Image quality of conventional images of dual-layer SPECTRAL CT: A phantom study

被引:31
|
作者
van Ommen, Fasco [1 ,2 ]
Bennink, Edwin [1 ,2 ]
Vlassenbroek, Alain [3 ]
Dankbaar, Jan Willem [1 ]
Schilham, Arnold M. R. [1 ]
Viergever, Max A. [2 ]
de Jong, Hugo W. A. M. [1 ,2 ]
机构
[1] UMC Utrecht, Radiol & Nucl Med, POB 85500, NL-3508 GA Utrecht, Netherlands
[2] UMC Utrecht, Image Sci Inst, POB 85500, NL-3508 GA Utrecht, Netherlands
[3] Philips Healthcare, CT Clin Sci, Brussels, Belgium
关键词
conventional images; dual-energy CT; dual-layer detector; image-quality; Spectral CT; ITERATIVE RECONSTRUCTION ALGORITHMS; COMPUTED-TOMOGRAPHY; DETECTOR CT;
D O I
10.1002/mp.12959
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeSpectral CT using a dual layer detector offers the possibility of retrospectively introducing spectral information to conventional CT images. In theory, the dual-layer technology should not come with a dose or image quality penalty for conventional images. In this study, we evaluate the influence of a dual-layer detector (IQon Spectral CT, Philips Healthcare) on the image quality of conventional CT images, by comparing these images with those of a conventional but otherwise technically comparable single-layer CT scanner (Brilliance iCT, Philips Healthcare), by means of phantom experiments. MethodsFor both CT scanners, conventional CT images were acquired using four adult scanning protocols: (a) body helical, (b) body axial, (c) head helical, and (d) head axial. A CATPHAN 600 phantom was scanned to conduct an assessment of image quality metrics at equivalent (CTDI) dose levels. Noise was characterized by means of noise power spectra (NPS) and standard deviation (SD) of a uniform region, and spatial resolution was evaluated with modulation transfer functions (MTF) of a tungsten wire. In addition, contrast-to-noise ratio (CNR), image uniformity, CT number linearity, slice thickness, slice spacing, and spatial linearity were measured and evaluated. Additional measurements of CNR, resolution and noise were performed in two larger phantoms. ResultsThe resolution levels at 50%, 10%, and 5% MTF of the iCT and IQon showed small, but significant differences up to 0.25 lp/cm for body scans, and up to 0.2 lp/cm for head scans in favor of the IQon. The iCT and IQon showed perfect CT linearity for body scans, but for head scans both scanners showed an underestimation of the CT numbers of materials with a high opacity. Slice thickness was slightly overestimated for both scanners. Slice spacing was comparable and reconstructed correctly. In addition, spatial linearity was excellent for both scanners, with a maximum error of 0.11 mm. CNR was higher on the IQon compared to the iCT for both normal and larger phantoms with differences up to 0.51. Spatial resolution did not change with phantom size, but noise levels increased significantly. For head scans, IQon had a noise level that was significantly lower than the iCT, on the other hand IQon showed noise levels significantly higher than the iCT for body scans. Still, these differences were well within the specified range of performance of iCT scanners. ConclusionsAt equivalent dose levels, this study showed similar quality of conventional images acquired on iCT and IQon for medium-sized phantoms and slightly degraded image quality for (very) large phantoms at lower tube voltages on the IQon. Accordingly, it may be concluded that the introduction of a dual-layer detector neither compromises image quality of conventional images nor increases radiation dose for normal-sized patients, and slightly degrades dose efficiency for large patients at 120 kVp and lower tube voltages.
引用
收藏
页码:3031 / 3042
页数:12
相关论文
共 50 条
  • [11] Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
    van Hamersvelt, Robbert W.
    Schilham, Arnold M. R.
    Engelke, Klaus
    den Harder, Annemarie M.
    de Keizer, Bart
    Verhaar, Harald J.
