Monte Carlo characterization of the M-19 high dose rate Iridium-192 brachytherapy source

被引:26
作者
Medich, David C.
Munro, John J., III
机构
[1] Univ Massachusetts, Radiat Lab, Lowell, MA 01854 USA
[2] Source Prod & Equipment Co Inc, St Rose, LA 70087 USA
关键词
brachytherapy; iridium-192; radiation dosimetry; Monte Carlo; MCNP;
D O I
10.1118/1.2733809
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The MCNP5 Monte Carlo code was used to simulate the dosimetry of an M-19 iridium-192 high dose rate brachytherapy source in both air/vacuum and water environments with the in-air photon spectrum filtered to remove low-energy photons below delta= 10 keV. Dosimetric data was organized into an away-along table and was used to derive the updated AAPM Task Group Report No. 43 (TG-43U1) parameters including S-K, D(r, theta), Lambda, g(L)(r), F(r, theta), phi(an)(r), and (phi) over bar (an) and their respective statistical uncertainties. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1999 / 2006
页数:8
相关论文
共 50 条
[31]   Three-year results of treatment for prostate cancer with low-dose rate temporary iridium-192 brachytherapy [J].
Momma, T ;
Saito, S ;
Toya, K ;
Yorozu, A ;
Dokiya, T ;
Murai, M .
INTERNATIONAL JOURNAL OF UROLOGY, 2006, 13 (03) :218-223
[32]   Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations [J].
Ahn, Woo Sang ;
Choi, Wonsik ;
Shin, Seong Soo ;
Jung, Jinhong .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2014, 65 (04) :557-564
[33]   Application of a pelvic phantom in brachytherapy dosimetry for high-dose-rate (HDR) 192Ir source based on Monte Carlo simulations [J].
Woo Sang Ahn ;
Wonsik Choi ;
Seong Soo Shin ;
Jinhong Jung .
Journal of the Korean Physical Society, 2014, 65 :557-564
[34]   A Monte Carlo study on dose distribution evaluation of Flexisource Ir-192 brachytherapy source [J].
Alizadeh, Majid ;
Ghorbani, Mandi ;
Haghparast, Abbas ;
Zare, Naser ;
Moghaddas, Toktam Ahmadi .
REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2015, 20 (03) :204-209
[35]   EXPERIMENTAL VALIDATION OF MONTE-CARLO DOSE CALCULATIONS ABOUT A HIGH-INTENSITY IR-192 SOURCE FOR PULSED DOSE-RATE BRACHYTHERAPY [J].
VALICENTI, RK ;
KIROV, AS ;
MEIGOONI, AS ;
MISHRA, V ;
DAS, RK ;
WILLIAMSON, JF .
MEDICAL PHYSICS, 1995, 22 (06) :821-829
[36]   A Monte Carlo investigation of the dose distribution for 60Co high dose rate brachytherapy source in water and in different media [J].
Badry, Hamza ;
Oufni, Lhoucine ;
Ouabi, Hmad ;
Hirayama, Hideo .
APPLIED RADIATION AND ISOTOPES, 2018, 136 :104-110
[37]   Postoperative single versus multiple fractions high-dose rate iridium-192 surface mould brachytherapy for keloid treatment: a comparative study [J].
Ahmad, Misbah ;
Ahmad, Habib ;
Khattak, Muhammad Rauf ;
Shah, Kamran Ali ;
Javed, Nabila ;
Shah, Syed Jawad Ali ;
Shaheen, Wajeeha ;
Mansoor, Kehkashan .
JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (01) :60-65
[38]   Two different sizes of Valencia applicators in non-melanoma skin cancer treatment with iridium-192 high-dose-rate brachytherapy [J].
Laliscia, Concetta ;
Coccia, Natalina ;
Fuentes, Taiusha ;
Perrone, Franco ;
Paiar, Fabiola .
JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2021, 13 (06) :615-619
[39]   Sub-second high dose rate brachytherapy Monte Carlo dose calculations with bGPUMCD [J].
Hissoiny, Sami ;
D'Amours, Michel ;
Ozell, Benoit ;
Despres, Philippe ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2012, 39 (07) :4559-4567
[40]   A Monte Carlo study of verification imaging in high dose rate brachytherapy [J].
Sheikh-Bagheri, D ;
Munro, P .
MEDICAL PHYSICS, 1998, 25 (04) :404-414