Validation of the Withings Sleep Analyzer, an under-the-mattress device for the detection of moderate-severe sleep apnea syndrome

被引:71
作者
Edouard, Paul [1 ]
Campo, David [1 ]
Bartet, Pierre [1 ]
Yang, Rui-Yi [1 ]
Bruyneel, Marie [2 ]
Roisman, Gabriel [3 ]
Escourrou, Pierre [4 ]
机构
[1] Withings, 2 Rue Maurice Hartmann, F-92130 Issy Les Moulineaux, France
[2] St Pierre Univ Hosp, Chest Serv, Brussels, Belgium
[3] Antoine Beclere Hosp, AP HP, Sleep Med Dept, Clamart, France
[4] CIS, Paris, France
来源
JOURNAL OF CLINICAL SLEEP MEDICINE | 2021年 / 17卷 / 06期
关键词
e-health; home monitoring; under-the-mattress sensor; sleep apnea; polysomnography; polygraphy; deep learning; screening; RESPIRATORY EVENTS; AMERICAN ACADEMY; EPIDEMIOLOGY; POPULATION; POLYSOMNOGRAPHY; AUTOTITRATION; TECHNOLOGY; PREVALENCE; AGREEMENT; DIAGNOSIS;
D O I
10.5664/jcsm.9168
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Objectives: To assess the diagnostic performance of a nonintrusive device placed under the mattress to detect sleep apnea syndrome. Methods: One hundred eighteen patients suspected to have obstructive sleep apnea syndrome completed a night at a sleep clinic with a simultaneous polysomnography (PSG) and recording with the Withings Sleep Analyzers. PSG nights were scored twice: first as simple polygraphy, then as PSG. Results: Average (standard deviation) apnea-hypopnea index from PSG was 31.2 events/h (25.0) and 32.8 events/h (29.9) according to the Withings Sleep Analyzers. The mean absolute error was 9.5 events/h. The sensitivity, specificity, and area under the receiver operating characteristic curve at thresholds of apnea-hypopnea index >= 15 events/h were, respectively, sensitivity (Se)(15) = 88.0%, specificity (Sp)(15) = 88.6%, and area under the receiver operating characteristic curve (AUROC)(15) = 0.926. At the threshold of apnea-hypopnea index 30 events/h, results included Se-30 = 86.0%, Sp(30) = 91.2%, AUROC(30) = 0.954. The average total sleep time from PSG and the Withings Sleep Analyzers was 366.6 (61.2) and 392.4 (67.2) minutes, sleep efficiency was 82.5% (11.6) and 82.6% (11.6), and wake after sleep onset was 62.7 (48.0) and 45.2 (37.3) minutes, respectively. Conclusions: Withings Sleep Analyzers accurately detect moderate-severe sleep apnea syndrome in patients suspected of sleep apnea syndrome. This simple and automated approach could be of great clinical value given the high prevalence of sleep apnea syndrome in the general population.
引用
收藏
页码:1217 / 1227
页数:11
相关论文
共 39 条
[1]  
American Academy of Sleep Medicine, 2016, HIDD HLTH CRIS COST
[2]   Estimation of the global prevalence and burden of obstructive sleep apnoea: a literature-based analysis [J].
Benjafield, Adam V. ;
Ayas, Najib T. ;
Eastwood, Peter R. ;
Heinzer, Raphael ;
Ip, Mary S. M. ;
Morrell, Mary J. ;
Nunez, Carlos M. ;
Patel, Sanjay R. ;
Penzel, Thomas ;
Pepin, Jean-Louis D. ;
Peppard, Paul E. ;
Sinha, Sanjeev ;
Tufik, Sergio ;
Valentine, Kate ;
Malhotra, Atul .
LANCET RESPIRATORY MEDICINE, 2019, 7 (08) :687-698
[3]  
Berry R.B., 2012, J.o.c.s.m, V8, P597
[4]  
Berry RB, 2008, SLEEP, V31, P1423
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Bousquet Jean., 2007, World Health Organization
[7]  
CATTERALL JR, 1982, LANCET, V1, P301
[8]   High STOP-Bang score indicates a high probability of obstructive sleep apnoea [J].
Chung, F. ;
Subramanyam, R. ;
Liao, P. ;
Sasaki, E. ;
Shapiro, C. ;
Sun, Y. .
BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 (05) :768-775
[9]   Obstructive Sleep Apnea Devices for Out-Of-Center (OOC) Testing: Technology Evaluation [J].
Collop, Nancy A. ;
Tracy, Sharon L. ;
Kapur, Vishesh ;
Mehra, Reena ;
Kuhlmann, David ;
Fleishman, Sam A. ;
Ojile, Joseph M. .
JOURNAL OF CLINICAL SLEEP MEDICINE, 2011, 7 (05) :531-548
[10]  
Davidovich MLY, 2016, 2016 COMP CARD C CIN