    Leiner, Tim
    de Jong, Pim A.
    Willemink, Martin J.
    EUROPEAN RADIOLOGY, 2017, 27 (10) : 4351 - 4359
  • [12] Accuracy of bone mineral density quantification using dual-layer spectral detector CT: a phantom study
    Robbert W. van Hamersvelt
    Arnold M. R. Schilham
    Klaus Engelke
    Annemarie M. den Harder
    Bart de Keizer
    Harald J. Verhaar
    Tim Leiner
    Pim A. de Jong
    Martin J. Willemink
    European Radiology, 2017, 27 : 4351 - 4359
  • [13] Objective image characterization of a spectral CT scanner with dual-layer detector
    Ozguner, Orhan
    Dhanantwari, Amar
    Halliburton, Sandra
    Wen, Gezheng
    Utrup, Steven
    Jordan, David
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (02):
  • [14] Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma
    Nagayama, Yasunori
    Tanoue, Shota
    Inoue, Taihei
    Oda, Seitaro
    Nakaura, Takeshi
    Utsunomiya, Daisuke
    Yamashita, Yasuyuki
    EUROPEAN RADIOLOGY, 2020, 30 (01) : 394 - 403
  • [15] Dual-layer spectral CT improves image quality of multiphasic pancreas CT in patients with pancreatic ductal adenocarcinoma
    Yasunori Nagayama
    Shota Tanoue
    Taihei Inoue
    Seitaro Oda
    Takeshi Nakaura
    Daisuke Utsunomiya
    Yasuyuki Yamashita
    European Radiology, 2020, 30 : 394 - 403
  • [16] Dual-layer spectral detector CT monoenergetic reconstruction improves image quality of non-contrast cerebral CT as compared with conventional single energy CT
    Zhao, Xue-Mei
    Wang, Man
    Wu, Run-Ze
    Dharaiya, Ekta
    Feng, Feng
    Li, Ming-Li
    You, Hui
    Wang, Yun
    Wang, Yi-Ning
    Jin, Zheng-Yu
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 103 : 131 - 138
  • [17] Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting
    Thuy Duong Do
    Rheinheimer, Stephan
    Kauczor, Hans-Ulrich
    Stiller, Wolfram
    Weber, Tim
    Skornitzke, Stephan
    EUROPEAN RADIOLOGY, 2020, 30 (10) : 5709 - 5719
  • [18] Image quality evaluation of dual-layer spectral CT in comparison to single-layer CT in a reduced-dose setting
    Thuy Duong Do
    Stephan Rheinheimer
    Hans-Ulrich Kauczor
    Wolfram Stiller
    Tim Weber
    Stephan Skornitzke
    European Radiology, 2020, 30 : 5709 - 5719
  • [19] Assessment of quantification accuracy and image quality of a full-body dual-layer spectral CT system
    Ehn, Sebastian
    Sellerer, Thorsten
    Muenzel, Daniela
    Fingerle, Alexander A.
    Kopp, Felix
    Duda, Manuela
    Mei, Kai
    Renger, Bernhard
    Herzen, Julia
    Dangelmaier, Julia
    Schwaiger, Benedikt J.
    Sauter, Andreas
    Riederer, Isabelle
    Renz, Martin
    Braren, Rickmer
    Rummeny, Ernst J.
    Pfeiffer, Franz
    Noeel, Peter B.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 204 - 217
  • [20] Virtual monochromatic images for coronary artery imaging with a spectral photon-counting CT in comparison to dual-layer CT systems: a phantom and a preliminary human study
    Joel Greffier
    Salim A. Si-Mohamed
    Hugo Lacombe
    Joey Labour
    Djamel Djabli
    Sara Boccalini
    Mohammad Varasteh
    Marjorie Villien
    Yoad Yagil
    Klaus Erhard
    Loic Boussel
    Jean-Paul Beregi
    Philippe C. Douek
    European Radiology, 2023, 33 : 5476 - 5488