Direct Detection of Benzene, Toluene, and Ethylbenzene at Trace Levels in Ambient Air by Atmospheric Pressure Chemical Ionization Using a Handheld Mass Spectrometer

被引:78
作者
Huang, Guangming [1 ,2 ]
Gao, Liang [1 ,2 ]
Duncan, Jason [1 ,2 ]
Harper, Jason D. [1 ,2 ]
Sanders, Nathaniel L. [1 ,2 ]
Ouyang, Zheng [3 ]
Cooks, R. Graham [1 ,2 ]
机构
[1] Purdue Univ, Dept Chem, W Lafayette, IN 47907 USA
[2] Purdue Univ, Ctr Analyt Instrumentat Dev, W Lafayette, IN 47907 USA
[3] Purdue Univ, Weldon Sch Biomed Engn, W Lafayette, IN 47907 USA
关键词
SOLID-PHASE MICROEXTRACTION; VOLATILE ORGANIC-COMPOUNDS; ON-SITE DETECTION; WARFARE AGENTS; MEMBRANE INLET; QUANTITATIVE-ANALYSIS; DETECTION SYSTEM; TOXIC COMPOUNDS; SAMPLE VOLUME; GASOLINE;
D O I
10.1016/j.jasms.2009.09.018
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The capabilities of a portable mass spectrometer for real-time monitoring of trace levels of benzene, toluene, and ethylbenzene in air are illustrated. An atmospheric pressure interface was built to implement atmospheric pressure chemical ionization for direct analysis of gas-phase samples on a previously described miniature mass spectrometer (Gao et al. Anal. Chem. 2006, 78, 5994-6002). Linear dynamic ranges, limits of detection and other analytical figures of merit were evaluated: for benzene, a limit of detection of 0.2 parts-per-billion was achieved for air samples without any sample preconcentration. The corresponding limits of detection for toluene and ethylbenzene were 0.5 parts-per-billion and 0.7 parts-per-billion, respectively. These detection limits are well below the compounds' permissible exposure levels, even in the presence of added complex mixtures of organics at levels exceeding the parts-per-million level. The linear dynamic ranges of benzene, toluene, and ethylbenzene are limited to approximately two orders of magnitude by saturation of the detection electronics. (J Am Soc Mass Spectrom 2010, 21, 132-135) (C) 2010 American Society for Mass Spectrometry
引用
收藏
页码:132 / 135
页数:4
相关论文
共 30 条
  • [1] Water solubilization of ethanol and BTEX from gasoline: on-line monitoring by membrane introduction mass spectrometry
    Alberici, RM
    Zampronio, CG
    Poppi, RJ
    Eberlin, MN
    [J]. ANALYST, 2002, 127 (02) : 230 - 234
  • [2] AUTOMATION AND OPTIMIZATION OF SOLID-PHASE MICROEXTRACTION
    ARTHUR, CL
    KILLAM, LM
    BUCHHOLZ, KD
    PAWLISZYN, J
    BERG, JR
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (17) : 1960 - 1966
  • [3] Sorption hysteresis of benzene in charcoal particles
    Braida, WJ
    Pignatello, JJ
    Lu, YF
    Ravikovitch, PI
    Neimark, AV
    Xing, BS
    [J]. ENVIRONMENTAL SCIENCE & TECHNOLOGY, 2003, 37 (02) : 409 - 417
  • [4] CHEMICAL IONIZATION MASS-SPECTROMETRY OF NON-VOLATILE ORGANIC-COMPOUNDS
    CARROLL, DI
    DZIDIC, I
    HORNING, MG
    MONTGOMERY, FE
    NOWLIN, JG
    STILLWELL, RN
    THENOT, JP
    HORNING, EC
    [J]. ANALYTICAL CHEMISTRY, 1979, 51 (11) : 1858 - 1860
  • [5] ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY - CORONA DISCHARGE ION-SOURCE FOR USE IN LIQUID CHROMATOGRAPH MASS SPECTROMETER-COMPUTER ANALYTICAL SYSTEM
    CARROLL, DI
    DZIDIC, I
    STILLWELL, RN
    HAEGELE, KD
    HORNING, EC
    [J]. ANALYTICAL CHEMISTRY, 1975, 47 (14) : 2369 - 2373
  • [6] SUBPICOGRAM DETECTION SYSTEM FOR GAS-PHASE ANALYSIS BASED UPON ATMOSPHERIC-PRESSURE IONIZATION (API) MASS-SPECTROMETRY
    CARROLL, DI
    DZIDIC, I
    STILLWELL, RN
    HORNING, MG
    HORNING, EC
    [J]. ANALYTICAL CHEMISTRY, 1974, 46 (06) : 706 - 710
  • [7] Analysis of gaseous toxic industrial compounds and chemical warfare agent simulants by atmospheric pressure ionization mass spectrometry
    Cotte-Rodríguez, I
    Justes, DR
    Nanita, SC
    Noll, RJ
    Mulligan, CC
    Sanders, NL
    Cooks, RG
    [J]. ANALYST, 2006, 131 (04) : 579 - 589
  • [8] Fast and direct screening of polyaromatic hydrocarbon (PAH)-contaminated sand using a miniaturized membrane inlet mass spectrometer (mini-MIMS)
    Frandsen, Helle
    Janfelt, Christian
    Lauritsen, Frants R.
    [J]. RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2007, 21 (10) : 1574 - 1578
  • [9] Handheld rectilinear ion trap mass spectrometer
    Gao, Liang
    Song, Qingyu
    Patterson, Garth E.
    Cooks, R. Graham
    Ouyang, Zheng
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (17) : 5994 - 6002
  • [10] Effect of sample volume on quantitative analysis by solid-phase microextraction .1. Theoretical considerations
    Gorecki, T
    Pawliszyn, J
    [J]. ANALYST, 1997, 122 (10) : 1079 - 1086