Echocardiographic outflow pump ramp test in centrifugal-flow left ventricular assist device

被引:3
作者
Iacovoni, Attilio [1 ]
Vittori, Claudia [1 ]
Fontana, Alessandra [1 ]
Carobbio, Alessandra [2 ]
Fino, Carlo [1 ]
D'Elia, Emilia [1 ]
Terzi, Amedeo [1 ]
Senni, Michele [1 ]
机构
[1] Reg Healthcare & Social Serv ASST, Cardiovasc Dept, Papa Giovanni Hosp 23, Bergamo, Italy
[2] Reg Healthcare & Social Serv ASST, FROM Res Fdn, Papa Giovanni Hosp 23, Bergamo, Italy
关键词
Advanced heart failure; Left ventricular assist device; Pump thrombosis; SPEED OPTIMIZATION; THROMBOSIS; SYSTEM;
D O I
10.5301/ijao.5000573
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study sought to develop a novel echocardiogram outflow ramp test to detect device malfunctions in centrifugal-flow left ventricular assist devices (LVADs). This new ramp pump test is based on the direct analyses of systolic and diastolic ratio (S/D) Doppler velocity in the outflow cannula in the HeartWare LVAD during progressive increases in speed. The results showed that in patients with normal pump function, the Doppler velocity S/D ratio gradually decreased during LVAD speed increases. This test is easily performed and seems promising to detect normal pump function in patients assisted by a centrifugal flow LVAD.