL SCIENCES, 2014, 346 (1-2) : 197 - 203
  • [22] Added diagnostic values of three-dimensional high-resolution proton density-weighted magnetic resonance imaging for unruptured intracranial aneurysms in the circle-of-Willis: Comparison with time-of-flight magnetic resonance angiography
    Yim, Younghee
    Jung, Seung Chai
    Kim, Jung Youn
    Kim, Seon-Ok
    Kim, Byung Jun
    Lee, Deok Hee
    Park, Wonhyoung
    Park, Jung Cheol
    Ahn, Jae Sung
    PLOS ONE, 2020, 15 (12):