Accelerated Brain DCE-MRI Using Iterative Reconstruction With Total Generalized Variation Penalty for Quantitative Pharmacokinetic Analysis: A Feasibility Study

被引:11
作者
Wang, Chunhao [1 ,2 ]
Yin, Fang-Fang [1 ,2 ]
Kirkpatrick, John P. [1 ,2 ]
Chang, Zheng [1 ,2 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Box 3295, Durham, NC 27710 USA
[2] Duke Univ, Med Phys Grad Program, Durham, NC USA
关键词
brain; DCE-MRI; pharmacokinetic analysis; total generalized variation; iterative reconstruction; CONTRAST-ENHANCED-MRI; ARTERIAL INPUT FUNCTION; DYNAMIC MRI; TEMPORAL-RESOLUTION; IMAGE-RECONSTRUCTION; PROSTATE-CANCER; SENSING MRI; PARAMETERS; BREAST; MODEL;
D O I
10.1177/1533034616649294
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To investigate the feasibility of using undersampled k-space data and an iterative image reconstruction method with total generalized variation penalty in the quantitative pharmacokinetic analysis for clinical brain dynamic contrast-enhanced magnetic resonance imaging. Methods: Eight brain dynamic contrast-enhanced magnetic resonance imaging scans were retrospectively studied. Two k-space sparse sampling strategies were designed to achieve a simulated image acquisition acceleration factor of 4. They are (1) a golden ratio-optimized 32-ray radial sampling profile and (2) a Cartesian-based random sampling profile with spatiotemporal-regularized sampling density constraints. The undersampled data were reconstructed to yield images using the investigated reconstruction technique. In quantitative pharmacokinetic analysis on a voxel-by-voxel basis, the rate constant K-trans in the extended Tofts model and blood flow F-B and blood volume V-B from the 2-compartment exchange model were analyzed. Finally, the quantitative pharmacokinetic parameters calculated from the undersampled data were compared with the corresponding calculated values from the fully sampled data. To quantify each parameter's accuracy calculated using the undersampled data, error in volume mean, total relative error, and cross-correlation were calculated. Results: The pharmacokinetic parameter maps generated from the undersampled data appeared comparable to the ones generated from the original full sampling data. Within the region of interest, most derived error in volume mean values in the region of interest was about 5% or lower, and the average error in volume mean of all parameter maps generated through either sampling strategy was about 3.54%. The average total relative error value of all parameter maps in region of interest was about 0.115, and the average cross-correlation of all parameter maps in region of interest was about 0.962. All investigated pharmacokinetic parameters had no significant differences between the result from original data and the reduced sampling data. Conclusion: With sparsely sampled k-space data in simulation of accelerated acquisition by a factor of 4, the investigated dynamic contrast-enhanced magnetic resonance imaging pharmacokinetic parameters can accurately estimate the total generalized variation-based iterative image reconstruction method for reliable clinical application.
引用
收藏
页码:446 / 460
页数:15
相关论文
共 82 条
[1]   Heterogeneity in DCE-MRI parametric maps: a biomarker for treatment response? [J].
Alic, L. ;
van Vliet, M. ;
van Dijke, C. F. ;
Eggermont, A. M. M. ;
Veenland, J. F. ;
Niessen, W. J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (06) :1601-1616
[2]  
Alic L, 2006, I S BIOMED IMAGING, P944
[3]  
[Anonymous], INT J POLYM SCI
[4]   Imaging anti-angiogenic treatment response with DCE-VCT, DCE-MRI and DWI in an animal model of breast cancer bone metastasis [J].
Baeuerle, Tobias ;
Bartling, Soenke ;
Berger, Martin ;
Schmitt-Graeff, Annette ;
Hilbig, Heidegard ;
Kauczor, Hans-Ulrich ;
Delorme, Stefan ;
Kiessling, Fabian .
EUROPEAN JOURNAL OF RADIOLOGY, 2010, 73 (02) :280-287
[5]  
Blanc-Feraud L., 1995, Proceedings. International Conference on Image Processing (Cat. No.95CB35819), P474, DOI 10.1109/ICIP.1995.529749
[6]   Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint [J].
Block, Kai Tobias ;
Uecker, Martin ;
Frahm, Jens .
MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) :1086-1098
[7]   Total Generalized Variation [J].
Bredies, Kristian ;
Kunisch, Karl ;
Pock, Thomas .
SIAM JOURNAL ON IMAGING SCIENCES, 2010, 3 (03) :492-526
[8]   Microcirculation and microvasculature in breast tumors: Pharmacokinetic analysis of dynamic MR image series [J].
Brix, G ;
Kiessling, F ;
Lucht, R ;
Darai, S ;
Wasser, K ;
Delorme, S ;
Griebel, J .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (02) :420-429
[9]   A review of image denoising algorithms, with a new one [J].
Buades, A ;
Coll, B ;
Morel, JM .
MULTISCALE MODELING & SIMULATION, 2005, 4 (02) :490-530
[10]   Concurrent Stereotactic Radiosurgery and Bevacizumab in Recurrent Malignant Gliomas: A Prospective Trial [J].
Cabrera, Alvin R. ;
Cuneo, Kyle C. ;
Desjardins, Annick ;
Sampson, John H. ;
McSherry, Frances ;
Herndon, James E., II ;
Peters, Katherine B. ;
Allen, Karen ;
Hoang, Jenny K. ;
Chang, Zheng ;
Craciunescu, Oana ;
Vredenburgh, James J. ;
Friedman, Henry S. ;
Kirkpatrick, John P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (05) :873-879