Fill volume, humidification and heat effects on aerosol delivery and fugitive emissions during noninvasive ventilation

被引:48
作者
Saeed, Haitham [1 ]
Mohsen, Marwa [1 ]
Fink, James B. [2 ]
Dailey, Patricia [3 ]
Eldin, Abeer Salah [4 ]
Abdelrahman, Maha M. [5 ]
Elberry, Ahmed A. [6 ]
Rabea, Hoda [1 ]
Hussein, Raghda R. S. [1 ]
Abdelrahim, Mohamed E. A. [1 ,7 ]
机构
[1] Beni Suef Univ, Fac Pharm, Dept Clin Pharm, Bani Suwayf, Egypt
[2] Georgia State Univ, Sch Hlth Profess, Div Resp Therapy, Atlanta, GA 30303 USA
[3] Aerogen Ltd, Med Affairs Clin, Galway, Ireland
[4] Beni Suef Univ, Fac Med, Resp Dept, Bani Suwayf, Egypt
[5] Beni Suef Univ, Fac Pharm, Dept Analyt Chem, Bani Suwayf, Egypt
[6] Beni Suef Univ, Fac Med, Clin Pharmacol, Bani Suwayf, Egypt
[7] Ahram Canadian Univ, Dept Clin Pharm, Fac Pharm, Giza, Egypt
关键词
Nebulizer; Non-invasive ventilation; Fill volume; Humidification; Heat; Fugitive aerosol; VIBRATING MESH NEBULIZER; MECHANICAL VENTILATION; BRONCHODILATOR DELIVERY; REALISTIC MODEL; DEPOSITION; FLOW; EXPOSURE; PERFORMANCE; PARTICLES; SETTINGS;
D O I
10.1016/j.jddst.2017.04.026
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
The aim of this study was to determine the-effect of fill-volume, heat-and-humidification on aerosol delivery and fugitive-aerosol-emissions from nebulizers placed proximal to the patient in single-limb-non-invasive-ventilation (NIV) adult-model. We compared a vibrating-mesh-nebulizer (VMN) and jet-nebulizer (JN) placed proximal to breathing-simulator with adult-settings (I:E 1:3, 15 breaths.min(-1), tidal-volume 500 mL). Ventilator was set to (20 cm.H2O peak-inspiratory-pressure, peak-expiratory-pressure 5cmH(2)O) in spontaneous-mode. Each nebulizer was used to compare total-inhaled-dose (TID) collected in inhalation-filter of 3 different fill-volumes (5000 mu g salbutamol diluted to 1, 2 and 4 mL using normal-saline) and 3 heat-and-humidification-conditions (no-heat and no-humidification, humidification with no-heat and heat-with humidification) using 1 mL respirable-solution containing 5000 mu g salbutamol diluted to 2 mL using normal-saline. TID with 1, 2 and 4 mL fill-volumes were similar with VMN but increased upto twofold with JN (p < 0.01). No significant effect of humidification or heat on TID was found for both VMN and JN. Upto 50% of emitted-aerosol escaped through the fixed-orifice-leak. Efficiency of JN varied with fill-volume. In contrast, the lower-residual-volume of VMN produced similar efficiency across fill-volumes with higher TID than JN. No need to switch-off the humidifier for aerosol-delivery in NIV, since variations of heat-and-humidity did not impact TID with both nebulizer. However, fugitive-aerosol-emissions represent a potential-risk to both healthcare-providers and acute-care-environment. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:372 / 378
页数:7
相关论文
共 46 条
[1]   Aerodynamic characteristics of nebulized terbutaline sulphate using the Andersen Cascade Impactor compared to the Next Generation Impactor [J].
Abdelrahim, Mohamed E. .
PHARMACEUTICAL DEVELOPMENT AND TECHNOLOGY, 2011, 16 (02) :137-145
[2]   In-vitro characterisation of the nebulised dose during non-invasive ventilation [J].
Abdelrahim, Mohamed E. ;
Plant, Paul ;
Chrystyn, Henry .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 2010, 62 (08) :966-972
[3]  
[Anonymous], 2017, PULM THER
[4]  
[Anonymous], J AM MED ASS
[5]  
[Anonymous], 2017, ARCH PULMONOLOGY RES, DOI DOI 10.17352/APRC.000017
[6]  
[Anonymous], AARC TIMES
[7]  
[Anonymous], RESP CARE
[8]  
Ari A., 2016, Indian J. Respir. Care, V5, P677
[9]   Quantifying Aerosol Delivery in Simulated Spontaneously Breathing Patients With Tracheostomy Using Different Humidification Systems With or Without Exhaled Humidity [J].
Ari, Arzu ;
Harwood, Robert ;
Sheard, Meryl ;
Alquaimi, Maher Mubarak ;
Alhamad, Bshayer ;
Fink, James B. .
RESPIRATORY CARE, 2016, 61 (05) :600-606
[10]   Factors affecting bronchodilator delivery in mechanically ventilated adults [J].
Ari, Arzu ;
Fink, James B. .
NURSING IN CRITICAL CARE, 2010, 15 (04) :192-203