Image quality of conventional images of dual-layer SPECTRAL CT: A phantom study

被引:31
|
作者
van Ommen, Fasco [1 ,2 ]
Bennink, Edwin [1 ,2 ]
Vlassenbroek, Alain [3 ]
Dankbaar, Jan Willem [1 ]
Schilham, Arnold M. R. [1 ]
Viergever, Max A. [2 ]
de Jong, Hugo W. A. M. [1 ,2 ]
机构
[1] UMC Utrecht, Radiol & Nucl Med, POB 85500, NL-3508 GA Utrecht, Netherlands
[2] UMC Utrecht, Image Sci Inst, POB 85500, NL-3508 GA Utrecht, Netherlands
[3] Philips Healthcare, CT Clin Sci, Brussels, Belgium
关键词
conventional images; dual-energy CT; dual-layer detector; image-quality; Spectral CT; ITERATIVE RECONSTRUCTION ALGORITHMS; COMPUTED-TOMOGRAPHY; DETECTOR CT;
D O I
10.1002/mp.12959
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeSpectral CT using a dual layer detector offers the possibility of retrospectively introducing spectral information to conventional CT images. In theory, the dual-layer technology should not come with a dose or image quality penalty for conventional images. In this study, we evaluate the influence of a dual-layer detector (IQon Spectral CT, Philips Healthcare) on the image quality of conventional CT images, by comparing these images with those of a conventional but otherwise technically comparable single-layer CT scanner (Brilliance iCT, Philips Healthcare), by means of phantom experiments. MethodsFor both CT scanners, conventional CT images were acquired using four adult scanning protocols: (a) body helical, (b) body axial, (c) head helical, and (d) head axial. A CATPHAN 600 phantom was scanned to conduct an assessment of image quality metrics at equivalent (CTDI) dose levels. Noise was characterized by means of noise power spectra (NPS) and standard deviation (SD) of a uniform region, and spatial resolution was evaluated with modulation transfer functions (MTF) of a tungsten wire. In addition, contrast-to-noise ratio (CNR), image uniformity, CT number linearity, slice thickness, slice spacing, and spatial linearity were measured and evaluated. Additional measurements of CNR, resolution and noise were performed in two larger phantoms. ResultsThe resolution levels at 50%, 10%, and 5% MTF of the iCT and IQon showed small, but significant differences up to 0.25 lp/cm for body scans, and up to 0.2 lp/cm for head scans in favor of the IQon. The iCT and IQon showed perfect CT linearity for body scans, but for head scans both scanners showed an underestimation of the CT numbers of materials with a high opacity. Slice thickness was slightly overestimated for both scanners. Slice spacing was comparable and reconstructed correctly. In addition, spatial linearity was excellent for both scanners, with a maximum error of 0.11 mm. CNR was higher on the IQon compared to the iCT for both normal and larger phantoms with differences up to 0.51. Spatial resolution did not change with phantom size, but noise levels increased significantly. For head scans, IQon had a noise level that was significantly lower than the iCT, on the other hand IQon showed noise levels significantly higher than the iCT for body scans. Still, these differences were well within the specified range of performance of iCT scanners. ConclusionsAt equivalent dose levels, this study showed similar quality of conventional images acquired on iCT and IQon for medium-sized phantoms and slightly degraded image quality for (very) large phantoms at lower tube voltages on the IQon. Accordingly, it may be concluded that the introduction of a dual-layer detector neither compromises image quality of conventional images nor increases radiation dose for normal-sized patients, and slightly degrades dose efficiency for large patients at 120 kVp and lower tube voltages.
引用
收藏
页码:3031 / 3042
页数:12
相关论文
共 50 条
  • [41] Effects of radiation dose levels and spectral iterative reconstruction levels on the accuracy of iodine quantification and virtual monochromatic CT numbers in dual-layer spectral detector CT: an iodine phantom study
    Lu, Xiaomei
    Lu, Zaiming
    Yin, Jiandong
    Gao, Yuying
    Chen, Xingbiao
    Guo, Qiyong
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2019, 9 (02) : 188 - 200
  • [42] SNR performance comparison of dual-layer detector and dual-kVp spectral CT
    Heismann, B. J.
    Wirth, S.
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3820 - 3822
  • [43] Effect of CT Acquisition Parameters on Iodine Density Measurement at Dual-Layer Spectral CT
    Kim, Hyungjin
    Park, Chang Min
    Kang, Chang Kyung
    Yoon, Jinyeong
    Chae, Kum Ju
    Goo, Jin Mo
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 211 (04) : 748 - 754
  • [44] Quantifying potential reduction in contrast dose with monoenergetic images synthesized from dual-layer detector spectral CT
    Tsang, Derek S.
    Merchant, Thomas E.
    Merchant, Sophie E.
    Yagil, Hanna Smith Yoad
    Yagil, Yoad
    Hua, Chia-Ho
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1078):
  • [45] Plaque burden assessment and attenuation measurement of carotid atherosclerotic plaque using virtual monoenergetic images in comparison to conventional polyenergetic images from dual-layer spectral detector CT
    Bai, Xiaoyan
    Gao, Peiyi
    Zhang, Dong
    Zhang, Shaosen
    Liang, Jiang
    Lu, Xiaomei
    Sui, Binbin
    EUROPEAN JOURNAL OF RADIOLOGY, 2020, 132
  • [46] Quantitative and qualitative assessment of plasma cell dyscrasias in dual-layer spectral CT
    S. C. Brandelik
    S. Skornitzke
    T. Mokry
    S. Sauer
    W. Stiller
    J. Nattenmüller
    H. U. Kauczor
    T. F. Weber
    T. D. Do
    European Radiology, 2021, 31 : 7664 - 7673
  • [47] Quantitative calcium-based assessment of osteoporosis in dual-layer spectral CT
    Do, T. D.
    Rahn, S.
    Melzig, C.
    Heussel, C. P.
    Stiller, W.
    Kauczor, H. U.
    Weber, T. F.
    Skornitzke, S.
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 178
  • [48] Material separation with dual-layer CT
    Carmi, Raz
    Naveh, Galit
    Altman, Ami
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1876 - 1878
  • [49] Quantitative and qualitative assessment of plasma cell dyscrasias in dual-layer spectral CT
    Brandelik, S. C.
    Skornitzke, S.
    Mokry, T.
    Sauer, S.
    Stiller, W.
    Nattenmuller, J.
    Kauczor, H. U.
    Weber, T. F.
    Do, T. D.
    EUROPEAN RADIOLOGY, 2021, 31 (10) : 7664 - 7673
  • [50] SUBJECTIVE EVALUATION OF IMAGE QUALITY BASED ON IMAGES OBTAINED WITH A BREAST-TISSUE PHANTOM - COMPARISON WITH A CONVENTIONAL IMAGE QUALITY PHANTOM