Polyethylene wear in Oxford unicompartmental knee replacement A RETRIEVAL STUDY OF 47 BEARINGS

被引:69
作者
Kendrick, B. J. L.
Longino, D.
Pandit, H.
Svard, U.
Gill, H. S. [1 ]
Dodd, C. A. F. [1 ]
Murray, D. W. [1 ]
Price, A. J. [1 ]
机构
[1] Nuffield Orthopaed Ctr, Biomed Res Unit, Nuffield Dept Orthopaed Rheumatol & Musculoskelet, Oxford OX3 7LD, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2010年 / 92B卷 / 03期
关键词
MOLECULAR-WEIGHT POLYETHYLENE; ARTHROPLASTY; COMPONENTS; PROSTHESES;
D O I
10.1302/0301-620X.92B3.22491
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The Oxford Unicompartmental Knee replacement (UKR) was introduced as a design to reduce polyethylene wear. There has been one previous retrieval study involving this implant, which reported very low rates of wear in some specimens but abnormal patterns of wear in others. There has been no further investigation of these abnormal patterns. The bearings were retrieved from 47 patients who had received a medial Oxford UKR for anteromedial osteoarthritis of the knee. None had been studied previously. The mean time to revision was 8.4 years (SD 4.1), with 20 having been implanted for over ten years. The macroscopic pattern of polyethylene wear and the linear penetration were recorded for each bearing. The mean rate of linear penetration was 0.07 mm/year. The patterns of wear fell into three categories, each with a different rate of linear penetration; 1) no abnormal macroscopic wear and a normal articular surface, n = 16 (linear penetration rate = 0.01 mm/year); 2) abnormal macroscopic wear and normal articular surfaces with extra-articular impingement, n = 16 (linear penetration rate = 0.05 mm/year); 3) abnormal macroscopic wear and abnormal articular surfaces with intra-articular impingement +/- signs of non-congruous articulation, n = 15 (linear penetration rate = 0.12 mm/year). The differences in linear penetration rate were statistically significant (p < 0.001). These results show that very low rates of polyethylene wear are possible if the device functions normally. However, if the bearing displays suboptimal function (extra-articular, intra-articular impingement or incongruous articulation) the rates of wear increase significantly.
引用
收藏
页码:367 / 373
页数:7
相关论文
共 26 条
[1]  
Altman DG, 1990, PRACTICAL STAT MED R, DOI DOI 10.1201/9780429258589
[2]  
[Anonymous], 1981, STAT METHODS RATES P
[3]   Polyethylene wear in a non-congruous unicompartmental knee replacement: a retrieval analysis [J].
Ashraf, T ;
Newman, JH ;
Desai, VV ;
Beard, D ;
Nevelos, JE .
KNEE, 2004, 11 (03) :177-181
[4]  
BARTLEY RE, 1994, CLIN ORTHOP RELAT R, P18
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]  
Deng M, 1998, J BIOMED MATER RES, V40, P214, DOI 10.1002/(SICI)1097-4636(199805)40:2<214::AID-JBM6>3.0.CO
[7]  
2-O
[8]   POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[9]   2009 Knee Society Presidential Guest Lecture Polyethylene Wear in Total Knees [J].
Fisher, John ;
Jennings, Louise M. ;
Galvin, Alison L. ;
Jin, Zhongmin M. ;
Stone, Martin H. ;
Ingham, Eileen .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (01) :12-18
[10]   Review article: Osteolysis after total knee arthroplasty [J].
Gupta, Sanjay K. ;
Chu, Alicc ;
Ranawat, Amar S. ;
Slamin, John ;
Ranawat, Chitranjan S. .
JOURNAL OF ARTHROPLASTY, 2007, 22 (06) :787-799