Image quality of conventional images of dual-layer SPECTRAL CT: A phantom study

被引:31
|
作者
van Ommen, Fasco [1 ,2 ]
Bennink, Edwin [1 ,2 ]
Vlassenbroek, Alain [3 ]
Dankbaar, Jan Willem [1 ]
Schilham, Arnold M. R. [1 ]
Viergever, Max A. [2 ]
de Jong, Hugo W. A. M. [1 ,2 ]
机构
[1] UMC Utrecht, Radiol & Nucl Med, POB 85500, NL-3508 GA Utrecht, Netherlands
[2] UMC Utrecht, Image Sci Inst, POB 85500, NL-3508 GA Utrecht, Netherlands
[3] Philips Healthcare, CT Clin Sci, Brussels, Belgium
关键词
conventional images; dual-energy CT; dual-layer detector; image-quality; Spectral CT; ITERATIVE RECONSTRUCTION ALGORITHMS; COMPUTED-TOMOGRAPHY; DETECTOR CT;
D O I
10.1002/mp.12959
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeSpectral CT using a dual layer detector offers the possibility of retrospectively introducing spectral information to conventional CT images. In theory, the dual-layer technology should not come with a dose or image quality penalty for conventional images. In this study, we evaluate the influence of a dual-layer detector (IQon Spectral CT, Philips Healthcare) on the image quality of conventional CT images, by comparing these images with those of a conventional but otherwise technically comparable single-layer CT scanner (Brilliance iCT, Philips Healthcare), by means of phantom experiments. MethodsFor both CT scanners, conventional CT images were acquired using four adult scanning protocols: (a) body helical, (b) body axial, (c) head helical, and (d) head axial. A CATPHAN 600 phantom was scanned to conduct an assessment of image quality metrics at equivalent (CTDI) dose levels. Noise was characterized by means of noise power spectra (NPS) and standard deviation (SD) of a uniform region, and spatial resolution was evaluated with modulation transfer functions (MTF) of a tungsten wire. In addition, contrast-to-noise ratio (CNR), image uniformity, CT number linearity, slice thickness, slice spacing, and spatial linearity were measured and evaluated. Additional measurements of CNR, resolution and noise were performed in two larger phantoms. ResultsThe resolution levels at 50%, 10%, and 5% MTF of the iCT and IQon showed small, but significant differences up to 0.25 lp/cm for body scans, and up to 0.2 lp/cm for head scans in favor of the IQon. The iCT and IQon showed perfect CT linearity for body scans, but for head scans both scanners showed an underestimation of the CT numbers of materials with a high opacity. Slice thickness was slightly overestimated for both scanners. Slice spacing was comparable and reconstructed correctly. In addition, spatial linearity was excellent for both scanners, with a maximum error of 0.11 mm. CNR was higher on the IQon compared to the iCT for both normal and larger phantoms with differences up to 0.51. Spatial resolution did not change with phantom size, but noise levels increased significantly. For head scans, IQon had a noise level that was significantly lower than the iCT, on the other hand IQon showed noise levels significantly higher than the iCT for body scans. Still, these differences were well within the specified range of performance of iCT scanners. ConclusionsAt equivalent dose levels, this study showed similar quality of conventional images acquired on iCT and IQon for medium-sized phantoms and slightly degraded image quality for (very) large phantoms at lower tube voltages on the IQon. Accordingly, it may be concluded that the introduction of a dual-layer detector neither compromises image quality of conventional images nor increases radiation dose for normal-sized patients, and slightly degrades dose efficiency for large patients at 120 kVp and lower tube voltages.
引用
收藏
页码:3031 / 3042
页数:12
相关论文
共 50 条
  • [21] Virtual monochromatic images for coronary artery imaging with a spectral photon-counting CT in comparison to dual-layer CT systems: a phantom and a preliminary human study
    Greffier, Joel
    Si-Mohamed, Salim A.
    Lacombe, Hugo
    Labour, Joey
    Djabli, Djamel
    Boccalini, Sara
    Varasteh, Mohammad
    Villien, Marjorie
    Yagil, Yoad
    Erhard, Klaus
    Boussel, Loic
    Beregi, Jean-Paul
    Douek, Philippe C.
    EUROPEAN RADIOLOGY, 2023, 33 (08) : 5476 - 5488
  • [22] Phantom-based quantification of the spectral accuracy in dual-layer spectral CT for pediatric imaging at 100 kVp
    Meyer, Sebastian
    Liu, Leening P.
    Litt, Harold, I
    Halliburton, Sandra S.
    Shapira, Nadav
    Noel, Peter B.
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2023, 13 (02) : 924 - 934
  • [23] Early rectal neoplasm in dual-layer spectral detector computed tomography: dual-layer spectral computed tomography (CT) images improve tumor detection and staging
    Pan, Xuelin
    Wu, Zhihan
    Zhao, Jin
    Zhang, Xinyi
    Zhang, Xiaodi
    Tang, Li
    Yang, Jinlin
    Deng, Kai
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (12) : 8260 - 8271
  • [24] Iodine Imaging at Spectral CT with a Dual-Layer Detector
    Ozguner, O.
    Dhanantwari, A.
    Halliburton, S.
    Wen, G.
    Utrup, S.
    Jordan, D.
    MEDICAL PHYSICS, 2016, 43 (06) : 3657 - 3657
  • [25] Initial exploration of coronary stent image subtraction using dual-layer spectral CT
    Qin, Le
    Gu, ShengJia
    Chen, ChiHua
    Zhang, Huan
    Zhu, ZhenBin
    Chen, XingBiao
    Han, Qun
    Yan, FuHua
    Yang, WenJie
    EUROPEAN RADIOLOGY, 2019, 29 (08) : 4239 - 4248
  • [26] Initial exploration of coronary stent image subtraction using dual-layer spectral CT
    Le Qin
    ShengJia Gu
    ChiHua Chen
    Huan Zhang
    ZhenBin Zhu
    XingBiao Chen
    Qun Han
    FuHua Yan
    WenJie Yang
    European Radiology, 2019, 29 : 4239 - 4248
  • [27] Iodine concentration, HU accuracy, and metal artifacts evaluation on second-generation dual-layer spectral detector CT images with metal implants: a phantom study
    Zhang, Xinming
    Li, Hanwei
    Wang, Haowen
    Liu, Xiaomin
    Deng, Weiwei
    Zhong, Jianqiu
    Fan, Litian
    Ling, Qingqing
    Hu, Debin
    Qi, Hongliang
    Chen, Hongwen
    ACTA RADIOLOGICA, 2024, 65 (09) : 1030 - 1038
  • [28] Dual-layer spectral detector CT arthrography of the shoulder: assessment of image quality and value in differentiating calcium from iodine
    Yoon, Hyeyoung
    Kang, Yusuhn
    Kim, Hyo Jin
    Lee, Eugene
    Ahn, Joong Mo
    Lee, Joon Woo
    ACTA RADIOLOGICA, 2023, 64 (02) : 638 - 647
  • [29] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    Robbert W. van Hamersvelt
    Martin J. Willemink
    Pim A. de Jong
    Julien Milles
    Alain Vlassenbroek
    Arnold M. R. Schilham
    Tim Leiner
    European Radiology, 2017, 27 : 3677 - 3686
  • [30] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    van Hamersvelt, Robbert W.
    Willemink, Martin J.
    de Jong, Pim A.
    Milles, Julien
    Vlassenbroek, Alain
    Schilham, Arnold M. R.
    Leiner, Tim
    EUROPEAN RADIOLOGY, 2017, 27 (09) : 3677 - 3686